Tonometric devices for central aortic systolic pressure measurements in patients with type 1 diabetes: comparison of the BPro and SphygmoCor devices

被引:9
|
作者
Theilade, Simone [1 ]
Hansen, Tine W. [1 ]
Joergensen, Christel [1 ]
Lajer, Maria [1 ]
Rossing, Peter [1 ,2 ,3 ]
机构
[1] Steno Diabet Ctr, DK-2820 Gentofte, Denmark
[2] Univ Copenhagen, Fac Hlth & Med Sci, Copenhagen, Denmark
[3] Aarhus Univ, Fac Hlth Sci, Aarhus, Denmark
关键词
central aortic systolic pressure; diabetes; tonometry; BLOOD-PRESSURE; EUROPEAN-SOCIETY; WAVE-FORM; ARTERIAL STIFFNESS; VALIDATION; DISEASE;
D O I
10.1097/MBP.0b013e328360fb19
中图分类号
R6 [外科学];
学科分类号
1002 ; 100210 ;
摘要
Objective Central blood pressure may be a better risk marker than brachial blood pressure and can be measured noninvasively by tonometric devices. We investigate whether tonometric measurements are feasible in patients with diabetes and whether the degree of albuminuria or increased arterial stiffness affects measurements. Patients and methods In 676 patients with type 1 diabetes, comparison of central aortic systolic pressure (CASP) measurements by the BPro and the SphygmoCor devices were made. The BPro device can obtain both office and 24-h measurements, whereas the SphygmoCor device is an accepted device for CASP measurements. Results Measurements of CASP with both BPro and SphygmoCor were available in 598 (88.5%) patients (mean age 54 years; mean diabetes duration 33 years; 45.2% women), and mean +/- SD of CASP was 122 +/- 17 and 118 +/- 17 mmHg, respectively (P < 0.001). Linear and intraclass correlation coefficients between CASP estimated from BPro and SphygmoCor were r equal to 0.96 and 0.95 (P< 0.001 for both). The mean +/- SD difference between devices was 3.6 +/- 4.8mmHg (P < 0.001). Analyses according to the level of albuminuria or degree of arterial stiffness were confirmatory. Conclusion In patients with type 1 diabetes, tonometric measurements of CASP with BPro and SphygmoCor showed strong correlations, although values differed by similar to 4mmHg between devices. Level of CASP, arterial stiffness, and degree of albuminuria did not interfere with the agreement between devices. In addition, the BPro device can obtain 24-h measurements and may thus be useful to assess the diurnal patterns of CASP. (c) 2013 Wolters Kluwer Health vertical bar Lippincott Williams & Wilkins.
引用
收藏
页码:156 / 160
页数:5
相关论文
共 25 条
  • [1] Assessment of Central Blood Pressure in Patients With Type 2 Diabetes: A Comparison Between Sphygmocor and Invasively Measured Values
    Laugesen, Esben
    Rossen, Niklas B.
    Peters, Christian D.
    Maeng, Michael
    Ebbehoj, Eva
    Knudsen, Soren T.
    Hansen, Klavs W.
    Botker, Hans E.
    Poulsen, Per L.
    AMERICAN JOURNAL OF HYPERTENSION, 2014, 27 (02) : 169 - 176
  • [2] Non-invasive estimates of central systolic blood pressure: Comparison of the Centron cBP301 and SphygmoCor devices
    Ann-Marie Mekhail
    Lisa M. Day
    Anna K. Goodhart
    Ian B. Wilkinson
    Carmel M. McEniery
    Artery Research, 2012, 6 (3) : 109 - 113
  • [3] Non-invasive estimates of central systolic blood pressure: Comparison of the Centron cBP301 and SphygmoCor devices
    Mekhail, Ann-Marie
    Day, Lisa M.
    Goodhart, Anna K.
    Wilkinson, Ian B.
    McEniery, Carmel M.
    ARTERY RESEARCH, 2012, 6 (03) : 109 - 113
  • [4] Estimates of arterial stiffness and central blood pressure in patients with type 2 diabetes: A comparison of SphygmoCor and Arteriograph
    Krogager, Christoffer
    Rossen, Niklas B.
    Hansen, Klavs W.
    Knudsen, Soren T.
    Peters, Christian D.
    Botker, Hans Erik
    Poulsen, Per L.
    Laugesen, Esben
    ARTERY RESEARCH, 2016, 16 : 18 - 24
  • [5] Discrepancy Between Tonometric Ambulatory and Cuff-Based Office Blood Pressure Measurements in Patients With Type 1 Diabetes
    Theilade, Simone
    Lajer, Maria
    Joergensen, Christel
    Persson, Frederik
    Rossing, Peter
    JOURNAL OF CLINICAL HYPERTENSION, 2012, 14 (10) : 686 - 693
  • [6] Calibration of noninvasive central blood pressure devices and negative aortic-to-brachial systolic pressure amplification
    Papaioannou, Theodore G.
    Vavuranakis, Manolis
    Tousoulis, Dimitrios
    KIDNEY INTERNATIONAL, 2017, 91 (01) : 253 - 254
  • [7] Influence of noninvasive central blood pressure devices for afterload monitoring with aortic velocity-pressure Loop in anesthetized patients
    Leve, Charlotte
    Hong, Alex
    Millasseau, Sandrine
    Joachim, Jona
    Touchard, Cyril
    Mateo, Joaquim
    Serrano, Jose
    Mebazaa, Alexandre
    Gayat, Etienne
    Vallee, Fabrice
    BLOOD PRESSURE MONITORING, 2020, 25 (04) : 184 - 194
  • [8] 24-hour central aortic systolic pressure and 24-hour central pulse pressure are related to diabetic complications in type 1 diabetes - a cross-sectional study
    Theilade, Simone
    Lajer, Maria
    Hansen, Tine Willum
    Joergensen, Christel
    Persson, Frederik
    Andresdottir, Gudbjorg
    Reinhard, Henrik
    Nielsen, Stine Elkjaer
    Lacy, Peter
    Williams, Bryan
    Rossing, Peter
    CARDIOVASCULAR DIABETOLOGY, 2013, 12
  • [9] Comparison of Two Noninvasive Devices for Measurement of Central Systolic Blood Pressure With Invasive Measurement During Cardiac Catheterization
    Ott, Christian
    Haetinger, Siegberto
    Schneider, Markus P.
    Pauschinger, Matthias
    Schmieder, Roland E.
    JOURNAL OF CLINICAL HYPERTENSION, 2012, 14 (09) : 575 - 579
  • [10] Central Systolic Blood Pressure Is Associated With Early Vascular Damage in Children and Adolescents With Type 1 Diabetes
    Tagetti, Angela
    Piona, Claudia A.
    Marcon, Denise
    Giontella, Alice
    Branz, Lorella
    Bortolotti, Stefano
    Morandi, Anita
    Maffeis, Claudio
    Fava, Cristiano
    FRONTIERS IN CARDIOVASCULAR MEDICINE, 2021, 8