Comprehensive methodology for commissioning modern 3D-image-based treatment planning systems for high dose rate gynaecological brachytherapy: A review

被引:3
作者
Kanani, Abolfazl [1 ]
Owrangi, Amir M. [2 ]
Mosleh-Shirazi, Mohammad Amin [1 ,3 ]
机构
[1] Shiraz Univ Med Sci, Ionizing & Nonionizing Radiat Protect Res Ctr, Sch Paramed Sci, Shiraz, Iran
[2] UT Southwestern Med Ctr, Dept Radiat Oncol, 2280 Inwood Rd,EC2-242, Dallas, TX 75235 USA
[3] Shiraz Univ Med Sci, Sch Med, Dept Radiooncol, Phys Unit, Shiraz 7193613311, Iran
来源
PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS | 2020年 / 77卷
关键词
HDR; Brachytherapy; TPS; Qualityassurance (QA); Commissioning; COLLAPSED-CONE ENGINE; SOCIETY CONSENSUS GUIDELINES; CERVIX CANCER BRACHYTHERAPY; LOCALLY ADVANCED-CARCINOMA; QUALITY-ASSURANCE; INTRACAVITARY BRACHYTHERAPY; AMERICAN BRACHYTHERAPY; TREATMENT DELIVERY; VOLUME PARAMETERS; RECOMMENDATIONS;
D O I
10.1016/j.ejmp.2020.07.031
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: Correct commissioning of treatment planning systems (TPSs) is important for reducing treatment failure events. There is currently no comprehensive and robust methodology available for TPS commissioning in modern brachytherapy. This review aimed to develop a comprehensive template for commissioning modern 3D-image-based brachytherapy TPSs for high dose rate (HDR) gynaecological applications. Methods: The literature relevant to TPS commissioning, including both external beam radiation therapy (EBRT) and brachytherapy, as well as guidelines by the International Atomic Energy Agency (IAEA), the American Association of Physicists in Medicine (AAPM), and the European Society for Radiotherapy and Oncology (ESTRO) were searched, studied and appraised. The applied relevant EBRT TPS commissioning tests were applied to brachytherapy. The developed template aimed to cover all dosimetric and non-dosimetric issues. Results: The essential commissioning items could be categorized into six parts: geometry, dose calculation, plan evaluation tools, plan optimization, TPS output, and end-to-end verification. The final template consists of 43 items. This paper presents the purpose and role of each test, as well as tolerance limits, to facilitate the use of the template. Conclusion: The information and recommendations available in a collection of publications over many years have been reviewed in order to develop a comprehensive template for commissioning complex modern 3D-image-based brachytherapy TPSs for HDR gynaecological applications. The up-to-date and concise information contained in the template can aid brachytherapy physicists during TPS commissioning as well as devising a regular quality assurance program and allocation of time and resources.
引用
收藏
页码:21 / 29
页数:9
相关论文
共 77 条
  • [61] A brief look at model-based dose calculation principles, practicalities, and promise
    Slobodo, Ron S.
    Morrison, Hali
    Cowston-Grant, Brie
    Menon, Geetha V.
    [J]. JOURNAL OF CONTEMPORARY BRACHYTHERAPY, 2017, 9 (01) : 79 - 88
  • [62] The influence of the dwell time deviation constraint (DTDC) parameter on dosimetry with IPSA optimisation for HDR prostate brachytherapy
    Smith, Ryan L.
    Panettieri, Vanessa
    Lancaster, Craig
    Mason, Natasha
    Franich, Rick D.
    Millar, Jeremy L.
    [J]. AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE, 2015, 38 (01) : 55 - 61
  • [63] Metal artefacts in MRI-guided brachytherapy of cervical cancer
    Soliman, Abraam S.
    Owrangi, Amir
    Ravi, Ananth
    Song, William Y.
    [J]. JOURNAL OF CONTEMPORARY BRACHYTHERAPY, 2016, 8 (04) : 363 - 369
  • [64] Advancements in brachytherapy
    Tanderup, Kari
    Menard, Cynthia
    Polgar, Csaba
    Lindegaard, Jacob Christian
    Kirisits, Christian
    Poetter, Richard
    [J]. ADVANCED DRUG DELIVERY REVIEWS, 2017, 109 : 15 - 25
  • [65] Tedgren AC, 2015, RADIOTHER ONCOL, V115, pS101
  • [66] Analysis of treatment delivery errors in brachytherapy using formal risk analysis techniques
    Thomadsen, B
    Lin, SW
    Laemmrich, P
    Waller, T
    Cheng, A
    Caldwell, B
    Rankin, R
    Stitt, J
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2003, 57 (05): : 1492 - 1508
  • [67] Monte Carlo dose calculations for high-dose-rate brachytherapy using GPU-accelerated processing
    Tian, Z.
    Zhang, M.
    Hrycushko, B.
    Albuquerque, K.
    Jiang, S. B.
    Jia, X.
    [J]. BRACHYTHERAPY, 2016, 15 (03) : 387 - 398
  • [68] Dosimetric verification of a high dose rate brachytherapy treatment planning system in homogeneous and heterogeneous media
    Uniyal, S. C.
    Sharma, S. D.
    Naithani, U. C.
    [J]. PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2013, 29 (02): : 171 - 177
  • [69] Accuracy requirements and uncertainties in radiotherapy: a report of the International Atomic Energy Agency
    van der Merwe, Debbie
    Van Dyk, Jacob
    Healy, Brendan
    Zubizarreta, Eduardo
    Izewska, Joanna
    Mijnheer, Ben
    Meghzifene, Ahmed
    [J]. ACTA ONCOLOGICA, 2017, 56 (01) : 1 - 6
  • [70] VANDYK J, 1993, INT J RADIAT ONCOL, V26, P261, DOI 10.1016/0360-3016(93)90206-B