Routine left atrial catheterization for the post-operative management of cardiac surgical patients: is the risk justified?

被引:16
作者
Santini, F [1 ]
Gatti, G [1 ]
Borghetti, V [1 ]
Oppido, G [1 ]
Mazzucco, A [1 ]
机构
[1] Univ Verona, Sch Med, Div Cardiac Surg, OCM Borgo Trento, I-37126 Verona, Italy
关键词
left atrial line; pressures monitoring; pulmonary vascular constriction; open heart operations; postoperative care;
D O I
10.1016/S1010-7940(99)00161-X
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Objective. To assess the risk/benefit ratio, including cost, associated with routine left atrial catheterization for the post-operative management of patients after cardiac surgery. Methods: From November 1991 to June 1998, out of 6187 open heart procedures performed at our institution, 5815 patients (94%) receive a left atrial monitoring line inserted invasively by a unique and reproducible technique. Catheters were removed on the first or second postoperative day before chest tube removal. A subgroup of 385 patients (7%) were switched to a Swan-Ganz catheter postoperatively. Results: A total of 14 patients suffered a complication related to the left monitoring line (0.24%). Bleeding necessitating transfusion occurred in 10 patients (0.17%), seven of whom required surgical reexploration (0.12%). Catheter retention occurred in four cases (0.07%), requiring a mediastinal reexploration in one and a minimally invasive procedures in three, for removal. No other complications related to the presence of a left atrial line emerged in this series. Apart from providing crucial hemodynamic information on a routine basis, in nine selected patients (0.15%) in a low cardiac output state with increased pulmonary vascular resistance and right ventricular failure, the left atrial line was used as a preferential route for catecholamine infusion, with significant hemodynamic improvement. Conclusions: Complications of left atrial monitoring catheters in cardiac surgery do occur but at a very low and acceptable rate. No mortality was correlated to their use in our series. Complication rate can be further lowered by a meticulous management of the device. The wealth of information and therapeutic options offered by this line appears to outweigh the associated risk. (C) 1999 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:218 / 221
页数:4
相关论文
共 29 条
[1]  
AKL BF, 1984, J THORAC CARDIOV SUR, V88, P1033
[2]  
ARNOLD J V, 1990, Journal of Audiovisual Media in Medicine, V13, P87, DOI 10.3109/17453059009055108
[3]   RETRIEVAL OF BROKEN LEFT-ATRIAL CATHETER USING A RIGID BRONCHOSCOPE [J].
BARNARD, SP ;
MAW, A ;
KVERNABA, K ;
KULATILAKE, ENP .
THORACIC AND CARDIOVASCULAR SURGEON, 1993, 41 (01) :71-72
[4]   Swan Ganz catheter - all in a knot [J].
Colbert, S ;
OHanlon, DM ;
Quill, DS ;
Keane, P .
EUROPEAN JOURNAL OF ANAESTHESIOLOGY, 1997, 14 (05) :518-520
[5]   SYSTEMIC GASEOUS MICROEMBOLI DURING LEFT ATRIAL CATHETERIZATION - A COMMON OCCURRENCE [J].
FEERICK, AE ;
CHURCH, JA ;
ZWISCHENBERGER, J ;
CONTI, V ;
JOHNSTON, WE .
JOURNAL OF CARDIOTHORACIC AND VASCULAR ANESTHESIA, 1995, 9 (04) :395-398
[6]   CATHETER-INDUCED PULMONARY-ARTERY PERFORATION - PATHOLOGICAL AND PATHOGENIC FEATURES [J].
FRASER, RS .
HUMAN PATHOLOGY, 1987, 18 (12) :1246-1251
[7]   TRANS-SEPTAL LEFT ATRIAL CATHETERIZATION [J].
FUKUTOME, T ;
KOHJIRO, M ;
SESE, A ;
UENO, Y .
ANAESTHESIA, 1993, 48 (05) :407-408
[8]   TRANS-THORACIC INTRACARDIAC MONITORING LINES IN PEDIATRIC SURGICAL PATIENTS - A 10-YEAR EXPERIENCE [J].
GOLD, JP ;
JONAS, RA ;
LANG, P ;
ELIXSON, EM ;
MAYER, JE ;
CASTANEDA, AR .
ANNALS OF THORACIC SURGERY, 1986, 42 (02) :185-191
[9]  
Haider W, 1993, J Cardiothorac Vasc Anesth, V7, P684, DOI 10.1016/1053-0770(93)90053-N
[10]  
Kulkarni R, 1998, CATHETER CARDIO DIAG, V44, P317, DOI 10.1002/(SICI)1097-0304(199807)44:3<317::AID-CCD15>3.0.CO