Is greater than 145° of deep knee flexion under weight-bearing conditions safe after total knee arthroplasty? A FLUOROSCOPIC ANALYSIS OF JAPANESE-STYLE DEEP KNEE FLEXION

被引:17
|
作者
Niki, Y. [1 ]
Takeda, Y. [1 ]
Udagawa, K. [1 ]
Enomoto, H. [1 ]
Toyama, Y. [1 ]
Suda, Y. [1 ]
机构
[1] Keio Univ, Sch Med, Dept Orthopaed Surg, Shinjuku Ku, Tokyo 1608582, Japan
关键词
POLYETHYLENE TIBIAL POST; IN-VIVO; KINEMATICS; FRACTURE;
D O I
10.1302/0301-620X.95B6.30757
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
We investigated the characteristics of patients who achieved Japanese-style deep flexion (seiza-sitting) after total knee replacement (TKR) and measured three-dimensional positioning and the contact positions of the femoral and tibial components. Seiza-sitting was achieved after surgery by 23 patients (29 knees) of a series of 463 TKRs in 341 patients. Pre-operatively most of these patients were capable of seiza-sitting, had a lower body mass index and a favourable attitude towards the Japanese lifestyle (27 of 29 knees). According to two-/three-dimensional image registration analysis in the seiza-sitting position, flexion, varus and internal rotation angles of the tibial component relative to the femoral component had means of 148 (sip 8.0), 1.9 degrees (6D 3.2) and 13.4 degrees (SD 5.9), respectively. Femoral surface contact positions tended to be close to the posterior edge of the tibial polyethylene insert, particularly in the lateral compartment, but only 8.3% (two of 24) of knees showed femoral subluxation over the posterior edge. The mean contact positions of the femoral cam on the tibial post were located 7.8 mm (SD 1.5) proximal to the lowest point of the polyethylene surface and 5.5 mm (6D 0.9) medial to the centre of the post, indicating that the post-cam contact position translated medially during seiza-sitting, but not proximally. Collectively, the seiza-sitting position seems safe against component dislocation, but the risks of posterior edge loading and breakage of the tibial polyethylene post remain.
引用
收藏
页码:782 / 787
页数:6
相关论文
共 41 条
  • [1] Factors affecting the achievement of Japanese-style deep knee flexion after total knee arthroplasty using posterior-stabilized prosthesis with high-flex knee design
    Niki, Yasuo
    Takeda, Yuki
    Harato, Kengo
    Suda, Yasunori
    JOURNAL OF ORTHOPAEDIC SCIENCE, 2015, 20 (06) : 1012 - 1018
  • [2] Fluoroscopic analysis of knee arthroplasty kinematics during deep flexion kneeling
    Kanekasu, K
    Banks, SA
    Honjo, S
    Nakata, O
    Kato, H
    JOURNAL OF ARTHROPLASTY, 2004, 19 (08) : 998 - 1003
  • [3] Evaluation of the Contact Area in Total Knee Arthroplasty Designed for Deep Knee Flexion
    Triwardono, Joko
    Supriadi, Sugeng
    Whulanza, Yudan
    Saragih, Agung Shamsuddin
    Novalianita, Deva Ariana
    Utomo, Muhammad Satrio
    Kartika, Ika
    INTERNATIONAL JOURNAL OF TECHNOLOGY, 2021, 12 (06) : 1312 - 1322
  • [4] Kinematics of bicruciate and posterior stabilized total knee arthroplasty during deep knee flexion and stair climbing
    Ishibashi, Teruya
    Tomita, Tetsuya
    Yamazaki, Takaharu
    Tsuji, Shigeyoshi
    Yoshikawa, Hideki
    Sugamoto, Kazuomi
    JOURNAL OF ORTHOPAEDIC RESEARCH, 2021, 39 (06) : 1262 - 1270
  • [5] In Vivo Kinematic Analysis of Cruciate-Retaining Total Knee Arthroplasty During Weight-Bearing and Non-Weight-Bearing Deep Knee Bending
    Horiuchi, Hiroshi
    Akizuki, Shaw
    Tomita, Tetsuya
    Sugamoto, Kazuomi
    Yamazaki, Takaharu
    Shimizu, Norimasa
    JOURNAL OF ARTHROPLASTY, 2012, 27 (06) : 1196 - 1202
  • [6] Patellofemoral kinematics during deep knee flexion after total knee replacement: a computational simulation
    Huang, Chang-Hung
    Hsu, Lin-I
    Lin, Kun-Jhih
    Chang, Ting-Kuo
    Cheng, Cheng-Kung
    Lu, Yung-Chang
    Chen, Chen-Sheng
    Huang, Chun-Hsiung
    KNEE SURGERY SPORTS TRAUMATOLOGY ARTHROSCOPY, 2014, 22 (12) : 3047 - 3053
  • [7] Patellofemoral kinematics during deep knee flexion after total knee replacement: a computational simulation
    Chang-Hung Huang
    Lin-I Hsu
    Kun-Jhih Lin
    Ting-Kuo Chang
    Cheng-Kung Cheng
    Yung-Chang Lu
    Chen-Sheng Chen
    Chun-Hsiung Huang
    Knee Surgery, Sports Traumatology, Arthroscopy, 2014, 22 : 3047 - 3053
  • [8] Effect of weight-bearing in bicruciate-retaining total knee arthroplasty during high-flexion activities
    Kono, Kenichi
    Inui, Hiroshi
    Tomita, Tetsuya
    Yamazaki, Takaharu
    Taketomi, Shuji
    Yamagami, Ryota
    Kawaguchi, Kohei
    Kage, Tomofumi
    Arakawa, Takahiro
    Tanaka, Sakae
    CLINICAL BIOMECHANICS, 2022, 92
  • [9] Weight-bearing knee flexion angle better correlates with patient-reported outcome measures than non-weight-bearing condition in total knee arthroplasty: a three-dimensional analysis study
    Kage, Tomofumi
    Inui, Hiroshi
    Tomita, Tetsuya
    Yamazaki, Takaharu
    Taketomi, Shuji
    Yamagami, Ryota
    Kono, Kenichi
    Kawaguchi, Kohei
    Sameshima, Shin
    Tanaka, Sakae
    BMC MUSCULOSKELETAL DISORDERS, 2021, 22 (01)
  • [10] In Vivo Knee Kinematics During Stair and Deep Flexion Activities in Patients With Bicruciate Substituting Total Knee Arthroplasty
    Kuroyanagi, Yuji
    Mu, Shang
    Hamai, Satoshi
    Robb, William J.
    Banks, Scott A.
    JOURNAL OF ARTHROPLASTY, 2012, 27 (01) : 122 - 128