Diagnostic accuracy of 128-slice dual-source CT coronary angiography: a randomized comparison of different acquisition protocols

被引:26
|
作者
Neefjes, Lisan A. [1 ,2 ,4 ]
Rossi, Alexia [1 ,2 ]
Genders, Tessa S. S. [2 ,3 ]
Nieman, Koen [1 ,2 ]
Papadopoulou, Stella L. [1 ,2 ]
Dharampal, Anoeshka S. [1 ,2 ]
Schultz, Carl J. [1 ]
Weustink, Annick C. [2 ]
Dijkshoorn, Marcel L. [2 ]
ten Kate, Gert-Jan R. [1 ,2 ,4 ]
Dedic, Admir [1 ,2 ]
van Straten, Marcel [2 ]
Cademartiri, Filippo [2 ]
Hunink, M. G. Myriam [2 ,3 ]
Krestin, Gabriel P. [2 ]
de Feyter, Pim J. [1 ,2 ]
Mollet, Nico R. [1 ,2 ]
机构
[1] Erasmus MC, Dept Cardiol, NL-3000 CA Rotterdam, Netherlands
[2] Erasmus MC, Dept Radiol, NL-3000 CA Rotterdam, Netherlands
[3] Erasmus MC, Dept Epidemiol, NL-3000 CA Rotterdam, Netherlands
[4] Interuniv Cardiol Inst Netherlands, Utrecht, Netherlands
关键词
Computed tomography; Coronary angiography; Coronary artery disease; Radiation dosage; Sensitivity and specificity; AMERICAN-HEART-ASSOCIATION; RADIATION-DOSE REDUCTION; COMPUTED-TOMOGRAPHY; HIGH-PITCH; IMAGE QUALITY; PERFORMANCE; EXPOSURE; STRATEGIES; CARDIOLOGY; RADIOLOGY;
D O I
10.1007/s00330-012-2663-3
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
To compare the diagnostic performance and radiation exposure of 128-slice dual-source CT coronary angiography (CTCA) protocols to detect coronary stenosis with more than 50 % lumen obstruction. We prospectively included 459 symptomatic patients referred for CTCA. Patients were randomized between high-pitch spiral vs. narrow-window sequential CTCA protocols (heart rate below 65 bpm, group A), or between wide-window sequential vs. retrospective spiral protocols (heart rate above 65 bpm, group B). Diagnostic performance of CTCA was compared with quantitative coronary angiography in 267 patients. In group A (231 patients, 146 men, mean heart rate 58 +/- 7 bpm), high-pitch spiral CTCA yielded a lower per-segment sensitivity compared to sequential CTCA (89 % vs. 97 %, P = 0.01). Specificity, PPV and NPV were comparable (95 %, 62 %, 99 % vs. 96 %, 73 %, 100 %, P > 0.05) but radiation dose was lower (1.16 +/- 0.60 vs. 3.82 +/- 1.65 mSv, P < 0.001). In group B (228 patients, 132 men, mean heart rate 75 +/- 11 bpm), per-segment sensitivity, specificity, PPV and NPV were comparable (94 %, 95 %, 67 %, 99 % vs. 92 %, 95 %, 66 %, 99 %, P > 0.05). Radiation dose of sequential CTCA was lower compared to retrospective CTCA (6.12 +/- 2.58 vs. 8.13 +/- 4.52 mSv, P < 0.001). Diagnostic performance was comparable in both groups. Sequential CTCA should be used in patients with regular heart rates using 128-slice dual-source CT, providing optimal diagnostic accuracy with as low as reasonably achievable (ALARA) radiation dose.
引用
收藏
页码:614 / 622
页数:9
相关论文
共 50 条
  • [1] Diagnostic accuracy of 128-slice dual-source CT coronary angiography: a randomized comparison of different acquisition protocols
    Lisan A. Neefjes
    Alexia Rossi
    Tessa S. S. Genders
    Koen Nieman
    Stella L. Papadopoulou
    Anoeshka S. Dharampal
    Carl J. Schultz
    Annick C. Weustink
    Marcel L. Dijkshoorn
    Gert-Jan R. ten Kate
    Admir Dedic
    Marcel van Straten
    Filippo Cademartiri
    M. G. Myriam Hunink
    Gabriël P. Krestin
    Pim J. de Feyter
    Nico R. Mollet
    European Radiology, 2013, 23 : 614 - 622
  • [2] Quantification and Limitation of CT Coronary Angiography Using 128-Slice Dual-Source CT at High Heart Rates: A Simulation Study
    Zhao, Lei
    Liu, Aishi
    Guo, Youmin
    JOURNAL OF MEDICAL IMAGING AND HEALTH INFORMATICS, 2018, 8 (06) : 1201 - 1206
  • [3] High-Pitch 128-Slice Dual-Source CT for the Assessment of Coronary Stents in a Phantom Model
    Donati, Olivio F.
    Burg, Matthias C.
    Desbiolles, Lotus
    Karlo, Christoph
    Stolzmann, Paul
    Bunck, Alexander
    Baumueller, Stephan
    Marincek, Borut
    Seifarth, Harald
    Alkadhi, Hatem
    Leschka, Sebastian
    Maintz, David
    ACADEMIC RADIOLOGY, 2010, 17 (11) : 1366 - 1374
  • [4] Accuracy, image quality and radiation dose comparison of high-pitch spiral and sequential acquisition on 128-slice dual-source CT angiography in children with congenital heart disease
    Nie, Pei
    Wang, Ximing
    Cheng, Zhaoping
    Ji, Xiaopeng
    Duan, Yanhua
    Chen, Jiuhong
    EUROPEAN RADIOLOGY, 2012, 22 (10) : 2057 - 2066
  • [5] Dual-source 128-slice MDCT neck: Radiation dose and image quality estimation of three different protocols
    Paul, Jijo
    Mbalisike, Emmanuel C.
    Nour-Eldin, Nour-Eldin A.
    Vogl, Thomas J.
    EUROPEAN JOURNAL OF RADIOLOGY, 2013, 82 (05) : 787 - 796
  • [6] Diagnostic accuracy of dual-source CT coronary angiography in patients with atrial fibrillation: Meta analysis
    Sun, Gang
    Li, Min
    Jiang, Zhi-wei
    Xu, Lin
    Peng, Zhao-hui
    Ding, Juan
    Li, Li
    Jin, Zhi-tao
    EUROPEAN JOURNAL OF RADIOLOGY, 2013, 82 (10) : 1749 - 1754
  • [7] Low-dose, 128-slice, dual-source CT coronary angiography: accuracy and radiation dose of the high-pitch and the step-and-shoot mode
    Alkadhi, Hatem
    Stolzmann, Paul
    Desbiolles, Lotus
    Baumueller, Stephan
    Goetti, Robert
    Plass, Andre
    Scheffel, Hans
    Feuchtner, Gudrun
    Falk, Volkmar
    Marincek, Borut
    Leschka, Sebastian
    HEART, 2010, 96 (12) : 933 - 938
  • [8] Detection of ischaemic myocardial lesions with coronary CT angiography and adenosine-stress dynamic perfusion imaging using a 128-slice dual-source CT: diagnostic performance in comparison with cardiac MRI
    Kim, S. M.
    Choi, J-H
    Chang, S-A
    Choe, Y. H.
    BRITISH JOURNAL OF RADIOLOGY, 2013, 86 (1032)
  • [9] Dose levels and image quality of second-generation 128-slice dual-source coronary CT angiography in clinical routine
    Wichmann, Julian L.
    Hu, Xiaohan
    Engler, Alexander
    Kerl, J. Matthias
    Beeres, Martin
    Frellesen, Claudia
    Luboldt, Wolfgang
    Vogl, Thomas J.
    Bauer, Ralf W.
    Lehnert, Thomas
    RADIOLOGIA MEDICA, 2015, 120 (12): : 1112 - 1121
  • [10] Accuracy, image quality and radiation dose comparison of high-pitch spiral and sequential acquisition on 128-slice dual-source CT angiography in children with congenital heart disease
    Pei Nie
    Ximing Wang
    Zhaoping Cheng
    Xiaopeng Ji
    Yanhua Duan
    Jiuhong Chen
    European Radiology, 2012, 22 : 2057 - 2066