Thermal noise in biomagnetic measurements

被引:49
|
作者
Nenonen, J [1 ]
Montonen, J [1 ]
Katila, T [1 ]
机构
[1] UNIV HELSINKI,CENT HOSP,BIOMAG LAB,SF-00290 HELSINKI,FINLAND
来源
REVIEW OF SCIENTIFIC INSTRUMENTS | 1996年 / 67卷 / 06期
关键词
D O I
10.1063/1.1147514
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Studies of weak magnetic fields are generally influenced by magnetic noise emanating from thermal agitation of electric charge (Johnson noise) in electrically conducting materials surrounding the magnetic-field sensor. In this article, the thermal magnetic noise fields generated by slabs with high electric conductivity (copper, aluminum) or high magnetic permeability (mu metal) are studied. The analysis is based both on a previously published phenomenological model and on measurements with an ultrasensitive superconducting magnetometer. Both the spectral densities and spatial correlations of the magnetic field fluctuations are evaluated. The computed correlation coefficients are utilized to develop a practical method for estimating the thermal noise due to thin conducting foils, such as thermal radiation shields in a cryogenic measurement Dewar. Also experiments to reduce the Dewar noise are reported. Finally, estimations are presented for the thermal noise fields arising in the walls of a magnetically shielded room. In practice, thermal magnetic noise, particularly due to the superinsulation in cryogenic Dewars, can be the limiting factor of sensitivity in measurements of weak biomagnetic signals arising in the human heart and brain. The results are useful in the estimation and minimization of the contribution of thermal noise. (C) 1996 American Institute of Physics.
引用
收藏
页码:2397 / 2405
页数:9
相关论文
共 50 条
  • [1] THERMAL MAGNETIC NOISE OF DEWARS FOR BIOMAGNETIC MEASUREMENTS
    KASAI, N
    SASAKI, K
    KIRYU, S
    SUZUKI, Y
    CRYOGENICS, 1993, 33 (02) : 175 - 179
  • [2] NOISE-REDUCTION OF MAGNETIC MODULATORS FOR BIOMAGNETIC MEASUREMENTS
    WADA, T
    BIOTELEMETRY AND PATIENT MONITORING, 1980, 7 (01) : 13 - 14
  • [3] SQUID BIOMAGNETIC MEASUREMENTS
    FARRELL, DE
    ZANZUCCHI, P
    CRC CRITICAL REVIEWS IN SOLID STATE AND MATERIALS SCIENCES, 1983, 11 (02): : 99 - 122
  • [4] Characteristics and prediction of magnetic noise due to DC electric railcars for biomagnetic measurements
    Yamazaki, K
    Kato, K
    Kobayashi, K
    Uchikawa, Y
    Kumagai, Y
    Haga, A
    Fujiwara, K
    IEEE TRANSACTIONS ON MAGNETICS, 2001, 37 (04) : 2884 - 2887
  • [5] RECONSTRUCTING CURRENT DISTRIBUTIONS FROM BIOMAGNETIC MEASUREMENTS UNDER LARGE EXTERNAL NOISE DISTURBANCES
    SEKIHARA, K
    SCHOLZ, B
    BRUDER, H
    GRAUMANN, R
    IEEE TRANSACTIONS ON MEDICAL IMAGING, 1994, 13 (01) : 144 - 151
  • [6] BIOMAGNETIC MEASUREMENTS USING SQUIDS
    ANNINOS, PA
    ANOGIANAKIS, G
    LEHNERTZ, K
    PANTEV, C
    HOKE, M
    INTERNATIONAL JOURNAL OF NEUROSCIENCE, 1987, 37 (3-4) : 149 - 168
  • [7] Cancellation technique of external noise inside a magnetically shielded room used for biomagnetic measurements
    Kandori, A
    Miyashita, T
    Tsukada, K
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2000, 71 (05): : 2184 - 2190
  • [8] SENSITIVITY LIMITS IN BIOMAGNETIC MEASUREMENTS
    KATILA, T
    LEINIO, M
    MONTONEN, J
    NENONEN, J
    ACTA OTO-LARYNGOLOGICA, 1991, : 36 - 42
  • [9] MAINS NOISE SUPPRESSION FOR SQUID BIOMAGNETIC MEASUREMENTS USING A PHASE-SENSITIVE DETECTOR TECHNIQUE
    PRANCE, RJ
    RUSHBY, H
    POTTS, MW
    MOORE, M
    CLARK, TD
    CRYOGENICS, 1981, 21 (04) : 238 - 240
  • [10] Low-noise biomagnetic measurements with a multichannel dc-SQUID system at 77 K
    Barthelmess, HJ
    Halverscheid, M
    Schiefenhövel, B
    Heim, E
    Schilling, M
    Zimmermann, R
    IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 2001, 11 (01) : 657 - 660