Structural Valve Deterioration in a Starr-Edwards Mitral Caged-Disk Valve Prosthesis

被引:3
作者
Aoyagi, Shigeaki [1 ]
Tayama, Kei-ichiro [1 ]
Okazaki, Teiji [1 ]
Shintani, Yusuke [1 ]
Kono, Michitaka [1 ]
Wada, Kumiko [1 ]
Kosuga, Ken-ichi [1 ]
Mori, Ryusuke [2 ]
Tanaka, Hiroyuki [2 ]
机构
[1] Munakata Suikokai Gen Hosp, Dept Cardiovasc Surg, Fukutsu, Japan
[2] Kurume Univ, Sch Med, Dept Surg, Kurume, Fukuoka 830, Japan
关键词
Caged-disk valve; Disk fracture; Disk wear; Prosthetic heart valve; Structural valve deterioration; REPLACEMENT; OUTCOMES;
D O I
10.1253/circj.CJ-12-0906
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: The durability of the Starr-Edwards (SE) mitral caged-disk valve, model 6520, is not clearly known, and structural valve deterioration in the SE disk valve is very rare. Methods and Results: Replacement of the SE mitral disk valve was performed in 7 patients 23-40 years after implantation. Macroscopic examination of the removed disk valves showed no structural abnormalities in 3 patients, in whom the disk valves were removed at <26 years after implantation. Localized disk wear was found at the sites where the disk abutted the struts of the cage, in disk valves excised >36 years after implantation in 4 patients. Disk fracture, a longitudinal split in the disk along its circumference at the site of incorporation of the titanium ring, was detected in the valves removed 36 and 40 years after implantation, respectively, and many cracks were also observed on the outflow aspect of the disk removed 40 years after implantation. Conclusions: Disk fracture and localized disk wear were found in the SE mitral disk valves implanted >36 years previously. The present results suggest that SE mitral caged-disk valves implanted >20 years previously should be carefully followed up, and that those implanted >30 years previously should be electively replaced with modern prosthetic valves. (Circ J 2013; 77: 105-108)
引用
收藏
页码:105 / 108
页数:4
相关论文
共 10 条
[1]  
Aokage K, 1985, JPN J ARTIF ORGANS, V14, P1393
[2]   Long-term survivors after valve replacement with a Starr-Edwards mitral disk valve prosthesis [J].
Aoyagi, S ;
Fukunaga, S ;
Tayama, E ;
Arinaga, K ;
Oda, T ;
Kosuga, T .
ARTIFICIAL ORGANS, 2006, 30 (06) :484-487
[3]  
Aoyagi S, 2011, J HEART VALVE DIS, V20, P474
[4]  
Kawachi Y, 1986, JPN J ARTIF ORGANS, V15, P794
[5]  
Kremkua EL, 1972, CIRCULATION S2, p[45, 178]
[6]  
Malouf JF, 2001, J HEART VALVE DIS, V10, P136
[7]  
Nakano S, 1978, JPN J THORAC SURG, V31, P440
[8]   Mid-Term Outcomes After Aortic Valve Replacement With the 17-mm St. Jude Medical Regent Valve [J].
Okamura, Homare ;
Yamaguchi, Atsushi ;
Nagano, Hiroshi ;
Itoh, Satoshi ;
Morita, Hideki ;
Naito, Kazuhiro ;
Yuri, Koichi ;
Adachi, Hideo .
CIRCULATION JOURNAL, 2012, 76 (02) :365-371
[9]   Clinical Outcomes and Hemodynamics of the 19-mm Perimount Magna Bioprosthesis in an Aortic Position - Comparison With the 19-mm Medtronic Mosaic Ultra Valve [J].
Okamura, Homare ;
Yamaguchi, Atsushi ;
Yoshizaki, Takamichi ;
Nagano, Hiroshi ;
Itoh, Satoshi ;
Morita, Hideki ;
Naito, Kazuhiro ;
Yuri, Koichi ;
Adachi, Hideo .
CIRCULATION JOURNAL, 2012, 76 (01) :102-108
[10]  
Starr A., 1969, Prosthetic heart valves, P148