Influence of the implant abutment types and the dynamic loading on initial screw loosening

被引:16
作者
Kim, Eun-Sook [1 ]
Shin, Soo-Yeon [1 ]
机构
[1] Dankook Univ, Coll Dent, Dept Prosthodont, Cheonan, South Korea
关键词
Computer-aided design; Dental implantation; Prosthesis failure; Removal torque; Screw loosening; INTERNAL CONICAL INTERFACE; 3-YEAR FOLLOW-UP; DENTAL IMPLANTS; FRAMEWORKS; SYSTEMS; FIT; CONNECTION; STRENGTH; DESIGN;
D O I
10.4047/jap.2013.5.1.21
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
PURPOSE. This study examined the effects of the abutment types and dynamic loading on the stability of implant prostheses with three types of implant abutments prepared using different fabrication methods by measuring removal torque both before and after dynamic loading. MATERIALS AND METHODS. Three groups of abutments were produced using different types of fabrication methods; stock abutment, gold cast abutment, and CAD/CAM custom abutment. A customized jig was fabricated to apply the load at 300 to the long axis. The implant fixtures were fixed to the jig, and connected to the abutments with a 30 Ncm tightening torque. A sine curved dynamic load was applied for 10 cycles between 25 and 250 N at 14 Hz. Removal torque before loading and after loading were evaluated. The SPSS was used for statistical analysis of the results. A Kruskal-Wallis test was performed to compare screw loosening between the abutment systems. A Wilcoxon signed-rank test was performed to compare screw loosening between before and after loading in each group (alpha=0.05). RESULTS. Removal torque value before loading and after loading was the highest in stock abutment, which was then followed by gold cast abutment and CAD/CAM custom abutment, but there were no significant differences. CONCLUSION. The abutment types did not have a significant influence on short term screw loosening. On the other hand, after 105 cycles dynamic loading, CAD/CAM custom abutment affected the initial screw loosening, but stock abutment and gold cast abutment did not. [J Adv Prosthodont 2013;5:21-8]
引用
收藏
页码:21 / 28
页数:8
相关论文
共 39 条
  • [1] Abduo J, 2011, INT J PROSTHODONT, V24, P207
  • [2] Al-Fadda SA, 2007, INT J PROSTHODONT, V20, P125
  • [3] COMPARATIVE-STUDY OF ANTIROTATIONAL SINGLE TOOTH ABUTMENTS
    BALFOUR, A
    OBRIEN, GR
    [J]. JOURNAL OF PROSTHETIC DENTISTRY, 1995, 73 (01) : 36 - 43
  • [4] Bates J F, 1976, J Oral Rehabil, V3, P57, DOI 10.1111/j.1365-2842.1976.tb00929.x
  • [5] Binon P P, 1995, Int J Prosthodont, V8, P162
  • [6] Binon PP, 1996, INT J PROSTHODONT, V9, P511
  • [7] TIGHTENING CHARACTERISTICS FOR SCREWED JOINTS IN OSSEOINTEGRATED DENTAL IMPLANTS
    BURGUETE, RL
    JOHNS, RB
    KING, T
    PATTERSON, EA
    [J]. JOURNAL OF PROSTHETIC DENTISTRY, 1994, 71 (06) : 592 - 599
  • [8] Examination of the implant-abutment interface after fatigue testing
    Cibirka, RM
    Nelson, SK
    Lang, BR
    Rueggeberg, FA
    [J]. JOURNAL OF PROSTHETIC DENTISTRY, 2001, 85 (03) : 268 - 275
  • [9] LIMITS OF HUMAN BITE STRENGTH
    GIBBS, CH
    MAHAN, PE
    MAUDERLI, A
    LUNDEEN, HC
    WALSH, EK
    [J]. JOURNAL OF PROSTHETIC DENTISTRY, 1986, 56 (02) : 226 - 229
  • [10] Haack J E, 1995, Int J Oral Maxillofac Implants, V10, P529