Using Monte Carlo methods to commission electron beams: A feasibility study

被引:38
作者
Antolak, JA [1 ]
Bieda, MR [1 ]
Hogstrom, KR [1 ]
机构
[1] Univ Texas, MD Anderson Canc Ctr, Dept Radiat Phys, Houston, TX 77030 USA
关键词
electron beams; commissioning; Monte Carlo;
D O I
10.1118/1.1469626
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The purpose of this study was to investigate the feasibility of using Monte Carlo methods to assist in the commissioning of electron beams for a medical linear accelerator. The EGS4/BEAM code system was used to model an installed linear accelerator at this institution. Following an initial tuning of the input parameters, dosimetry data normally measured during the machine commissioning was calculated using the Monte Carlo code. All commissioning data was calculated for 6- and 12-MeV electron beams, and a subset of the commissioning data was calculated for the 20-MeV electron beams. On central axis, calculated percentage depth dose, cross-beam profiles, cone-insert ratios, and air-gap factors were generally within 2% of D-max or 1 mm of the measured commissioning data; however, calculated open-cone ratios were not within 2%, in most cases. Calculated off-axis dose profiles for small fields were generally within the 2% (1-mm) criteria; however, calculated dose profiles for larger (open cone) fields frequently failed the 2% (1-mm) criteria. The remaining discrepancies between Monte Carlo calculations and measurement could be due to flaws in the Monte Carlo code, inaccuracies in the simulation geometry, the approximation of the initial source configuration, or a combination of the above. Although agreement between Monte Carlo calculated and measured doses was impressive and similar to previously published comparisons., our results did not prove our hypothesis that Monte Carlo calculations can generate electron commissioning data that is accurate within 2% of D-max or 0.1 cm over the entire range of clinical treatment parameters. Although we believe that this hypothesis can be proved, it remains a challenge for the medical physics community. We intend to pursue this further by developing systematic methods for isolating causes of these differences. (C) 2002 American Association of Physicists in Medicine.
引用
收藏
页码:771 / 786
页数:16
相关论文
共 26 条
  • [1] [Anonymous], 1983, MED PHYS, V10, P741
  • [2] [Anonymous], 1984, 35 ICRU
  • [3] Bevington P., 2002, Data Reduction and Error Analysis for the Physical Sciences, V3rd ed.
  • [4] The effect of scattering foil parameters on electron-beam Monte Carlo calculations
    Bieda, MR
    Antolak, JA
    Hogstrom, KR
    [J]. MEDICAL PHYSICS, 2001, 28 (12) : 2527 - 2534
  • [5] BIEDA MR, 1998, MED PHYS, V25, P1577
  • [6] Bielajew A., 1990, The Dosimetry of Ionizing Radiation, Volume, P427
  • [7] BIELAJEW AF, 1987, NUCL INSTRUM METH B, V18, P165
  • [8] DING GX, 1995, PIRS439 NAT RES COUN
  • [9] HOGSTROM KR, 1991, ADV RAD ONCOLOGY PHY, P390
  • [10] X-RAY SOURCES OF MEDICAL LINEAR ACCELERATORS - FOCAL AND EXTRA-FOCAL RADIATION
    JAFFRAY, DA
    BATTISTA, JJ
    FENSTER, A
    MUNRO, P
    [J]. MEDICAL PHYSICS, 1993, 20 (05) : 1417 - 1427