A combined dose calculation and verification method for a small animal precision irradiator based on onboard imaging

被引:37
作者
Granton, Patrick V. [1 ]
Podesta, Mark [1 ]
Landry, Guillaume [1 ]
Nijsten, Sebastiaan [1 ]
Bootsma, Gregory [2 ,3 ]
Verhaegen, Frank [1 ,4 ]
机构
[1] Maastricht Univ Med Ctr, Dept Radiat Oncol MAASTRO, GROW, Sch Oncol & Dev Biol, NL-6201 BN Maastricht, Netherlands
[2] Univ Toronto, Dept Med Biophys, Toronto, ON M5G 2M9, Canada
[3] Princess Margaret Hosp, Radiat Med Program, Toronto, ON M5G 2M9, Canada
[4] McGill Univ, Dept Oncol, Montreal, PQ H3G 1A4, Canada
基金
加拿大自然科学与工程研究理事会;
关键词
portal dosimetry; small animal micro-IR; Monte Carlo; COMPUTED-TOMOGRAPHY; RADIOTHERAPY; BEAM; RADIATION; SCATTER; SYSTEM; PERFORMANCE; DOSIMETRY; QUALITY; EGSNRC;
D O I
10.1118/1.4725710
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: Novel small animal precision microirradiators (micro-IR) are becoming available for preclinical use and are often equipped with onboard imaging (OBI) devices. We investigated the use of OBI as a means to infer the accuracy of the delivered treatment plan. Methods: Monte Carlo modeling of the micro-IR including an elliptical Gaussian electron beam incident on the x-ray tube was used to score dose and to continue photon transport to the plane of the OBI device. A model of the OBI detector response was used to generate simulated onboard images. Experimental OBI was performed at 225 kVp, gain/offset and scatter-glare were corrected. Simulated and experimentally obtained onboard images of phantoms and a mouse specimen were compared for a range of photon beam sizes from 2.5 cm down to 0.1 cm. Results: Simulated OBI can be used in small animal radiotherapy to determine if a treatment plan was delivered according to the prescription within an uncertainty of 5% for beams as small as 4 mm in diameter. For collimated beams smaller than 4 mm, beam profile differences remain primarily in the penumbra region of the smallest beams, which may be tolerable for specific preclinical micro-IR investigations. Conclusions: Comparing simulated to acquired OBI during small animal treatment radiotherapy represents a useful treatment delivery tool. (C) 2012 American Association of Physicists in Medicine. [http://dx.doi.org/10.1118/1.4725710]
引用
收藏
页码:4155 / 4166
页数:12
相关论文
共 34 条
[1]   Efficiency improvements of x-ray simulations in EGSnrc user-codes using bremsstrahlung cross-section enhancement (BCSE) [J].
Ali, E. S. M. ;
Rogers, D. W. O. .
MEDICAL PHYSICS, 2007, 34 (06) :2143-2154
[2]   A comprehensive study of the Gafchromic EBT2 radiochromic film. A comparison with EBT [J].
Andres, C. ;
del Castillo, A. ;
Tortosa, R. ;
Alonso, D. ;
Barquero, R. .
MEDICAL PHYSICS, 2010, 37 (12) :6271-6278
[3]   The effects of compensator and imaging geometry on the distribution of x-ray scatter in CBCT [J].
Bootsma, G. J. ;
Verhaegen, F. ;
Jaffray, D. A. .
MEDICAL PHYSICS, 2011, 38 (02) :897-914
[4]   The effect of scatter and glare on image quality in contrast-enhanced breast imaging using an a-Si/CsI(Tl) full-field flat panel detector [J].
Carton, Ann-Katherine ;
Acciavatti, Raymond ;
Kuo, Johnny ;
Maidment, Andrew D. A. .
MEDICAL PHYSICS, 2009, 36 (03) :920-928
[5]   Characterization of image quality and image-guidance performance of a preclinical microirradiator [J].
Clarkson, R. ;
Lindsay, P. E. ;
Ansell, S. ;
Wilson, G. ;
Jelveh, S. ;
Hill, R. P. ;
Jaffray, D. A. .
MEDICAL PHYSICS, 2011, 38 (02) :845-856
[6]   Small fields:: Nonequilibrium radiation dosimetry [J].
Das, Indra J. ;
Ding, George X. ;
Ahnesjo, Anders .
MEDICAL PHYSICS, 2008, 35 (01) :206-215
[7]   The indirect use of CT numbers to establish material properties needed for Monte Carlo calculation of dose distributions in patients [J].
du Plessis, FCP ;
Willemse, CA ;
Lotter, MG ;
Goedhals, L .
MEDICAL PHYSICS, 1998, 25 (07) :1195-1201
[8]   PRACTICAL CONE-BEAM ALGORITHM [J].
FELDKAMP, LA ;
DAVIS, LC ;
KRESS, JW .
JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 1984, 1 (06) :612-619
[9]   Characterization of scattered radiation in kV CBCT images using Monte Carlo simulations [J].
Jarry, Genevieve ;
Graham, Sean A. ;
Moseley, Douglas J. ;
Jaffray, David J. ;
Siewerdsen, Jeffrey H. ;
Verhaegen, Frank .
MEDICAL PHYSICS, 2006, 33 (11) :4320-4329
[10]   Accurate condensed history Monte Carlo simulation of electron transport. I. EGSnrc, the new EGS4 version [J].
Kawrakow, I .
MEDICAL PHYSICS, 2000, 27 (03) :485-498