Influence of the Peek Abutments on Mechanical Behavior of the Internal Connections Single Implant

被引:5
作者
Melo de Matos, Jefferson David [1 ,2 ]
Scalzer Lopes, Guilherme da Rocha [2 ]
Queiroz, Daher Antonio [3 ]
Jesus Pereira, Andre Luiz [4 ]
Coelho Sinhoreti, Mario Alexandre [5 ]
Ramos, Nathalia de Carvalho [2 ,6 ,7 ]
Lino, Vinicius [2 ]
de Oliveira, Flavio Rosa [2 ]
Souto Borges, Alexandre Luiz [2 ]
Bottino, Marco Antonio [2 ]
机构
[1] Univ Florida, Ctr Dent Biomat, Dept Restorat Dent Sci, UF Hlth, Gainesville, FL 32601 USA
[2] Sao Paulo State Univ, Inst Sci & Technol, Dept Biomat Dent Mat & Prosthodont, UNESP, BR-12228900 Sao Jose Dos Campos, SP, Brazil
[3] Univ Texas Hlth Sci Ctr Houston UTHlth, Sch Dent, Dept Restorat Dent & Prosthodont, Houston, TX 77054 USA
[4] Aeronaut Technol Inst ITA, Phys Dept, Plasmas & Proc Lab, 50 Marechal Eduardo Gomes Sq, BR-12228900 Sao Jose Dos Campos, SP, Brazil
[5] FOP UNICAMP, Dept Restorat Dent, Dent Mat Div, Piracicaba Dent Sch, BR-13416000 Piracicaba, SP, Brazil
[6] Univ Sao Francisco USF, Dept Dent, BR-12916900 Braganca Paulista, SP, Brazil
[7] Univ Taubate UNITAU, Dept Dent, Postgrad Program Dent, BR-12080000 Taubate, SP, Brazil
关键词
bioengineering; dental abutments; dental implants; dental materials; polyetheretherketone; IN-VITRO EVALUATION; STRESS-DISTRIBUTION; FRAMEWORK; FORCE; DEFORMATION; PROSTHESIS; INTERFACE; DESIGN; ALLOY;
D O I
10.3390/ma15228133
中图分类号
O64 [物理化学(理论化学)、化学物理学];
学科分类号
070304 ; 081704 ;
摘要
The present study aimed to evaluate the biomechanical behavior of PEEK abutments with different heights on single titanium implants. To investigate the implant surface, different tests (scanning electron microscopy, energy-dispersive X-ray, and X-ray diffraction) were adopted. Herein, 20 implants received the 4.5 x 4.0 mm PEEK short abutment (SA) and 20 received the 4.5 x 5.5 mm PEEK long abutment (LA). The abutments were installed using dual-cure resin cement. To determine the fatigue test, two specimens from each group were submitted to the single load fracture test. For this, the samples were submitted to a compressive load of (0.5 mm/min; 30 degrees) in a universal testing machine. For the fatigue test, the samples received 2,000,000 cycles (2 Hz; 30 degrees). The number of cycles and the load test was analyzed by the reliability software SPSS statistics using Kaplan-Meier and Mantel-Cox tests (log-rank) (p < 0.05). The maximum load showed no statistically significant differences (p = 0.189) for the SA group (64.1 kgf) and the LA group (56.5 kgf). The study groups were statistically different regarding the number of cycles (p = 0.022) and fracture strength (p = 0.001). PEEK abutments can be indicated with caution for implant-supported rehabilitation and may be suitable as temporary rehabilitation.
引用
收藏
页数:14
相关论文
共 50 条
  • [41] Axial Displacement of Abutments into Implants and Implant Replicas, with the Tapered Cone-Screw Internal Connection, as a Function of Tightening Torque
    Dailey, Bruno
    Jordan, Laurence
    Blind, Olivier
    Tavernier, Bruno
    INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 2009, 24 (02) : 251 - 256
  • [42] Reverse Torque and Wear of Internal Connections of Metal-Based Abutments Under Cyclic Loading
    Queiroz, Daher Antonio
    Freitas, Mariana Itaborai Moreira
    Ockey, Kaden Thomas
    Lee, Damian Jaewhan
    Zheng, Fengyuan
    JOURNAL OF ESTHETIC AND RESTORATIVE DENTISTRY, 2025,
  • [43] Effect of material and dimensions on the mechanical behavior of a dental implant
    Hedia, Hassan S.
    Najjar, Ismail M. R.
    Fouda, Noha
    Al-Thobiani, Fisal W.
    Timraz, Hattan A.
    MATERIALS TESTING, 2020, 62 (08) : 775 - 782
  • [44] Influence of CAD/CAM Abutment Heights on the Biomechanical Behavior of Zirconia Single Crowns
    de Matos, Jefferson David Melo
    Gomes, Leonardo Silva
    Ramos, Nathalia de Carvalho
    Queiroz, Daher Antonio
    Tribst, Joao Paulo Mendes
    Campos, Tiago Moreira Bastos
    Borges, Alexandre Luiz Souto
    Scalzer Lopes, Guilherme da Rocha
    Bottino, Marco Antonio
    Paes Junior, Tarcisio Jose Arruda
    METALS, 2022, 12 (12)
  • [45] Characteristics of implant-CAD/CAM abutment connections of two different internal connection systems
    Sumi, T.
    Braian, M.
    Shimada, A.
    Shibata, N.
    Takeshita, K.
    Vandeweghe, S.
    Coelho, P. G.
    Wennerberg, A.
    Jimbo, R.
    JOURNAL OF ORAL REHABILITATION, 2012, 39 (05): : 391 - 398
  • [46] Implants With Internal Hexagon and Conical Implant-Abutment Connections: An In Vitro Study of the Bacterial Contamination
    D'Ercole, Simonetta
    Scarano, Antonio
    Perrotti, Vittoria
    Mulatinho, Jorge
    Piattelli, Adriano
    Iezzi, Giovanna
    Tripodi, Domenico
    JOURNAL OF ORAL IMPLANTOLOGY, 2014, 40 (01) : 31 - 36
  • [47] Randomized controlled clinical study of veneered zirconia abutments for single implant crowns: Clinical, histological, and microbiological outcomes
    Thoma, Daniel S.
    Sailer, Irena
    Muhlemann, Sven
    Gil, Alfonso
    Jung, Ronald E.
    Hammerle, Christoph H. F.
    CLINICAL IMPLANT DENTISTRY AND RELATED RESEARCH, 2018, 20 (06) : 988 - 996
  • [48] Stress distribution in implant-supported prosthesis with external and internal implant-abutment connections
    Goiato, Marcelo Coelho
    Pesqueira, Aldieris Alves
    Falcon-Antenucci, Rosse Mary
    dos Santos, Daniela Micheline
    Haddad, Marcela Filie
    Bannwart, Lisiane Cristina
    Moreno, Amalia
    ACTA ODONTOLOGICA SCANDINAVICA, 2013, 71 (02) : 283 - 288
  • [49] The Esthetic Effect of Veneered Zirconia Abutments for Single-Tooth Implant Reconstructions: A Randomized Controlled Clinical Trial
    Thoma, Daniel S.
    Brandenberg, Francine
    Fehmer, Vincent
    Knechtle, Nathalie
    Hammerle, Christoph H. F.
    Sailer, Irena
    CLINICAL IMPLANT DENTISTRY AND RELATED RESEARCH, 2016, 18 (06) : 1210 - 1217
  • [50] Three-Dimensional Finite Element Analysis of Stress Distribution in Dental Implant Prosthesis and Surrounding Bone Using PEEK Abutments
    Hong, Min-Ho
    Choi, Hyunsuk
    BIOMIMETICS, 2024, 9 (08)