Noninvasive and nonocclusive determination of blood pressure using laser Doppler flowmetry

被引:2
|
作者
Elter, P [1 ]
Stork, W [1 ]
Müller-Glaser, KD [1 ]
Lutter, N [1 ]
机构
[1] Univ Karlsruhe, Inst Informat Proc Technol, D-76128 Karlsruhe, Germany
来源
SPECIALTY FIBER OPTICS FOR MEDICAL APPLICATIONS, PROCEEDINGS OF | 1999年 / 3596卷
关键词
blood pressure; noninvasive; laser; Doppler; flowmetry;
D O I
10.1117/12.346715
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
This report describes an approach determining blood pressure noninvasively without cuff. Regarding an elastic, fluid-filled tube as a model of an arterial segment, the solution of the Navier Stokes differential equations delivers a relation between the pressure and velocity pulse. There, simulations prove a minimal sensitivity of blood pressure concerning blood density, blood viscosity and damping. Hence, these parameters can be regarded interindividually as constants. Blood pressure is essentially sensitive on the pulse wave velocity, the velocity pulse, the arterial diameter and the reflection coefficient. To perform measurements, a system was built up comprising at least one laser Doppler blood flow sensor, a high performance DSP hardware and a PC. After individual initial Riva Rocci calibration, arterial diameter and reflection coefficient can be determined. Flow and pulse wave velocity and thus blood pressure can be calculated measuring continuously at least one velocity pulse with the laser Doppler flow sensor at a superficial artery like the a. radialis and simultaneously another cardiovascular signal like an ECG or another flow pulse at a different site of the artery. As a first result, high linear correlations between systolic blood pressure and pulse transit time were obtained.
引用
收藏
页码:188 / 196
页数:9
相关论文
共 50 条
  • [41] Evaluation of the A&D UA-767 and Welch Allyn Spot Vital Signs noninvasive blood pressure monitors using a blood pressure simulator
    Davis, PD
    Dennis, JL
    Railton, R
    JOURNAL OF HUMAN HYPERTENSION, 2005, 19 (03) : 197 - 203
  • [42] Continuous Noninvasive blood pressure measurement by pulse transit time
    Fung, P
    Dumont, G
    Ries, C
    Mott, C
    Ansermino, M
    PROCEEDINGS OF THE 26TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-7, 2004, 26 : 738 - 741
  • [43] Pulsed photoacoustic Doppler flowmetry using a cross-correlation method
    Brunker, J.
    Beard, P.
    PHOTONS PLUS ULTRASOUND: IMAGING AND SENSING 2010, 2010, 7564
  • [44] Acoustic resolution photoacoustic Doppler flowmetry for assessment of patient rectal cancer blood perfusion
    Kou, Sitai
    Leng, Xiandong
    Luo, Hongbo
    Nie, Haolin
    Zhu, Quing
    JOURNAL OF BIOMEDICAL OPTICS, 2024, 29
  • [45] Mucosal blood flow measurements using laser Doppler perfusion monitoring
    Hoff, Dag Arne Lihaug
    Gregersen, Hans
    Hatlebakk, Jan Gunnar
    WORLD JOURNAL OF GASTROENTEROLOGY, 2009, 15 (02) : 198 - 203
  • [46] Noninvasive Blood Pressure Estimation Using Ultrasound and Simple Finite Element Models
    Zakrzewski, Aaron M.
    Anthony, Brian W.
    IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2018, 65 (09) : 2011 - 2022
  • [47] Noninvasive Blood Pressure Classification Based on Photoplethysmography Using Machine Learning Techniques
    Mohammadi, Hanieh
    Tarvirdizadeh, Bahram
    Alipour, Khalil
    Ghamari, Mohammad
    2024 32ND INTERNATIONAL CONFERENCE ON ELECTRICAL ENGINEERING, ICEE 2024, 2024, : 213 - 219
  • [48] Noninvasive Cuffless Estimation of Blood Pressure using Photoplethysmography without Electrocardiograph Measurement
    Samria, Rohan
    Jain, Ridhi
    Jha, Ankita
    Saini, Sandeep
    Chowdhury, Shubhajit Roy
    2014 IEEE REGION 10 SYMPOSIUM, 2014, : 254 - 257
  • [49] MEMS characterization using laser Doppler vibrometry
    Lawrence, EM
    Speller, KE
    Yu, DL
    RELIABILITY, TESTING, AND CHARACTERIZATION OF MEMS/MOEMS II, 2003, 4980 : 51 - 62
  • [50] Noninvasive and cuffless measurements of blood pressure for telemedicine
    Chan, KW
    Hung, K
    Zhang, YT
    PROCEEDINGS OF THE 23RD ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-4: BUILDING NEW BRIDGES AT THE FRONTIERS OF ENGINEERING AND MEDICINE, 2001, 23 : 3592 - 3593