Micro- and nano-volume samples by electrothermal, near-torch vaporization sample introduction using removable, interchangeable and portable rhenium coiled-filament assemblies and axially-viewed inductively coupled plasma-atomic emission spectrometry

被引:30
作者
Badiei, Hamid R. [1 ]
Lai, Bryant [1 ]
Karanassios, Vassili [1 ]
机构
[1] Univ Waterloo, Dept Chem, Waterloo, ON N2L 3G1, Canada
基金
加拿大自然科学与工程研究理事会;
关键词
Electrothermal vaporization; Near-torch vaporization (NTV); Inductively coupled plasma-atomic emission; spectrometry (ICP-AES); Microsamples; Nanosamples; MASS-SPECTROMETRY; BIOLOGICAL SAMPLES; ITV SAMPLE; TRACE; INSERTION; ELEMENTS; JUICE; NANOPARTICLES; MODIFIER; ABLATION;
D O I
10.1016/j.sab.2012.07.025
中图分类号
O433 [光谱学];
学科分类号
0703 ; 070302 ;
摘要
An electrothermal, near-torch vaporization (NW) sample introduction for micro- or nano-volume samples is described. Samples were pipetted onto coiled-filament assemblies that were purposely developed to be removable and interchangeable and were dried and vaporized into a small-volume vaporization chamber that clips onto any ICP torch with a ball joint. Interchangeable assemblies were also constructed to be small-size (e.g., less than 3 cm long with max diameter of 0.65 cm) and light-weight (1.4 g) so that they can be portable. Interchangeable assemblies with volume-capacities in three ranges (i.e., <1 mu L, 1-10 mu l. and 10-100 mu L) were fabricated and used. The horizontally-operated NIV sample introduction was interfaced to an axially-viewed ICP-AES (inductively coupled plasma-atomic emission spectrometry) system and NTV was optimized using ICP-AES and 8 elements (Pb, Cd, Zn, V. Ba, Mg, Be and Ca). Precision was 1.0-2.3% (peak height) and 1.1-2.4% (peak area). Detection limits (obtained using 5 mu L volumes) expressed in absolute-amounts ranged between 4 pg for Pb to 0.3 fg (similar to 5 million atoms) for Ca. Detection limits expressed in concentration units (obtained using 100 mu L volumes of diluted, single-element standard solutions) were: 50 pg/mL for Pb; 10 pg/mL for Cd; 9 pg/mL for Zn: 1 pg/mL for V; 0.9 pg/mL for Ba; 0.5 pg/mL for Mg; 50 fg/mL for Be; and 3 fg/mL for Ca. Analytical capability and utility was demonstrated using the determination of Pb in pg/mL levels of diluted natural water Certified Reference Material (CRM) and the determination of Zn in 80 nL volumes of the liquid extracted from an individual vesicle. It is shown that portable and interchangeable assemblies with dried sample residues on them can be transported without analyte loss (for the concentrations tested), thus opening up the possibility for "taking part of the lab to the sample" applications, such as testing for Cu concentration-compliance with the lead-copper rule of the Environmental Protection Agency. It is also shown that interchangeable assemblies with volume-capacities in the ranges mentioned above can be used interchangeably in the same calibration curve, thus in some cases and for on-site applications, volume-of-sample can be substituted for sample-dilution. (C) 2012 Elsevier B.V. All rights reserved.
引用
收藏
页码:19 / 30
页数:12
相关论文
共 35 条
[1]  
[Anonymous], APPL NOT, V1
[2]   Electrothermal vaporization-inductively coupled plasma-mass spectrometry: A versatile tool for tackling challenging samples A critical review [J].
Aramendia, Maite ;
Resano, Martin ;
Vanhaecke, Frank .
ANALYTICA CHIMICA ACTA, 2009, 648 (01) :23-44
[3]   Simultaneous electrothermal vaporization and nebulizer sample introduction system for inductively coupled plasma mass spectrometry [J].
Arnquist, Isaac J. ;
Kreschollek, Thomas E. ;
Holcombe, James A. .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2011, 66 (3-4) :255-260
[4]   Rhenium-cup, in-torch vaporization (ITV) sample introduction for axially viewed ICP-AES and its application to the analysis of a microscopic, ng-weight solid sample [J].
Badiei, HR ;
Smith, AT ;
Karanassios, V .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2002, 17 (09) :1030-1036
[5]   Calcium content of individual, microscopic, (sub) nanoliter volume Paramecium sp. cells using rhenium-cup in-torch vaporization (ITV) sample introduction and axially viewed ICP-AES [J].
Badiei, HR ;
Rutzke, MA ;
Karanassios, V .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2002, 17 (09) :1007-1010
[6]  
Barnes KW, 1997, ATOM SPECTROSC, V18, P84
[7]  
Beauchemin D., 2000, Discrete Sample Introduction Techniques for Inductively Coupled Plasma Mass Spectrometry
[8]   ELECTROTHERMAL VAPORIZATION FOR SAMPLE INTRODUCTION IN PLASMA SOURCE SPECTROMETRY [J].
CAREY, JM ;
CARUSO, JA .
CRITICAL REVIEWS IN ANALYTICAL CHEMISTRY, 1992, 23 (05) :397-439
[9]   Effects of sample probe insertion on plasma conditions in direct sample insertion-inductively coupled plasma-atomic emission spectrometry [J].
Chan, GCY ;
Fan, MN ;
Chan, WT .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2001, 56 (01) :13-25
[10]  
Curtis C, 2012, P SOC PHOTO-OPT INS, V8366