摘要
This report describes an approach determining blood pressure noninvasively without cuff. Regarding an elastic, fluid-filled tube as a model of an arterial segment, the solution of the Navier Stokes differential equations delivers a relation between the pressure and velocity pulse. There, simulations prove a minimal sensitivity of blood pressure concerning blood density, blood viscosity and damping. Hence, these parameters can be regarded interindividually as constants. Blood pressure is essentially sensitive on the pulse wave velocity, the velocity pulse, the arterial diameter and the reflection coefficient. To perform measurements, a system was built up comprising at least one laser Doppler blood flow sensor, a high performance DSP hardware and a PC. After individual initial Riva Rocci calibration, arterial diameter and reflection coefficient can be determined. Flow and pulse wave velocity and thus blood pressure can be calculated measuring continuously at least one velocity pulse with the laser Doppler flow sensor at a superficial artery like the a. radialis and simultaneously another cardiovascular signal like an ECG or another flow pulse at a different site of the artery. As a first result, high linear correlations between systolic blood pressure and pulse transit time were obtained.
引用
收藏
页码:188 / 196
页数:9
相关论文
共 50 条
  • [21] Wide-field stabilized Doppler flowmetry with the tracking scanning laser ophthalmoscope (TSLO)
    Ferguson, RD
    Hammer, DX
    Elsner, AE
    Burns, SA
    Webb, RH
    Weiter, JJ
    OPHTHALMIC TECHNOLOGIES XIV, 2004, 5314 : 170 - 178
  • [22] EFFECT OF OXYMETAZOLINE ON NASAL AND SINUS MUCOSAL BLOOD-FLOW IN THE RABBIT AS MEASURED WITH LASER-DOPPLER FLOWMETRY
    AKERLUND, A
    BENDE, M
    ARFORS, KE
    INTAGLIETTA, M
    ANNALS OF OTOLOGY RHINOLOGY AND LARYNGOLOGY, 1993, 102 (02) : 123 - 126
  • [23] Reliability of different image analysis methods for scanning laser Doppler flowmetry
    Kreis, Andreas J.
    Nguyen, Thanh
    Rogers, Sophie
    Wang, Jie Jin
    Harazny, Joanna
    Michelson, Georg
    Farouque, H. M. Omar
    Wong, Tien Y.
    CURRENT EYE RESEARCH, 2008, 33 (5-6) : 493 - 499
  • [24] Gingival microcirculation response to tooth brushing measured by laser Doppler flowmetry
    Perry, DA
    McDowell, J
    Goodis, HE
    JOURNAL OF PERIODONTOLOGY, 1997, 68 (10) : 990 - 995
  • [25] LASER-DOPPLER FLOWMETRY - A REVIEW OF ITS APPLICATION FOR MEASURING CEREBRAL AND SPINAL-CORD BLOOD-FLOW
    FRERICHS, KU
    FEUERSTEIN, GZ
    MOLECULAR AND CHEMICAL NEUROPATHOLOGY, 1990, 12 (01) : 55 - 70
  • [26] A non-invasive heart rate measurement system using laser Doppler blood flowmetry with husbandry training of the masked palm civet (Parguma larvata)
    Nogami, Hirofumi
    Ohgata, Satoru
    Saito, Aya
    Ban, Kazuyuki
    Akiyama, Terukazu
    Hiejima, Takumi
    Takigawa, Ryo
    Hosoya, Tadatsugu
    JAPANESE JOURNAL OF APPLIED PHYSICS, 2021, 60 (SC)
  • [27] NONINVASIVE ESTIMATION OF THE BLOOD PRESSURE WAVEFORM IN THE CAROTID ARTERY USING CONTINUOUS FINGER BLOOD PRESSURE MONITORING
    Idzenga, Tim
    Reesink, Koen D.
    van Swelm, Youri
    Hansen, Hendrik H. G.
    Holewijn, Suzanne
    de Korte, Chris L.
    ULTRASOUND IN MEDICINE AND BIOLOGY, 2012, 38 (11) : 1998 - 2006
  • [28] Noninvasive Estimation of Blood Pressure in HeartMate 3 Patients
    Pinsino, Alberto
    Gaudig, Antonia
    Castagna, Francesco
    Mondellini, Giulio M.
    Mcdonnell, Barry J.
    Stoehr, Eric J.
    Cockcroft, John
    Kormos, Robert L.
    Sayer, Gabriel T.
    Uriel, Nir
    Naka, Yoshifumi
    Takeda, Koji
    Yuzefpolskaya, Melana
    Colombo, Paolo C.
    ASAIO JOURNAL, 2024, 70 (02) : e18 - e20
  • [29] Evaluation of a noninvasive transcranial Doppler and blood pressure-based method for the assessment of cerebral perfusion pressure in pregnant women
    Belfort, MA
    Tooke-Miller, C
    Varner, M
    Saade, G
    Grunewald, C
    Nisell, H
    Herd, JA
    HYPERTENSION IN PREGNANCY, 2000, 19 (03) : 331 - 340
  • [30] New Simple Phenomenological Model for Laser Doppler Measurements of Blood Flow in Tissue
    Lapitan, Denis
    Rogatkin, Dmitry
    Persheyev, Saydulla
    Rogatkin, Andrey
    PROCEEDINGS OF THE 10TH INTERNATIONAL JOINT CONFERENCE ON BIOMEDICAL ENGINEERING SYSTEMS AND TECHNOLOGIES, VOL 1: BIODEVICES, 2017, : 98 - 103