Short-term in vivo wear of cross-linked polyethylene

被引:121
作者
Heisel, C [1 ]
Silva, M [1 ]
dela Rosa, MA [1 ]
Schmalzried, TP [1 ]
机构
[1] Orthoped Hosp, Joint Replacement Inst, Los Angeles, CA 90007 USA
关键词
D O I
10.2106/00004623-200404000-00012
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background: Cross-linked polyethylene was developed to reduce volumetric wear in prosthetic joints. Hip simulator studies have shown promising results with regard to wear reduction. This study evaluated the short-term in vivo wear of a moderately cross-linked polyethylene. Methods: Linear head penetration, as an assessment of in vivo polyethylene wear, was measured in two groups of patients after total hip replacement. Twenty-four hips received a conventional polyethylene insert and thirty-four, a crosslinked polyethylene liner; both inserts were manufactured by the same company. Linear and volumetric wear rates were measured on radiographs with use of a validated computer-assisted technique and were adjusted for patient-related factors. Patient activity was assessed by a computerized two-dimensional accelerometer worn on the ankle. Results: Patients with a conventional polyethylene insert showed a mean linear wear rate of 0.13 mm per year and a mean volumetric wear rate of 87.6 mm(3) per year. The group with a cross-linked polyethylene liner showed a mean linear wear rate of 0.02 mm per year and a mean volumetric wear rate of 17.0 mm(3) per year. Wear in the group with cross-linked polyethylene was 81% lower than that in the group with conventional polyethylene (p < 0.00001). Accounting for differences in patient activity, the adjusted wear rates per million cycles for a patient weight of 70 kg were 53 mm(3) per million cycles for conventional polyethylene and 15 mm(3) per million cycles for cross-linked polyethylene, a 72% reduction (p = 0.0002). No factor, other than the type of polyethylene, was found to influence the difference in wear rates between the two groups. Conclusions: The results of this study are promising. The in vivo wear reduction with this cross-linked polyethylene is consistent with the predictions of hip simulator studies. Level of Evidence: Therapeutic study, Level II-1 (prospective cohort study). See Instructions to Authors for a complete description of levels of evidence.
引用
收藏
页码:748 / 751
页数:4
相关论文
共 32 条
  • [1] TREATMENT OF PRIMARY OSTEOARTHRITIS OF THE HIP - A COMPARISON OF TOTAL JOINT AND SURFACE REPLACEMENT ARTHROPLASTY
    AMSTUTZ, HC
    THOMAS, BJ
    JINNAH, R
    KIM, W
    GROGAN, T
    YALE, C
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1984, 66A (02) : 228 - 241
  • [2] Baker DA, 1999, J BIOMED MATER RES, V46, P573, DOI 10.1002/(SICI)1097-4636(19990915)46:4<573::AID-JBM16>3.0.CO
  • [3] 2-A
  • [4] POLYETHYLENE WEAR IN CEMENTED METAL-BACKED ACETABULAR CUPS
    CATES, HE
    FARIS, PM
    KEATING, EM
    RITTER, MA
    [J]. JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1993, 75 (02): : 249 - 253
  • [5] DEVANE PA, 1999, CLIN ORTHOP RELAT R, V369, P59
  • [6] Digas G., 2003, T ORTHOP RES SOC, V28, P1430
  • [7] The need to account for patient activity when evaluating the results of total hip arthroplasty with survivorship analysis
    Dorey, FJ
    Amstutz, HC
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2002, 84A (05) : 709 - 710
  • [8] Validation of two and three-dimensional radiographic techniques for measuring polyethylene wear after total hip arthroplasty
    Hui, AJ
    McCalden, FW
    Martell, JM
    MacDonald, SJ
    Bourne, RB
    Rorabeck, CH
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2003, 85A (03) : 505 - 511
  • [9] JASTY M, 1997, T SOC BIOMATER, V20, P157
  • [10] Complications of total hip arthroplasty associated with the use of an acetabular component with a Hylamer liner
    Livingston, BJ
    Chmell, MJ
    Spector, M
    Poss, R
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1997, 79A (10) : 1529 - 1538