The Accuracy of Dynamic Contour Tonometry Over Soft Contact Lenses

被引:2
作者
Gogniat, Fabrice [1 ]
Steinegger, Daniela [1 ]
Nosch, Daniela S. [1 ]
Joos, Roland [1 ]
Goldschmidt, Michael [1 ]
机构
[1] Univ Appl Sci NW Switzerland, Inst Optometry, CH-4600 Olten, Switzerland
关键词
dynamic contour tonometer; cornea; intraocular pressure; ocular pulse amplitude; topical anesthesia; hydrogel; silicone hydrogel; contact lens; OCULAR PULSE AMPLITUDE; NORMAL-TENSION GLAUCOMA; OPEN-ANGLE GLAUCOMA; PRESSURE; HYPERTENSION;
D O I
10.1097/OPX.0b013e31827cdaa7
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose. Dynamic contour tonometry (DCT) has been shown to measure the intraocular pressure (IOP) independently of corneal thickness. This study aimed to investigate if DCT remains accurate when the IOP measurement is taken over soft contact lenses (CLs) of different thicknesses and material characteristics. Methods. This was a prospective clinical study that included 42 patients. Subject age was 22 to 59 years (26.5 +/- 6.3 years). Intraocular pressure and ocular pulse amplitude (OPA) measurements were taken under topical anesthesia without CLs and over various daily disposable CLs with -0.50, +5.00, and -5.00 diopters (D) in hydrogel (Nelfilcon A) and in silicone hydrogel (Narafilcon A) materials. Results. No statistically significant differences were found when comparing the IOP measurements obtained using either of the different CL powers of -0.50 or -5.00 D, irrespective of which CL material was being used. However, the difference of 0.62 mm Hg observed when the Nelfilcon A with a power of +5.00 D was used turned out to be highly statistically significant (p = 0.0002), whereas the Narafilcon A with the same power of +5.00 D, with a small difference of -0.16 mm Hg, was not. Regarding OPA measurements, no significant differences were found between measurements with and without CL neither for different materials nor for change in dioptrical power (F = 0, p = 1.000). Conclusions. This study showed good reliability of IOP and OPA measurements over CLs with varying thickness profiles and different soft materials when using the DCT. Only a small but statistically significant difference of 0.62 mm Hg was found for the IOP measurement with the hydrogel CL of +5.00 D compared with "no CL." (Optom Vis Sci 2013;90:125-130)
引用
收藏
页码:125 / 130
页数:6
相关论文
共 25 条
[1]  
Almubrad TM, 2007, CLIN OPHTHALMOL, V1, P305
[2]  
[Anonymous], 2004, MODERN APPL STAT S
[3]  
[Anonymous], 2011, R LANG ENV STAT COMP
[4]   Ocular hypertension and central corneal thickness [J].
Argus, WA .
OPHTHALMOLOGY, 1995, 102 (12) :1810-1812
[5]   A study comparing ocular pressure pulse and ocular fundus pulse in dependence of axial eye length and ocular volume [J].
Berisha, Fatmire ;
Findl, Oliver ;
Lasta, Michael ;
Kiss, Barbara ;
Schmetterer, Leopold .
ACTA OPHTHALMOLOGICA, 2010, 88 (07) :766-772
[6]   Dynamic contour tonometry in comparison to intracameral IOP measurements [J].
Boehm, Andreas G. ;
Weber, Anja ;
Pillunat, Lutz E. ;
Koch, Rainer ;
Spoerl, Eberhard .
INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2008, 49 (06) :2472-2477
[7]   Measurement of ocular blood flow [J].
Cioffi, GA ;
Alm, A .
JOURNAL OF GLAUCOMA, 2001, 10 (05) :S62-S64
[8]  
EHLERS N, 1975, ACTA OPHTHALMOL, V53, P34
[9]   Pulsatile ocular blood flow investigation in asymmetric normal tension glaucoma and normal subjects [J].
Fontana, L ;
Poinoosawmy, D ;
Bunce, CV ;
O'Brien, C ;
Hitchings, RA .
BRITISH JOURNAL OF OPHTHALMOLOGY, 1998, 82 (07) :731-736
[10]  
GOLDMANN H, 1957, Ophthalmologica, V134, P221