Study on a Multi-Ions Sensing System for Monitoring of Blood Electrolytes With Wireless Home-Care System

被引:12
作者
Cheng, Jui-Fu [1 ,2 ]
Chou, Jung-Chuan [3 ]
Sun, Tai-Ping [4 ]
Hsiung, Shen-Kan [5 ]
Kao, Hui-Ling [1 ]
机构
[1] Chung Yuan Christian Univ, Inst Elect Engn, Chungli 320, Taiwan
[2] Natl Army Acad, Dept Elect Engn, Chungli 320, Taiwan
[3] Natl Yunlin Univ Sci & Technol, Grad Sch Elect Engn, Touliu 640, Taiwan
[4] Natl Chi Nan Univ, Inst Elect Engn, Nantou 545, Taiwan
[5] Chang Jung Christian Univ, Dept Engn & Management Adv Technol, Tainan 711, Taiwan
关键词
Bluetooth module; Lab-VIEW software; multi-ions sensing system; portable module; wireless home-care system; FIELD-EFFECT TRANSISTOR; SENSOR NETWORKS; IMPLEMENTATION; BIOSENSORS; MEMBRANE; INTERNET; MOBILE; GLASS;
D O I
10.1109/JSEN.2011.2161633
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
This study developed a multi-ions sensing system that includes pH, potassium, sodium, and chloride ion sensors with a wireless home-care system. The pH sensor was based on a tin oxide (SnO2)/indium tin oxide (ITO) glass, and the potassium, sodium, and chloride ion sensors were based on the pH sensor with sensing membrane. The tin oxide (SnO2) sensing membrane was deposited on indium tin oxide (ITO) substrate using a sputtering system. The sensing membranes of the ion sensors were prepared by mixing poly (vinyl chloride) (PVC), bis (2-ethylhexyl) sebacate (DOS), ionophores, and additives. According to the experimental results, sensitivities of the sensors all were over the value 51 mV/decade. The linear range of pH sensor was between pH2 and pH12, the linear ranges of potassium, sodium, and chloride sensors all were over from 1 mM to 1 M. Therefore, the multi-ions sensing system is suitable for monitoring of blood electrolytes. The correlations of the multi-ion sensors under different test solutions also were discussed. Moreover, this study also investigated sensors module, portable module, Bluetooth module, personal computer (PC) terminal, and portable unit of the wireless home-care system. A graphical measurement interface for data recorded and measurement results displayed was designed by National Instrument Lab-VIEW software. According to the experimental results, the measurement data by the wireless home-care system are near the real values. The wireless home-care system could successfully monitor and transmitted the sensing signals.
引用
收藏
页码:967 / 977
页数:11
相关论文
共 50 条
[1]   Wireless Passive Sensor Networks [J].
Akan, Ozgur B. ;
Isik, M. Talha ;
Baykal, Buyurman .
IEEE COMMUNICATIONS MAGAZINE, 2009, 47 (08) :92-99
[2]   Routing techniques in wireless sensor networks: A survey [J].
Al-Karaki, JN ;
Kamal, AE .
IEEE WIRELESS COMMUNICATIONS, 2004, 11 (06) :6-28
[3]   Electrochemical sensors [J].
Bakker, E .
ANALYTICAL CHEMISTRY, 2004, 76 (12) :3285-3298
[4]   Work function changes and surface chemistry of oxygen, hydrogen, and carbon on indium tin oxide [J].
Chaney, JA ;
Pehrsson, PE .
APPLIED SURFACE SCIENCE, 2001, 180 (3-4) :214-226
[5]  
Chen JJ, 2010, PROCEEDINGS OF 2009 INTERNATIONAL TEXTILE SCIENCE AND TECHNOLOGY FORUM, P304
[6]  
Cheng J.-F., IEEE SENSOR IN PRESS
[7]  
Cheng JF, 2006, RARE METAL MAT ENG, V35, P240
[8]   All-solid-state separated potassium electrode based on SnO2/ITO glass [J].
Cheng, Jui-Fu ;
Chou, Jung-Chuan ;
Sun, Tai-Ping ;
Hsiunga, Shen-Kan .
JOURNAL OF THE ELECTROCHEMICAL SOCIETY, 2007, 154 (11) :J369-J374
[9]   Study on All-Solid-State Chloride Sensor Based on Tin Oxide/Indium Tin Oxide Glass [J].
Cheng, Jui-Fu ;
Chou, Jung-Chuan ;
Sun, Tai-Ping ;
Hsiung, Shen-Kan ;
Kao, Hui-Ling .
JAPANESE JOURNAL OF APPLIED PHYSICS, 2011, 50 (03)
[10]   Study on extended gate field effect transistor with tin oxide sensing membrane [J].
Chi, LL ;
Chou, JC ;
Chung, WY ;
Sun, TP ;
Hsiung, SK .
MATERIALS CHEMISTRY AND PHYSICS, 2000, 63 (01) :19-23