Glucose and urea biosensors based on all solid-state PVC-NH2 membrane electrodes

被引:57
作者
Tinkilic, N [1 ]
Cubuk, O [1 ]
Isildak, I [1 ]
机构
[1] Ondokuz Mayis Univ, Fac Sci, Dept Chem, TR-55139 Kurupelit, Turkey
关键词
potentiometric biosensors; urea biosensor; glucose biosensor; all solid-state contact; PVC-NH2 membrane electrode;
D O I
10.1016/S0003-2670(01)01441-6
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Miniaturized urea and glucose sensors prepared by immobilization of glucose oxidase or urease directly onto all solid-state contact PVC-NH2 membrane ammonium and hydrogen ion selective electrodes are described. The resulting biosensing membranes function equivalently to normal PVC membranes in terms of potentiometric response properties. The most important features of the glucose and urea sensors were high sensitivity, long life-time, easily built at a low cost, microconstruction and short response time. The characteristics of the glucose and urea sensors were examined in several buffer solutions at different concentrations and pH values. The influence of immobilization conditions on the dynamic response properties and life-time of the electrodes was studied. Under optimal conditions, the urea electrode showed a linear response between 5 x 10(-2) and 5 x 10(-4) M urea, while the glucose electrode showed a linear response between 5 x 10(-2) and 1 X 10(-4) M glucose. (C) 2002 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:29 / 34
页数:6
相关论文
共 13 条
[1]   RESPONSE CHARACTERISTICS OF CONDUCTIVE POLYMER COMPOSITE SUBSTRATE ALL-SOLID-STATE POLY(VINYL CHLORIDE) MATRIX MEMBRANE ION-SELECTIVE ELECTRODES IN AERATED AND NITROGEN-SATURATED SOLUTIONS [J].
ALEGRET, S ;
FLORIDO, A .
ANALYST, 1991, 116 (05) :473-476
[2]   MINIATURIZED DISPOSABLE BIOSENSORS [J].
BILITEWSKI, U ;
CHEMNITIUS, GC ;
RUGER, P ;
SCHMID, RD .
SENSORS AND ACTUATORS B-CHEMICAL, 1992, 7 (1-3) :351-355
[3]   ENZYME UREA BIOSENSOR BASED ON A MODIFIED POTENTIOMETRIC PVC-NONACTIN MEMBRANE-ELECTRODE FOR ASSAY OF UREA IN BLOOD [J].
BUTT, SB ;
CAMMANN, K .
ANALYTICAL LETTERS, 1992, 25 (09) :1597-1615
[4]   PLANAR MICRO SENSORS FOR IN-VIVO MYOCARDIAL PH MEASUREMENTS [J].
COSOFRET, VV ;
LINDNER, E ;
JOHNSON, TA ;
NEUMAN, MR .
TALANTA, 1994, 41 (06) :931-938
[5]  
GUILBAULT GG, 1977, HDB ENZYMATIC METHOD, pCH2
[6]  
IUPAC, 1979, PURE APPL CHEM, V48, P127, DOI DOI 10.1351/PAC197648010127
[7]   COMPARISON OF PH-MEMBRANE ENZYME ELECTRODES FOR UREA WITH COVALENTLY BOUND ENZYME [J].
KONCKI, R ;
KOPCZEWSKA, E ;
GLAB, S .
ANALYTICAL LETTERS, 1994, 27 (03) :475-486
[8]  
MA SC, 1990, MIKROCHIM ACTA, V1, P197
[9]   RESPONSE PROPERTIES OF ION-SELECTIVE POLYMERIC MEMBRANE ELECTRODES PREPARED WITH AMINATED AND CARBOXYLATED POLYVINYL-CHLORIDE) [J].
MA, SC ;
CHANIOTAKIS, NA ;
MEYERHOFF, ME .
ANALYTICAL CHEMISTRY, 1988, 60 (20) :2293-2299
[10]  
Magalhaes JMCS, 1997, ANAL CHIM ACTA, V355, P241