Kinematic analysis of mobile-bearing and fixed-bearing knee prostheses by simulation

被引:7
作者
Higashijima, K
Ishida, A [1 ]
Fukuoka, Y
Hoshino, A
Minamitani, H
机构
[1] Keio Univ, Grad Sch Sci & Technol, Inst Biomed Engn, Yokohama, Kanagawa, Japan
[2] Tokyo Med & Dent Univ, Inst Biomat & Bioengn, Tokyo, Japan
[3] Kawaguchi Kohgyo Gen Hosp, Kawaguchi, Japan
关键词
mobile-bearing; fixed-bearing; prosthesis; kinematics; simulation;
D O I
10.1007/BF02347691
中图分类号
TP39 [计算机的应用];
学科分类号
081203 ; 0835 ;
摘要
The object of the study was to model fixed-bearing knee prostheses (FBKs) and mobile-bearing knee prostheses (MBKs) during weight-bearing deep knee bends and to analyse and compare the kinematics of the two prosthesis types. To obtain quantitative data, an overall model of a leg was constructed, and this included a three-dimensional model of the tibiofemoral joint and simplified two-dimensional models of the ankle and patellofemoral joint. The simulated movement pattern of the tibiofemoral contact point in the FBK was analysed to show the posterior contact position on the tibia at full extension and anterior translation as the knee was flexed from 30degrees to 90degrees. The simulated maximum displacements of the medial and lateral contact positions of the FBK were 5.6 and 6.2 mm, respectively. These results were almost in agreement with experimental studies. Compared with the FBK, the movement pattern of the tibiofemoral contact point in the MBK for the anterior contact position on the tibia at full extension and posterior translation, with respect to the tibia as the knee was flexed, gave results closer to those of the normal knee. The simulated displacements of the medial and lateral contact positions of the MBK with respect to the tibia were 9.0 and 13.0 mm from full extension to 90degrees flexion, respectively. The difference in the kinematic results between the FBK and the MBK could be accounted for by movement of the insert and the larger force of the posterior cruciate ligament on the MBK.
引用
收藏
页码:22 / 28
页数:7
相关论文
共 22 条
[1]   MODEL STUDIES OF THE STIFFNESS CHARACTERISTICS OF THE HUMAN KNEE-JOINT [J].
ANDRIACCHI, TP ;
MIKOSZ, RP ;
HAMPTON, SJ ;
GALANTE, JO .
JOURNAL OF BIOMECHANICS, 1983, 16 (01) :23-29
[2]   In vivo kinematics of cruciate-retaining and -substituting knee arthroplasties [J].
Banks, SA ;
Markovich, GD ;
Hodge, WA .
JOURNAL OF ARTHROPLASTY, 1997, 12 (03) :297-304
[3]  
DENNIS DA, 1996, CLIN ORTHOP RELAT R, V331, P107
[4]  
DENNIS DA, 1998, CLIN ORTHOP RELAT R, V356, P47
[5]   INTERACTION BETWEEN INTRINSIC KNEE MECHANICS AND THE KNEE EXTENSOR MECHANISM [J].
DRAGANICH, LF ;
ANDRIACCHI, TP ;
ANDERSSON, GBJ .
JOURNAL OF ORTHOPAEDIC RESEARCH, 1987, 5 (04) :539-547
[6]  
EIJDEN TMG, 1985, J BIOMECH, V19, P219
[7]   A MATHEMATICAL-MODEL FOR THE EVALUATION OF THE BEHAVIOR DURING FLEXION OF CONDYLAR-TYPE KNEE PROSTHESES [J].
ESSINGER, JR ;
LEYVRAZ, PF ;
HEEGARD, JH ;
ROBERTSON, DD .
JOURNAL OF BIOMECHANICS, 1989, 22 (11-12) :1229-&
[8]  
*FDA, 1984, PREM APPL LCS MEN BE
[9]  
KAPANDJI IA, 1982, PHYSL JOINT, V2, P68
[10]   Total knee arthroplasty kinematics - Computer simulation and intraoperative evaluation [J].
Martelli, S ;
Ellis, RE ;
Marcacci, M ;
Zaffagnini, S .
JOURNAL OF ARTHROPLASTY, 1998, 13 (02) :145-155