Adverse local tissue reactions in metal-on-polyethylene total hip arthroplasty due to trunnion corrosion THE RISK OF MISDIAGNOSIS

被引:104
作者
Whitehouse, M. R. [1 ]
Endo, M. [1 ]
Zachara, S. [2 ]
Nielsen, T. O. [3 ]
Greidanus, N. V. [1 ]
Masri, B. A. [1 ]
Garbuz, D. S. [1 ]
Duncan, C. P. [1 ]
机构
[1] Vancouver Gen Hosp, Vancouver, BC, Canada
[2] Vancouver Gen Hosp, Anat Pathol, Vancouver, BC V5Z 1M9, Canada
[3] Vancouver Gen Hosp, Pathol & Lab Med, Anat Pathol, Vancouver, BC V5Z 1M9, Canada
关键词
FRETTING CORROSION; PSEUDOTUMOR; REVISION; COMPONENTS; PRODUCTS; BEARINGS; CREVICE;
D O I
10.1302/0301-620X.97B8.34682
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Adverse reaction to wear and corrosion debris is a cause for concern in total hip arthroplasty (THA). Modular junctions are a potential source of such wear products and are associated with secondary pseudotumour formation. We present a consecutive series of 17 patients treated at our unit for this complication following metal-on-highly cross-linked polyethylene (MoP) THA. We emphasise the risk of misdiagnosis as infection, and present the aggregate laboratory results and pathological findings in this series. The clinical presentation was pain, swelling or instability. Solid, cystic and mixed soft-tissue lesions were noted on imaging and confirmed intra-operatively. Corrosion at the head-neck junction was noted in all cases. No bacteria were isolated on multiple pre- and intra-operative samples yet the mean erythrocyte sedimentation rate was 49 (9 to 100) and C-reactive protein 32 (0.6 to 106) and stromal polymorphonuclear cell counts were noted in nine cases. Adverse soft-tissue reactions can occur in MoP THA owing to corrosion products released from the head-neck junction. The diagnosis should be carefully considered when investigating pain after THA. This may avoid the misdiagnosis of periprosthetic infection with an unidentified organism and mitigate the unnecessary management of these cases with complete single-or two-stage exchange.
引用
收藏
页码:1024 / 1030
页数:7
相关论文
共 39 条
  • [1] [Anonymous], DIAGNOSIS PERIPROSTH
  • [2] [Anonymous], 2009, IMATRI MED ANTEVERSI
  • [3] High failure rates with a large-diameter hybrid metal-on-metal total hip replacement CLINICAL, RADIOLOGICAL AND RETRIEVAL ANALYSIS
    Bolland, B. J. R. F.
    Culliford, D. J.
    Langton, D. J.
    Millington, J. P. S.
    Arden, N. K.
    Latham, J. M.
    [J]. JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2011, 93B (05): : 608 - 615
  • [4] FRETTING CORROSION ACCELERATES CREVICE CORROSION OF MODULAR HIP TAPERS
    BROWN, SA
    FLEMMING, CAC
    KAWALEC, JS
    PLACKO, HE
    VASSAUX, C
    MERRITT, K
    PAYER, JH
    KRAAY, MJ
    [J]. JOURNAL OF APPLIED BIOMATERIALS, 1995, 6 (01) : 19 - 26
  • [5] Failed Metal-on-Metal Hip Arthroplasties: A Spectrum of Clinical Presentations and Operative Findings
    Browne, James A.
    Bechtold, C. Dustin
    Berry, Daniel J.
    Hanssen, Arlen D.
    Lewallen, David G.
    [J]. CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2010, 468 (09) : 2313 - 2320
  • [6] In vitro comparison of frictional torque and torsional resistance of aged conventional gamma-in-nitrogen sterilized polyethylene versus aged highly crosslinked polyethylene articulating against head sizes larger than 32 mm
    Burroughs, Brian R.
    Muratoglu, Orhun K.
    Bragdon, Charles R.
    Wannomae, Keith K.
    Christensen, Steven
    Lozynsky, Andrew J.
    Harris, William H.
    [J]. ACTA ORTHOPAEDICA, 2006, 77 (05) : 710 - 718
  • [7] Histological Features of Pseudotumor-like Tissues From Metal-on-Metal Hips
    Campbell, Pat
    Ebramzadeh, Edward
    Nelson, Scott
    Takamura, Karren
    De Smet, Koen
    Amstutz, Harlan C.
    [J]. CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2010, 468 (09) : 2321 - 2327
  • [8] A NEW METHOD OF CLASSIFYING PROGNOSTIC CO-MORBIDITY IN LONGITUDINAL-STUDIES - DEVELOPMENT AND VALIDATION
    CHARLSON, ME
    POMPEI, P
    ALES, KL
    MACKENZIE, CR
    [J]. JOURNAL OF CHRONIC DISEASES, 1987, 40 (05): : 373 - 383
  • [9] COLLIER JP, 1991, CLIN ORTHOP RELAT R, P305
  • [10] CORROSION BETWEEN THE COMPONENTS OF MODULAR FEMORAL HIP PROSTHESES
    COLLIER, JP
    SURPRENANT, VA
    JENSEN, RE
    MAYOR, MB
    SURPRENANT, HP
    [J]. JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1992, 74 (04): : 511 - 517