The Production of a Uniform Magnetic Field Using a System of Axial Coils for Calibrating Magnetometers

被引:18
作者
Baranova, V. E. [1 ]
Baranov, P. F. [1 ]
Muravyov, S. V. [1 ]
Uchaikin, S. V. [2 ]
机构
[1] Natl Res Tomsk Polytech Univ, Tomsk, Russia
[2] D Wave Syst Inc, Burnaby, BC, Canada
关键词
uniform magnetic field; axial coils; Helmholtz coils; magnetometer; calibration;
D O I
10.1007/s11018-015-0752-9
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
The construction of an axial 8-coil system based on Helmholtz coils is described, which enables a magnetic field with a nonuniformity of not more than 0.1% to be produced at a distance of half the radius from the center of the coil system. The results obtained are confirmed by experimental investigations.
引用
收藏
页码:550 / 555
页数:6
相关论文
共 14 条
  • [1] [Anonymous], 803091 GOST GSI
  • [2] Baranov P. F., 2012, IZV TOMSK POLITEKH U, V320, P89
  • [3] Barker J.R., 1949, REV SCI INSTRUM, V26, P213
  • [4] COIL-SYSTEM DESIGN FOR PRODUCTION OF UNIFORM MAGNETIC-FIELDS
    CARTER, RG
    [J]. PROCEEDINGS OF THE INSTITUTION OF ELECTRICAL ENGINEERS-LONDON, 1976, 123 (11): : 1279 - 1283
  • [5] COILS FOR GENERATING UNIFORM-FIELDS IN A CYLINDRICAL FERROMAGNETIC SHIELD
    HOSOYA, M
    GOTO, E
    [J]. REVIEW OF SCIENTIFIC INSTRUMENTS, 1991, 62 (10) : 2472 - 2475
  • [6] Kalantarov P. L., 1986, Calculation of Inductances: Handbook
  • [7] Square loop coil system for balancing and calibration of second-order SQUID gradiometers
    Lee, Soon Gul
    Kang, Chan Seok
    Chang, Jung Won
    [J]. IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 2007, 17 (02) : 3769 - 3772
  • [8] Combinations of circular currents for producing uniform magnetic fields
    McKeehan, LW
    [J]. REVIEW OF SCIENTIFIC INSTRUMENTS, 1936, 7 (03) : 150 - 153
  • [9] AC magnetic flux density standards in the low frequency range
    Park, Po Gyu
    Kim, Y. G.
    Kalabin, V. N.
    Shifrin, V. Ya.
    [J]. 2008 CONFERENCE ON PRECISION ELECTROMAGNETIC MEASUREMENTS DIGEST, 2008, : 456 - +
  • [10] Park PG, 2010, 2010 CONFERENCE ON PRECISION ELECTROMAGNETIC MEASUREMENTS CPEM, P312, DOI 10.1109/CPEM.2010.5543554