Improvements in the corrosion resistance and biocompatibility of biomedical Ti-6Al-7Nb alloy using an electrochemical anodization treatment

被引:52
作者
Huang, Her-Hsiung [1 ,2 ,3 ]
Wu, Chia-Ping [4 ]
Sun, Ying-Sui [4 ]
Lee, Tzu-Hsin [5 ,6 ]
机构
[1] Natl Yang Ming Univ, Dept Dent, Taipei 112, Taiwan
[2] Taipei City Hosp, Dept Dent, Taipei 115, Taiwan
[3] Taipei Vet Gen Hosp, Dept Stomatol, Taipei 112, Taiwan
[4] Natl Yang Ming Univ, Dept Oral Biol, Taipei 112, Taiwan
[5] Chung Shan Med Univ, Sch Dent, Taichung 402, Taiwan
[6] Chung Shan Med Univ Hosp, Dept Dent, Taichung 402, Taiwan
关键词
Ti-6Al-7Nb alloy; Anodization; Nanoporous; Corrosion; Biocompatibility; TITANIUM SURFACE; TIO2; NANOTUBES; LAYER; IMPLANTS; VANADIUM; METAL;
D O I
10.1016/j.tsf.2012.08.063
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
The biocompatibility of an implant material is determined by its surface characteristics. This study investigated the application of an electrochemical anodization surface treatment to improve both the corrosion resistance and biocompatibility of Ti-6Al-7Nb alloy for implant applications. The electrochemical anodization treatment produced an Al-free oxide layer with nanoscale porosity on the Ti-6Al-7Nb alloy surface. The surface topography and microstructure of Ti-6Al-7Nb alloy were analyzed. The corrosion resistance was investigated using potentiodynamic polarization curve measurements in simulated blood plasma (SBP). The adhesion and proliferation of human bone marrow mesenchymal stem cells to test specimens were evaluated using various biological analysis techniques. The results showed that the presence of a nanoporous oxide layer on the anodized Ti-6Al-7Nb alloy increased the corrosion resistance (i.e., increased the corrosion potential and decreased both the corrosion rate and the passive current) in SBP compared with the untreated Ti-6Al-7Nb alloy. Changes in the nanotopography also improved the cell adhesion and proliferation on the anodized Ti-6Al-7Nb alloy. We conclude that a fast and simple electrochemical anodization surface treatment improves the corrosion resistance and biocompatibility of Ti-6Al-7Nb alloy for biomedical implant applications. (c) 2012 Elsevier B.V. All rights reserved.
引用
收藏
页码:157 / 162
页数:6
相关论文
共 29 条
[1]  
[Anonymous], 1996, PRINCIPALS PREVENTIO
[2]   Activation of integrin function by nanopatterned adhesive interfaces [J].
Arnold, M ;
Cavalcanti-Adam, EA ;
Glass, R ;
Blümmel, J ;
Eck, W ;
Kantlehner, M ;
Kessler, H ;
Spatz, JP .
CHEMPHYSCHEM, 2004, 5 (03) :383-388
[3]   TiO2 nanotubes:: Tailoring the geometry in H3PO4/HF electrolytes [J].
Bauer, Sebastian ;
Kleber, Sebastian ;
Schmuki, Patrik .
ELECTROCHEMISTRY COMMUNICATIONS, 2006, 8 (08) :1321-1325
[4]   In vitro toxic effects of metal compounds on kinetic traits and ultrastructure of rabbit spermatozoa [J].
Castellini, Cesare ;
Mourvaki, Evangelia ;
Sartini, Barbara ;
Cardinali, Raffaella ;
Moretti, Elena ;
Collodel, Giulia ;
Fortaner, Salvador ;
Sabbioni, Enrico ;
Renieri, Tommaso .
REPRODUCTIVE TOXICOLOGY, 2009, 27 (01) :46-54
[5]   Protection against vanadium-induced testicular toxicity by testosterone propionate in rats [J].
Chandra, Amar K. ;
Ghosh, Rituparna ;
Chatterjee, Aparajita ;
Sarkar, Mahitosh .
TOXICOLOGY MECHANISMS AND METHODS, 2010, 20 (06) :306-315
[6]   Fibroblast cell attachment and growth on nanoengineered sculptured thin films [J].
Demirel, M. C. ;
So, E. ;
Ritty, T. M. ;
Naidu, S. H. ;
Lakhtakia, A. .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH PART B-APPLIED BIOMATERIALS, 2007, 81B (01) :219-223
[7]   Biocompatibility of β-stabilizing elements of titanium alloys [J].
Eisenbarth, E ;
Velten, D ;
Müller, M ;
Thull, R ;
Breme, J .
BIOMATERIALS, 2004, 25 (26) :5705-5713
[8]   Osteoinduction of porous bioactive titanium metal [J].
Fujibayashi, S ;
Neo, M ;
Kim, HM ;
Kokubo, T ;
Nakamura, T .
BIOMATERIALS, 2004, 25 (03) :443-450
[9]   Ion implantation and annealing for an efficient N-doping of TiO2 nanotubes [J].
Ghicov, Andrei ;
Macak, Jan M. ;
Tsuchiya, Hiroaki ;
Kunze, Julia ;
Haeublein, Volker ;
Frey, Lothar ;
Schmuki, Patrik .
NANO LETTERS, 2006, 6 (05) :1080-1082
[10]   Cellular Ti-6Al-4V structures with interconnected macro porosity for bone implants fabricated by selective electron beam melting [J].
Heinl, Peter ;
Mueller, Lenka ;
Koerner, Carolin ;
Singer, Robert F. ;
Mueller, Frank A. .
ACTA BIOMATERIALIA, 2008, 4 (05) :1536-1544