The visible signal responsible for proton therapy dosimetry using bare optical fibers is not Cerenkov radiation

被引:57
作者
Darafsheh, Arash [1 ]
Taleei, Reza [2 ]
Kassaee, Alireza [1 ]
Finlay, Jarod C. [1 ]
机构
[1] Univ Penn, Dept Radiat Oncol, Philadelphia, PA 19104 USA
[2] UT Southwestern Med Ctr, Dept Radiat Oncol, Div Med Phys & Engn, Dallas, TX 75390 USA
关键词
Cerenkov radiation; dosimetry; proton therapy; fiber optics; luminescence; spectroscopy; ionization quenching; PLASTIC SCINTILLATION DETECTORS; HIGH-ENERGY; QUENCHING CORRECTION; CHERENKOV RADIATION; LIGHT-EMISSION; SENSOR; RADIOTHERAPY; PERFORMANCE; BEAMS; ARRAY;
D O I
10.1118/1.4964453
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: Proton beam dosimetry using bare plastic optical fibers has emerged as a simple approach to proton beam dosimetry. The source of the signal in this method has been attributed to Cerenkov radiation. The aim of this work was a phenomenological study of the nature of the visible light responsible for the signal in bare fiber optic dosimetry of proton therapy beams. Methods: Plastic fiber optic probes embedded in solid water phantoms were irradiated with proton beams of energies 100, 180, and 225 MeV produced by a proton therapy cyclotron. Luminescence spectroscopy was performed by a CCD-coupled spectrometer. The spectra were acquired at various depths in phantom to measure the percentage depth dose (PDD) for each beam energy. For comparison, the PDD curves were acquired using a standard multilayer ion chamber device. In order to further analyze the contribution of the Cerenkov radiation in the spectra, Monte Carlo simulation was performed using FLUKA Monte Carlo code to stochastically simulate radiation transport, ionizing radiation dose deposition, and optical emission of Cerenkov radiation. Results: The measured depth doses using the bare fiber are in agreement with measurements performed by the multilayer ion chamber device, indicating the feasibility of using bare fiber probes for proton beam dosimetry. The spectroscopic study of proton-irradiated fibers showed a continuous spectrum with a shape different from that of Cerenkov radiation. The Monte Carlo simulations confirmed that the amount of the generated Cerenkov light does not follow the radiation absorbed dose in a medium. Conclusions: The source of the optical signal responsible for the proton dose measurement using bare optical fibers is not Cerenkov radiation. It is fluorescence of the plastic material of the fiber. (C) 2016 American Association of Physicists in Medicine.
引用
收藏
页码:5973 / 5980
页数:8
相关论文
共 66 条
[1]  
[Anonymous], 1958, CERENKOV RAD ITS APP
[2]  
[Anonymous], 2015, P SPIE
[3]   Cherenkov radiation: from discovery to RICH [J].
Bashmakov, Yu A. .
PHYSICS-USPEKHI, 2015, 58 (05) :467-471
[4]   CERENKOV LIGHT GENERATED IN OPTICAL FIBERS AND OTHER LIGHT PIPES IRRADIATED BY ELECTRON-BEAMS [J].
BEDDAR, AS ;
MACKIE, TR ;
ATTIX, FH .
PHYSICS IN MEDICINE AND BIOLOGY, 1992, 37 (04) :925-935
[5]  
Beddar Sam., 2016, Scintillation dosimetry, DOI DOI 10.1201/9781315372655
[6]   Measurement of Neutron Yields From UF4 [J].
Bell, Z. W. ;
Ziock, K. -P. ;
Ohmes, Martin F. ;
Xu, Yunlin ;
Downar, T. ;
Pozzi, S. A. .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2010, 57 (04) :2239-2246
[7]  
Birks J. B., 1964, THEORY PRACTICE SCIN, DOI [DOI 10.1016/C2013-0-01791-4, 10.1016/C2013-0-01791-4]
[8]   SCINTILLATIONS FROM ORGANIC CRYSTALS - SPECIFIC FLUORESCENCE AND RELATIVE RESPONSE TO DIFFERENT RADIATIONS [J].
BIRKS, JB .
PROCEEDINGS OF THE PHYSICAL SOCIETY OF LONDON SECTION A, 1951, 64 (382) :874-877
[9]   The FLUKA Code: Developments and Challenges for High Energy and Medical Applications [J].
Boehlen, T. T. ;
Cerutti, F. ;
Chin, M. P. W. ;
Fosso, A. ;
Ferrari, A. ;
Ortega, P. G. ;
Mairani, A. ;
Sala, P. R. ;
Smirnov, G. ;
Vlachoudisl, V. .
NUCLEAR DATA SHEETS, 2014, 120 :211-214
[10]   Systematic evaluation of photodetector performance for plastic scintillation dosimetry [J].
Boivin, Jonathan ;
Beddar, Sam ;
Guillemette, Maxime ;
Beaulieu, Luc .
MEDICAL PHYSICS, 2015, 42 (11) :6211-6220