Automated control of inspired oxygen in ventilated preterm infants: crossover physiological study

被引:42
作者
Lal, Mithilesh [1 ]
Tin, Win [1 ,2 ]
Sinha, Sunil [1 ,2 ]
机构
[1] James Cook Univ Hosp, Dept Neonatal Med, Middlesbrough TS4 3BW, Cleveland, England
[2] Univ Durham, Paediat & Neonatal Med, Durham, England
关键词
Closed loop; Hyperoxaemia; Hypoxaemia; Oxygen saturation; Premature infants; BIRTH-WEIGHT INFANTS; SATURATION; EPISODES; TARGET;
D O I
10.1111/apa.13137
中图分类号
R72 [儿科学];
学科分类号
100202 ;
摘要
AimTo evaluate the efficacy of automated control of the fraction of inspired oxygen (FiO(2)) in comparison with manual FiO(2) control in maintaining target pulse oxygen saturation (SpO(2)) range. MethodsCrossover physiological study involving preterm infants requiring mechanical ventilation and supplemental oxygen. Each infant was studied for two consecutive 12hours in a random sequence. Outcome measures included the proportion of time with SpO(2) within and outside the target range of 90-95%, extreme hypoxaemia (<80%) and hyperoxaemia (98%). ResultsComplete data set was available in 27 infants. The percentage of time (median, IQR) within the target range was higher during automated control 72.8 (58.8-82.6) compared to manual control 59.6 (49.3-73.3), p=0.031. Corresponding reduction in per cent time below the target range was 18.1 (12.7-23.6) versus 25.9 (17.8-30.7), p=0.028, and above the target range 4.8 (3-16) versus 10.1 (6.4-22.5), p=0.026. Median (IQR) per cent time spent with severe hypoxaemia (SpO(2)<80%) and severe hyperoxaemia (SpO(2)98%) was 1.3 (0.1-2.9) versus 3.2 (1.4-6.1), p=0.022 and 0.08 (0.05-0.36) versus 1.7 (0.7-6.8), p=0.001, respectively. Median number of manual adjustments of FiO(2) per 12hour was 0 and 63, respectively. ConclusionAutomated control of FiO(2) significantly improved compliance of oxygen saturation targeting and significantly reduced exposure to hypoxaemia as well as hyperoxaemia.
引用
收藏
页码:1084 / 1089
页数:6
相关论文
共 16 条
  • [1] Target Ranges of Oxygen Saturation in Extremely Preterm Infants.
    Carlo, Waldemar A.
    Finer, Neil N.
    Walsh, Michele C.
    Rich, Wade
    Gantz, Marie G.
    Laptook, Abbot R.
    Yoder, Bradley A.
    Faix, Roger G.
    Das, Abhik
    Poole, W. Kenneth
    Schibler, Kurt
    Newman, Nancy S.
    Ambalavanan, Namasivayam
    Frantz, Ivan D., III
    Piazza, Anthony J.
    Sanchez, Pablo J.
    Morris, Brenda H.
    Laroia, Nirupama
    Phelps, Dale L.
    Poindexter, Brenda B.
    Cotten, C. Michael
    Van Meurs, Krisa P.
    Duara, Shahnaz
    Narendran, Vivek
    Sood, Beena G.
    O'Shea, T. Michael
    Bell, Edward F.
    Ehrenkranz, Richard A.
    Watterberg, Kristi L.
    Higgins, Rosemary D.
    Jobe, A. H.
    Caplan, M. S.
    Oh, W.
    Hensman, A. M.
    Gingras, D.
    Barnett, S.
    Lillie, S.
    Francis, K.
    Andrews, D.
    Angela, K.
    Fanaroff, A. A.
    Siner, B. S.
    Zadell, A.
    DiFiore, J.
    Donovan, E. F.
    Bridges, K.
    Alexander, B.
    Grisby, C.
    Mersmann, M. W.
    Mincey, H. L.
    [J]. NEW ENGLAND JOURNAL OF MEDICINE, 2010, 362 (21) : 1959 - 1969
  • [2] Closed-loop controlled inspired oxygen concentration for mechanically ventilated very low birth weight infants with frequent episodes of hypoxemia
    Claure, N
    Gerhardt, T
    Everett, R
    Musante, G
    Herrera, C
    Bancalari, E
    [J]. PEDIATRICS, 2001, 107 (05) : 1120 - 1124
  • [3] Multicenter Crossover Study of Automated Control of Inspired Oxygen in Ventilated Preterm Infants
    Claure, Nelson
    Bancalari, Eduardo
    D'Ugard, Carmen
    Nelin, Leif
    Stein, Melanie
    Ramanathan, Rangasamy
    Hernandez, Richard
    Donn, Steven M.
    Becker, Michael
    Bachman, Thomas
    [J]. PEDIATRICS, 2011, 127 (01) : E76 - E83
  • [4] Automated Adjustment of Inspired Oxygen in Preterm Infants with Frequent Fluctuations in Oxygenation: A Pilot Clinical Trial
    Claure, Nelson
    D'Ugard, Carmen
    Bancalari, Eduardo
    [J]. JOURNAL OF PEDIATRICS, 2009, 155 (05) : 640 - U75
  • [5] Di FJ, 2012, J PEDIAT, V161, P1047
  • [6] Achieved versus intended pulse oximeter saturation in infants born less than 28 weeks' gestation: The AVIOx study
    Hagadorn, James I.
    Furey, Anne M.
    Nghiem, Tuyet-Hang
    Schmid, Christopher H.
    Phelps, Dale L.
    Pillers, De-Ann M.
    Cole, Cynthia H.
    [J]. PEDIATRICS, 2006, 118 (04) : 1574 - 1582
  • [7] Jain D, 2014, PAS
  • [8] Oxygen targeting in preterm infants using the Masimo SET Radical pulse oximeter
    Johnston, Ewen D.
    Boyle, Breidge
    Juszczak, Ed
    King, Andy
    Brocklehurst, Peter
    Stenson, Ben J.
    [J]. ARCHIVES OF DISEASE IN CHILDHOOD-FETAL AND NEONATAL EDITION, 2011, 96 (06): : F429 - F433
  • [9] Pulse oximetry in very low birth weight infants: can oxygen saturation be maintained in the desired range?
    Laptook, A. R.
    Salhab, W.
    Allen, J.
    Saha, S.
    Walsh, M.
    [J]. JOURNAL OF PERINATOLOGY, 2006, 26 (06) : 337 - 341
  • [10] EPISODES OF SPONTANEOUS DESATURATIONS IN INFANTS WITH CHRONIC LUNG-DISEASE AT 2 DIFFERENT LEVELS OF OXYGENATION