Knee joint laxity is restored in a bi-cruciate retaining TKA-design

被引:17
作者
Arnout, N. [1 ,2 ]
Victor, J. [1 ,2 ]
Vermue, H. [1 ,2 ]
Pringels, L. [1 ,2 ]
Bellemans, J. [3 ]
Verstraete, M. A. [2 ]
机构
[1] Univ Hosp Ghent, Corneel Heymanslaan 10, B-9000 Ghent, Belgium
[2] Univ Ghent, Ghent, Belgium
[3] ZOL, Schiepse Bos 6, B-3600 Genk, Belgium
关键词
Cruciate ligaments; Total knee arthroplasty; Knee laxity; PATIENT-REPORTED OUTCOMES; TIBIOFEMORAL JOINT; ARTHROPLASTY; REPLACEMENT; SURVIVORSHIP; SATISFACTION; BIOMECHANICS; KINEMATICS; YOUNGER;
D O I
10.1007/s00167-019-05639-4
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Purpose The goal is to evaluate the passive stability of a bicruciate retaining, cruciate retaining and bicruciate substituting TKA design in relation to the native knee stability in terms of the laxity envelope. A bicruciate retaining knee prosthesis was hypothesized to offer a closer to normal knee stability in vitro. Methods Fourteen cadaveric knee specimens have been tested under passive conditions with and without external loads, involving a varus/valgus and an external/internal rotational torque, distraction/compression force and an anteroposterior shear force. Subsequently, the native knee, bicruciate retaining, cruciate retaining and finally a bicruciate substituting total knee arthroplasty were tested. Results Through the range of motion, the width of the varus/valgus and internal/external laxity envelope for the native knee and the bicruciate retaining knee were almost equivalent, whereas the cruciate retaining and the bicruciate substituting knee displayed less laxity and more joint distraction. In all prosthetic knees, an equal anteroposterior laxity was seen for the lateral and medial side whereas in the native knee, a difference in laxity was seen between the stable medial side and the more mobile lateral side. Conclusion Bicruciate retaining knee prostheses can restore normal laxity and thus have the potential to offer more normal knee function. Restoration of natural peri-articular soft-tissue tension is clinically important because of its obvious effects on joint stability and range of motion. Furthermore, the results of this study could help to establish the ideal ligament tension and laxity in more conventional implants by approaching the normal values for passive knee evaluation as presented here.
引用
收藏
页码:2863 / 2871
页数:9
相关论文
共 34 条
[1]  
[Anonymous], 10 ANN REPORT 2013 N
[2]   Post-cam mechanics and tibiofemoral kinematics: a dynamic in vitro analysis of eight posterior-stabilized total knee designs [J].
Arnout, N. ;
Vanlommel, L. ;
Vanlommel, J. ;
Luyckx, J. P. ;
Labey, L. ;
Innocenti, B. ;
Victor, J. ;
Bellemans, J. .
KNEE SURGERY SPORTS TRAUMATOLOGY ARTHROSCOPY, 2015, 23 (11) :3343-3353
[3]   THE ENVELOPE OF PASSIVE KNEE-JOINT MOTION [J].
BLANKEVOORT, L ;
HUISKES, R ;
DELANGE, A .
JOURNAL OF BIOMECHANICS, 1988, 21 (09) :705-+
[4]   LIGAMENTOUS RESTRAINTS TO ANTERIOR-POSTERIOR DRAWER IN THE HUMAN KNEE - BIOMECHANICAL STUDY [J].
BUTLER, DL ;
NOYES, FR ;
GROOD, ES .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1980, 62 (02) :259-270
[5]   Collateral ligament strains during knee joint laxity evaluation before and after TKA [J].
Delport, Hendrik ;
Labey, Luc ;
De Corte, Ronny ;
Innocenti, Bernardo ;
Sloten, Jos Vander ;
Bellemans, Johan .
CLINICAL BIOMECHANICS, 2013, 28 (07) :777-782
[6]  
Font-Rodriguez DE, 1997, CLIN ORTHOP RELAT R, P79
[7]   Patient-Reported Outcomes, Quality of Life, and Satisfaction Rates in Young Patients Aged 50 Years or Younger After Total Knee Arthroplasty [J].
Goh, Graham Seow-Hng ;
Liow, Ming Han Lincoln ;
Bin Abd Razak, Hamid Rahmatullah ;
Tay, Darren Keng-Jin ;
Lo, Ngai-Nung ;
Yeo, Seng-Jin .
JOURNAL OF ARTHROPLASTY, 2017, 32 (02) :419-425
[8]   A JOINT COORDINATE SYSTEM FOR THE CLINICAL DESCRIPTION OF 3-DIMENSIONAL MOTIONS - APPLICATION TO THE KNEE [J].
GROOD, ES ;
SUNTAY, WJ .
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 1983, 105 (02) :136-144
[9]   Anteroposterior Laxity After Bicruciate-Retaining Total Knee Arthroplasty Is Closer to the Native Knee Than ACL-Resecting TKA: A Biomechanical Cadaver Study [J].
Halewood, Camilla ;
Traynor, Alison ;
Bellemans, Johan ;
Victor, Jan ;
Amis, Andrew A. .
JOURNAL OF ARTHROPLASTY, 2015, 30 (12) :2315-2319
[10]   Native rotational knee kinematics are lost in bicruciate-retaining total knee arthroplasty when the tibial component is replaced [J].
Hamada, Daisuke ;
Wada, Keizo ;
Takasago, Tomoya ;
Goto, Tomohiro ;
Nitta, Akihiro ;
Higashino, Kosaku ;
Fukui, Yoshihiro ;
Sairyo, Koichi .
KNEE SURGERY SPORTS TRAUMATOLOGY ARTHROSCOPY, 2018, 26 (11) :3249-3256