Prospects of implant with locking plate in fixation of subtrochanteric fracture: experimental demonstration of its potential benefits on synthetic femur model with supportive hierarchical nonlinear hyperelastic finite element analysis

被引:10
作者
Latifi, Mohammed Hadi [1 ,2 ]
Ganthel, Kunalan [1 ,2 ]
Rukmanikanthan, Shanmugam [1 ,2 ]
Mansor, Azura [1 ,2 ]
Kamarul, Tunku [1 ,2 ]
Bilgen, Mehmet [1 ]
机构
[1] Univ Malaya, Fac Med, Dept Orthoped Surg, Kuala Lumpur 50603, Malaysia
[2] Univ Malaya, Natl Orthopaed Ctr Excellence Res & Learning, Kuala Lumpur 50603, Malaysia
关键词
Locking plate; Angle blade plate; Dynamic condylar screw plate; Subtrochanteric fracture; Biomechanics; Finite element analysis; Hierarchical finite element modelling; DYNAMIC CONDYLAR SCREW; FEMORAL FRACTURES; INTACT FEMUR; BIOMECHANICS; MANAGEMENT;
D O I
10.1186/1475-925X-11-23
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Background: Effective fixation of fracture requires careful selection of a suitable implant to provide stability and durability. Implant with a feature of locking plate (LP) has been used widely for treating distal fractures in femur because of its favourable clinical outcome, but its potential in fixing proximal fractures in the subtrochancteric region has yet to be explored. Therefore, this comparative study was undertaken to demonstrate the merits of the LP implant in treating the subtrochancteric fracture by comparing its performance limits against those obtained with the more traditional implants; angle blade plate (ABP) and dynamic condylar screw plate (DCSP). Materials and Methods: Nine standard composite femurs were acquired, divided into three groups and fixed with LP (n=3), ABP (n=3) and DCSP (n=3). The fracture was modeled by a 20 mm gap created at the subtrochanteric region to experimentally study the biomechanical response of each implant under both static and dynamic axial loading paradigms. To confirm the experimental findings and to understand the critical interactions at the boundaries, the synthetic femur/implant systems were numerically analyzed by constructing hierarchical finite element models with nonlinear hyperelastic properties. The predictions from the analyses were then compared against the experimental measurements to demonstrate the validity of each numeric model, and to characterize the internal load distribution in the femur and load bearing properties of each implant. Results: The average measurements indicated that the constructs with ABP, DCPS and LP respectively had overall stiffness values of 70.9, 110.2 and 131.4 N/mm, and exhibited reversible deformations of 12.4, 4.9 and 4.1 mm when the applied dynamic load was 400 N and plastic deformations of 11.3, 2.4 and 1.4 mm when the load was 1000 N. The corresponding peak cyclic loads to failure were 1100, 1167 and 1600 N. The errors between the displacements measured experimentally or predicted by the nonlinear hierarchical hyperelastic model were less than 18%. In the implanted femur heads, the principal stresses were spatially heterogeneous for ABP and DCSP but more homogenous for LP, meaning LP had lower stress concentrations. Conclusion: When fixed with the LP implant, the synthetic femur model of the subtrochancteric fracture consistently exceeds in the key biomechanical measures of stability and durability. These capabilities suggest increased resistance to fatigue and failure, which are highly desirable features expected of functional implants and hence make the LP implant potentially a viable alternative to the conventional ABP or DCSP in the treatment of subtrochancteric femur fractures for the betterment of clinical outcome.
引用
收藏
页数:18
相关论文
共 23 条
[1]   Reversed LISS-DF in selected cases of complex proximal femur fractures [J].
Acklin, Yves P. ;
Bereiter, Heinz ;
Sommer, Christoph .
INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2010, 41 (04) :427-429
[2]   Influence of materials for fixation implants on local infection - An experimental study of steel versus titanium DCP in rabbits [J].
Arens, S ;
Schlegel, U ;
Printzen, G ;
Ziegler, WJ ;
Perren, SM ;
Hansis, M .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1996, 78B (04) :647-651
[3]   A biomechanical comparison of the Surgical Implant Generation Network (SIGN) tibial nail with the standard hollow nail [J].
Calafi, L. A. ;
Antkowiak, T. ;
Curtiss, S. ;
Neu, C. P. ;
Moehring, D. .
INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2010, 41 (07) :753-757
[4]   A Biomechanical Comparison of Locked Plate Fixation With Percutaneous Insertion Capability Versus the Angled Blade Plate in a Subtrochanteric Fracture Gap Model [J].
Crist, Brett D. ;
Khalafi, Afshin ;
Hazelwood, Scott J. ;
Lee, Mark A. .
JOURNAL OF ORTHOPAEDIC TRAUMA, 2009, 23 (09) :622-627
[5]   Biomechanics of locked plates and screws [J].
Egol, KA ;
Kubiak, EN ;
Fulkerson, E ;
Kummer, FJ ;
Koval, KJ .
JOURNAL OF ORTHOPAEDIC TRAUMA, 2004, 18 (08) :488-493
[6]   Biomechanical Comparison of Proximal Locking Plates and Blade Plates for the Treatment of Comminuted Subtrochanteric Femoral Fractures [J].
Floyd, John C. P. ;
O'Toole, Robert V. ;
Stall, Alec ;
Forward, Daren P. ;
Nabili, Marjan ;
Shillingburg, Daniel ;
Hsieh, Adam ;
Nascone, Jason W. .
JOURNAL OF ORTHOPAEDIC TRAUMA, 2009, 23 (09) :628-633
[7]   Management of periprosthetic femur fractures with a first generation locking plate [J].
Fulkerson, Eric ;
Tejwani, Nirmal ;
Stuchin, Steven ;
Egol, Kenneth .
INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2007, 38 (08) :965-972
[8]  
Ganthel K, 2010, BIOMECHANICAL COMP C
[9]   An experimental study on the interface strength between titanium mesh cage and vertebra in reference to vertebral bone mineral density [J].
Hasegawa, K ;
Abe, M ;
Washio, T ;
Hara, T .
SPINE, 2001, 26 (08) :957-963
[10]   Structural properties of a new design of composite replicate femurs and tibias [J].
Heiner, AD ;
Brown, TD .
JOURNAL OF BIOMECHANICS, 2001, 34 (06) :773-781