Towards estimating fiducial localization error of point-based registration in image-guided neurosurgery

被引:3
|
作者
Zhi, Deng [1 ]
机构
[1] Fudan Univ, Zhongshan Hosp, Emergency Dept, Shanghai 200433, Peoples R China
关键词
Image-guided neurosurgery; fiducial localization error; point-based registration; TARGET REGISTRATION;
D O I
10.3233/BME-151388
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Fiducial Localization Error (FLE) is one of the major reasons of inaccuracy in point-based spatial registration of Image-Guided Neurosurgery System (IGNS), and minimizing FLE is the fundamental way to improve spatial registration accuracy. A reliable estimation of FLE is needed, as it cannot be measured directly in real application of IGNS. In this paper, we propose a method to estimate the FLE in a point-based registration of IGNS. Test fiducial point sets were generated in one coordinate system around the given fiducial point set by utilizing simple random sampling. Further, these points were registered to the fiducial point set in the other coordinate system. The average position of the test fiducial point sets with small FRE are calculated and its displacement from the given fiducial point set is the parameter used to estimate the FLE of each fiducial point. The correlation between the displacement and the FLE of each fiducial point is greater than 0.75 when nine or more fiducial points were utilized. This correlation gradually increases up to 0.9 with the increase of the number of fiducial points.
引用
收藏
页码:S943 / S949
页数:7
相关论文
共 50 条
  • [1] The Study of Fiducial Localization Error of Image in Point-based Registration
    Liu, Wenbo
    Ding, Hui
    Han, Hongyan
    Xue, Qinghua
    Sun, Zhaohui
    Wang, Guangzhi
    2009 ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-20, 2009, : 5088 - +
  • [2] Fiducial Optimization for Minimal Target Registration Error in Image-Guided Neurosurgery
    Shamir, Reuben R.
    Joskowicz, Leo
    Shoshan, Yigal
    IEEE TRANSACTIONS ON MEDICAL IMAGING, 2012, 31 (03) : 725 - 737
  • [3] Toward Estimating MRI-Ultrasound Registration Error in Image-Guided Neurosurgery
    Bierbrier, Joshua
    Eskandari, Mohammadreza
    Di Giovanni, Daniel A.
    Collins, D. Louis
    IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2023, 70 (09) : 999 - 1015
  • [4] Distribution of Fiducial Registration Error in Rigid-Body Point-Based Registration
    Moghari, Mehdi Hedjazi
    Abolmaesumi, Purang
    IEEE TRANSACTIONS ON MEDICAL IMAGING, 2009, 28 (11) : 1791 - 1801
  • [5] Understanding the Effect of Bias in Fiducial Localization Error on Point-Based Rigid-Body Registration
    Moghari, Mehdi Hedjazi
    Abolmaesumi, Purang
    IEEE TRANSACTIONS ON MEDICAL IMAGING, 2010, 29 (10) : 1730 - 1738
  • [6] Estimating Maximum Target Registration Error Under Uniform Restriction of Fiducial Localization Error in Image Guided System
    Qiu, Lei
    Zhang, Yu
    Xu, Lihui
    Niu, Xiaohui
    Zhang, Qing
    Zhang, Li
    IEEE TRANSACTIONS ON MEDICAL IMAGING, 2018, 37 (04) : 881 - 892
  • [7] Improving target registration accuracy in image-guided neurosurgery by optimizing the distribution of fiducial points
    Wang, Manning
    Song, Zhijian
    INTERNATIONAL JOURNAL OF MEDICAL ROBOTICS AND COMPUTER ASSISTED SURGERY, 2009, 5 (01): : 26 - 31
  • [8] A statistical model for point-based target registration error with anisotropic fiducial localizer error
    Wiles, Andrew D.
    Likholyot, Alexander
    Frantz, Donald D.
    Peters, Terry M.
    IEEE TRANSACTIONS ON MEDICAL IMAGING, 2008, 27 (03) : 378 - 390
  • [9] Guidelines for the placement of fiducial points in image-guided neurosurgery
    Wang, Manning
    Song, Zhijian
    INTERNATIONAL JOURNAL OF MEDICAL ROBOTICS AND COMPUTER ASSISTED SURGERY, 2010, 6 (02): : 142 - 149
  • [10] Distribution Templates of the Fiducial Points in Image-Guided Neurosurgery
    Wang Manning
    Song Zhijian
    NEUROSURGERY, 2010, 66 (03) : 143 - 150