Fracture of nickel-titanium superelastic alloy in sodium hypochlorite solution

被引:25
作者
Yokoyama, K
Kaneko, K
Yabuta, E
Asaoka, K
Sakai, J
机构
[1] Univ Tokushima, Sch Dent, Dept Dent Engn, Tokushima 7708504, Japan
[2] Univ Tokushima, Sch Dent, Dept Orthodont, Tokushima 7708504, Japan
[3] Waseda Univ, Dept Mat Sci & Engn, Shinjuku Ku, Tokyo 1698555, Japan
来源
MATERIALS SCIENCE AND ENGINEERING A-STRUCTURAL MATERIALS PROPERTIES MICROSTRUCTURE AND PROCESSING | 2004年 / 369卷 / 1-2期
关键词
Ni-Ti; superelastic alloy; fracture; corrosion; sodium hypochlorite;
D O I
10.1016/j.msea.2003.10.320
中图分类号
TB3 [工程材料学];
学科分类号
0805 ; 080502 ;
摘要
Fracture of the Ni-Ti superelastic alloy for endodontic instruments such as files was investigated with a sustained tensile-loading test in sodium hypochlorite (NaOCl) solution of various concentrations. It was found that the time to fracture was reduced when the applied stress exceeded the critical stress for martensite transformation. When the applied stress was higher than the critical stress, the 0.3 mm diameter wires of the Ni-Ti superelastic alloy sometimes fractured within 60 min. From the results of observations of the fracture surface using a scanning electron microscope, it was revealed that the fracture of the Ni-Ti superelastic alloy is significantly influenced by corrosion when the applied stress was higher than the critical stress for martensite transformation. The results of the present study suggest that one of the causes of the fracture of Ni-Ti files during clinical use is corrosion under the applied stress above the critical stress for martensite transformation in NaOCI solution. (C) 2003 Elsevier B.V. All rights reserved.
引用
收藏
页码:43 / 48
页数:6
相关论文
共 23 条
[11]   Accelerated failure in high strength steel by alternating hydrogen-charging potential [J].
Nagumo, M ;
Uyama, H ;
Yoshizawa, M .
SCRIPTA MATERIALIA, 2001, 44 (06) :947-952
[12]  
Nakasa K, 1999, TETSU TO HAGANE, V85, P479
[13]  
Oshida Y, 1991, ZAIRYO TO KANKYO, V40, P834, DOI DOI 10.3323/JCORR1991.40.834
[14]   Cyclic fatigue testing of nickel-titanium endodontic instruments [J].
Pruett, JP ;
Clement, DJ ;
Carnes, DL .
JOURNAL OF ENDODONTICS, 1997, 23 (02) :77-85
[15]  
Rondelli G, 2000, J BIOMED MATER RES, V51, P47, DOI 10.1002/(SICI)1097-4636(200007)51:1<47::AID-JBM7>3.3.CO
[16]  
2-G
[17]   THE CORROSION BEHAVIOR OF NICKEL TITANIUM SHAPE MEMORY ALLOYS [J].
RONDELLI, G ;
VICENTINI, B ;
GIGADA, A .
CORROSION SCIENCE, 1990, 30 (8-9) :805-812
[18]   Corrosion resistance tests on NiTi shape memory alloy [J].
Rondelli, G .
BIOMATERIALS, 1996, 17 (20) :2003-2008
[19]   Torsional properties of stainless steel and nickel-titanium endodontic files [J].
Rowan, MB ;
Nicholls, JI ;
Steiner, J .
JOURNAL OF ENDODONTICS, 1996, 22 (07) :341-345
[20]   ANODIC OXIDE-FILMS OF NICKEL IN ALKALINE ELECTROLYTE [J].
VISSCHER, W ;
BARENDRECHT, E .
SURFACE SCIENCE, 1983, 135 (1-3) :436-452