Clinical outcomes of totally implantable venous access port placement via the axillary vein in patients with head and neck malignancy

被引:19
作者
Hong, Sun [1 ]
Seo, Tae-Seok [1 ]
Song, Myung Gyu [1 ]
Seol, Hae-Young [1 ]
Suh, Sang Il [1 ]
Ryoo, In-Seon [1 ]
机构
[1] Korea Univ, Guro Hosp, Dept Radiol, Coll Med, 148 Gurodongro, Seoul 152703, South Korea
关键词
Totally implantable venous access port; axillary vein; central vein stenosis; head and neck malignancy; RISK-FACTORS; CATHETER; COMPLICATIONS; DEVICES; CANNULATION; ULTRASOUND; STENOSIS; SYSTEMS; RATES; RATIO;
D O I
10.1177/1129729818781270
中图分类号
R6 [外科学];
学科分类号
1002 ; 100210 ;
摘要
Purpose: To evaluate the clinical outcomes and complications of totally implantable venous access port implantation via the axillary vein in patients with head and neck malignancy. Materials and methods: A total of 176 totally implantable venous access ports were placed via the axillary vein in 171 patients with head and neck malignancy between May 2012 and June 2015. The patients included 133 men and 38 women, and the mean age was 58.8 years (range: 19-84 years). Medical records were retrospectively reviewed. Results: This study included a total of 93,237 totally implantable venous access port catheter-days (median 478 catheter-days, range: 13-1380 catheter-days). Of the 176 implanted totally implantable venous access port, complications developed in nine cases (5.1%), with the overall incidence of 0.097 events/1000 catheter-days. The complications were three central line-associated blood-stream infection cases, one case of keloid scar at the needling access site, and five cases of central vein stenosis or thrombosis on neck computed tomography images. The 133 cases for which neck computed tomography images were available had a total of 59,777 totally implantable venous access port catheter-days (median 399 catheter-days, range: 38-1207 catheter-days). On neck computed tomography evaluation, the incidence of central vein stenosis or thrombosis was 0.083 events/1000 catheter-days. Thrombosis developed in four cases, yielding an incidence of 0.067 events/1000 catheter-days. All four patients presented with thrombus in the axillary or subclavian vein. Stenosis occurred in one case yielding an incidence of 0.017 events/1000 catheter-days. One case was catheter-related brachiocephalic vein stenosis, and the other case was subclavian vein stenosis due to extrinsic compression by tumor progression. Of the nine complication cases, six underwent port removal. Conclusion: These data indicate that totally implantable venous access port implantation via the axillary vein in patients with head and neck malignancy is safe and feasible, with a low axillary vein access-related complication rate.
引用
收藏
页码:134 / 139
页数:6
相关论文
共 26 条
  • [1] Central vein stenosis: A nephrologist's perspective
    Agarwal, Anil K.
    Patel, Bhairavi M.
    Haddad, Nabil J.
    [J]. SEMINARS IN DIALYSIS, 2007, 20 (01) : 53 - 62
  • [2] COMPLICATIONS AND SIDE-EFFECTS ASSOCIATED WITH LARGE-BORE CATHETERS IN THE SUBCLAVIAN AND INTERNAL JUGULAR VEINS
    BAMBAUER, R
    INNIGER, R
    PIRRUNG, KJ
    SCHIEL, R
    DAHLEM, R
    [J]. ARTIFICIAL ORGANS, 1994, 18 (04) : 318 - 321
  • [3] Best choice of central venous insertion site for the prevention of catheter-related complications in adult patients who need cancer therapy: a randomized trial
    Biffi, R.
    Orsi, F.
    Pozzi, S.
    Pace, U.
    Bonomo, G.
    Monfardini, L.
    Della Vigna, P.
    Rotmensz, N.
    Radice, D.
    Zampino, M. G.
    Fazio, N.
    de Braud, F.
    Andreoni, B.
    Goldhirsch, A.
    [J]. ANNALS OF ONCOLOGY, 2009, 20 (05) : 935 - 940
  • [4] Totally implantable vascular access devices 30 years after the first procedure. What has changed and what is still unsolved?
    Biffi, Roberto
    Toro, Adriana
    Pozzi, Simonetta
    Di Carlo, Isidoro
    [J]. SUPPORTIVE CARE IN CANCER, 2014, 22 (06) : 1705 - 1714
  • [5] Catheter tip position as a risk factor for thrombosis associated with the use of subcutaneous infusion ports
    Caers, J
    Fontaine, C
    Vinh-Hung, V
    De Mey, J
    Ponnet, G
    Oost, C
    Lamote, J
    De Greve, J
    Van Camp, B
    Lacor, P
    [J]. SUPPORTIVE CARE IN CANCER, 2005, 13 (05) : 325 - 331
  • [6] Ultrasound for central venous cannulation: economic evaluation of cost-effectiveness
    Calvert, N
    Hind, D
    McWilliams, R
    Davidson, A
    Beverley, CA
    Thomas, SM
    [J]. ANAESTHESIA, 2004, 59 (11) : 1116 - 1120
  • [7] Increased Use of Percutaneous Technique for Totally Implantable Venous Access Devices. Is It Real Progress? A 27-Year Comprehensive Review on Early Complications
    Di Carlo, Isidoro
    Pulvirenti, Elia
    Mannino, Maurizio
    Toro, Adriana
    [J]. ANNALS OF SURGICAL ONCOLOGY, 2010, 17 (06) : 1649 - 1656
  • [8] Reasons for explantation of totally implantable access ports:: A multivariate analysis of 385 consecutive patients
    Fischer, Lars
    Knebel, Phillip
    Schroeder, Steffen
    Bruckner, Thomas
    Diener, Markus K.
    Hennes, Roland
    Buhl, Klaus
    Schmied, Bruno
    Seiler, Christoph M.
    [J]. ANNALS OF SURGICAL ONCOLOGY, 2008, 15 (04) : 1124 - 1129
  • [9] Systematic review: malfunction of totally implantable venous access devices in cancer patients
    Goossens, Godelieve Alice
    Stas, Marguerite
    Jerome, Martine
    Moons, Philip
    [J]. SUPPORTIVE CARE IN CANCER, 2011, 19 (07) : 883 - 898
  • [10] IMPLANTABLE CATHETER SYSTEM FOR LONG-TERM INTRAVENOUS CHEMOTHERAPY
    HALL, P
    CEDERMARK, B
    SWEDENBORG, J
    [J]. JOURNAL OF SURGICAL ONCOLOGY, 1989, 41 (01) : 39 - 41