On the use of image-derived input functions in oncological fluorine-18 fluorodeoxyglucose positron emission tomography studies

被引:39
作者
Hoekstra, CJ
Hoekstra, OS
Lammertsma, AA
机构
[1] Free Univ Amsterdam Hosp, Clin PET Ctr, NL-1007 MB Amsterdam, Netherlands
[2] Free Univ Amsterdam Hosp, Dept Pulm Med, NL-1007 MB Amsterdam, Netherlands
[3] Free Univ Amsterdam Hosp, Dept Clin Epidemiol, NL-1007 MB Amsterdam, Netherlands
关键词
D O I
10.1007/s002590050484
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
There is growing interest in monitoring response to therapy in oncology patients using positron emission tomography and fluorine-ls fluorodeoxyglucose (FDG). Quantification is required, and various methods have been described. Arterial sampling, which provides the most accurate input function, often is not feasible in patients undergoing chemotherapy. In the thorax an image-derived input function can be obtained from the large vascular structures. In many studies the left ventricle (LV) is used, but the obtained data are rarely validated. In this study a simple quality control procedure for the image-derived input function was developed and compared with a standard LV curve. Twenty dynamic FDG scans were obtained in nine patients with nonsmall cell lung cancer stage IIIA-N2. Three venous blood samples were taken as a quality control for the image-derived input function. Regions of interest (ROIs) were defined for aorta, LV and left atrium. Input curves were generated according the the standard use of the LV curve and by applying the quality control method to all vascular ROIs. Tumour ROIs were defined and both input functions were used to calculate tumour glucose metabolism (MRglu), using both standard non-linear regression and Patlak analyses. Mean differences in MRglu using "standard" LV and "quality control approved" input functions were 16.2% and 17.5% for non-linear regression and Patlak analyses, respectively, iondicating that the use of "standard" LV curve might lead to significant errors. It is concluded that care should be taken in using image derived input functions without appropriate quality control. The proposed procedure is simple and results in significantly more accurate MRglu data.
引用
收藏
页码:1489 / 1492
页数:4
相关论文
共 19 条
[1]  
Avril N, 1997, J NUCL MED, V38, P1186
[2]  
Brix G, 1997, J NUCL MED, V38, P1614
[3]   Whole-body 18FDG positron emission tomography in the staging of non-small cell lung cancer [J].
Bury, T ;
Dowlati, A ;
Paulus, P ;
Corhay, JL ;
Hustinx, R ;
Ghaye, B ;
Radermecker, M ;
Rigo, P .
EUROPEAN RESPIRATORY JOURNAL, 1997, 10 (11) :2529-2534
[4]   Noninvasive quantification of the cerebral metabolic rate for glucose using positron emission tomography, 18F-fluoro-2-deoxyglucose, the Patlak method, and an image-derived input function [J].
Chen, K ;
Bandy, D ;
Reiman, E ;
Huang, SC ;
Lawson, M ;
Feng, D ;
Yun, LS ;
Palant, A .
JOURNAL OF CEREBRAL BLOOD FLOW AND METABOLISM, 1998, 18 (07) :716-723
[5]   PET and [F-18]-FDG in oncology: A clinical update [J].
Conti, PS ;
Lilien, DL ;
Hawley, K ;
Keppler, J ;
Grafton, ST ;
Bading, JR .
NUCLEAR MEDICINE AND BIOLOGY, 1996, 23 (06) :717-735
[6]  
GAMBHIR SS, 1989, J NUCL MED, V30, P359
[7]   QUANTITATION IN POSITRON EMISSION COMPUTED-TOMOGRAPHY .1. EFFECT OF OBJECT SIZE [J].
HOFFMAN, EJ ;
HUANG, SC ;
PHELPS, ME .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1979, 3 (03) :299-308
[8]  
Hunter GJ, 1998, CLIN CANCER RES, V4, P949
[9]   Estimation of myocardial glucose utilisation with PET using the left ventricular time-activity curve as a non-invasive input function [J].
Li, X ;
Feng, D ;
Lin, KP ;
Huang, SC .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1998, 36 (01) :112-117
[10]   CORRECTION OF SPILLOVER RADIOACTIVITIES FOR ESTIMATION OF THE BLOOD TIME-ACTIVITY CURVE FROM THE IMAGED LV CHAMBER IN CARDIAC DYNAMIC FDG PET STUDIES [J].
LIN, KP ;
HUANG, SC ;
CHOI, Y ;
BRUNKEN, RC ;
SCHELBERT, HR ;
PHELPS, ME .
PHYSICS IN MEDICINE AND BIOLOGY, 1995, 40 (04) :629-642