Incidence, mechanisms, and clinical impact of inappropriate shocks in patients with a subcutaneous defibrillator

被引:10
|
作者
Rudic, Boris [1 ,2 ]
Tueluemen, Erol [1 ,2 ]
Fastenrath, Fabian [1 ,2 ]
Roeger, Susanne [1 ,2 ]
Goranova, Diana [1 ,2 ]
Akin, Ibrahim [1 ,2 ]
Borggrefe, Martin [1 ,2 ]
Kuschyk, Juergen [1 ,2 ]
机构
[1] Univ Med Ctr Mannheim, Dept Med Cardiol 1, D-68167 Mannheim, Germany
[2] German Ctr Cardiovasc Res DZHK, Partner Site Heidelberg Mannheim, Mannheim, Germany
来源
EUROPACE | 2020年 / 22卷 / 05期
关键词
Subcutaneous implantable cardioverter-defibrillator; Ventricular tachyarrhythmias; Inappropriate shock; Programming; Sensing vector; IMPLANTABLE CARDIOVERTER-DEFIBRILLATOR; PREVENTION; THERAPY;
D O I
10.1093/europace/euaa026
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Aims Inappropriate shocks (IAS) remain a challenge for patients and physicians after implantation of the subcutaneous implantable cardioverter-defibrfflator (S-ICD). The aims were to assess and characterize different patterns of IAS. Methods and results Two hundred and thirty-nine patients were implanted with an S-ICD between 2010 and 2018 for primary and secondary prevention. Follow-up data of at least 6 months were analysed. During a mean follow-up of 34.9 +/- 16.0 months, a total of 73 shocks occurred in 38 patients (6%). Forty-three (59%) shocks were considered appropriate due to ventricular tachycardia/ventricular fibrillation, while 30 (41%) were inappropriate and occurred in 19 patients (8%). Myopotentials/noise was the most frequent cause of inappropriate shocks (n = 8), followed by T-wave oversensing (n = 6) and undersensing of the QRS, resulting in adaptation of the automatic gain control and inappropriate shock (n = 5). Seventy-four percent of all IAS occurred on the primary vector, while no IAS occurred on the alternate vector. In seven of eight patients (88%), IAS related to myopotentials have occurred on the primary sensing vector. Multivariate analysis identified taller patients, primary sensing vector and first-generation SICD device as predictors for IAS. SMART pass effectively reduced the occurrence of IAS in the second-generation S-ICD system. Conclusions Inappropriate therapies are less frequently observed on the alternate vector. The primary vector seems to be unfavourable with regard to oversensing caused by myopotentials. Inappropriate shocks were associated with an increased rate of rehospitalization but not mortality. These observations have implications for the prevention of inappropriate S-ICD shocks.
引用
收藏
页码:761 / 768
页数:8
相关论文
共 50 条
  • [1] Incidence and causes of inappropriate shocks in patients with implantable cardioverter/defibrillator
    Schukro, C.
    Siebermair, J.
    Stix, G.
    Kastner, J.
    Pezawas, T.
    Wolzt, M.
    Schmidinger, H.
    EUROPEAN HEART JOURNAL, 2008, 29 : 645 - 645
  • [2] Inappropriate Shocks due to Subcutaneous Air in a Patient With a Subcutaneous Cardiac Defibrillator
    Zipse, Matthew M.
    Sauer, William H.
    Varosy, Paul D.
    Aleong, Ryan G.
    Nguyen, Duy T.
    CIRCULATION-ARRHYTHMIA AND ELECTROPHYSIOLOGY, 2014, 7 (04): : 768 - 770
  • [3] Predictors and Clinical Impact of Inappropriate Implantable Cardioverter-Defibrillator Shocks in Korean Patients
    Yang, Jeong Hoon
    Byeon, Kyeongmin
    Yim, Hye Ran
    Park, Jung Wae
    Park, Seung-Jung
    Huh, June
    Kim, June Soo
    On, Young Keun
    JOURNAL OF KOREAN MEDICAL SCIENCE, 2012, 27 (06) : 619 - 624
  • [4] Inappropriate Implantable Cardioverter-Defibrillator Shocks Incidence, Predictors, and Impact on Mortality
    van Rees, Johannes B.
    Borleffs, C. Jan Willem
    de Bie, Mihaly K.
    Stijnen, Theo
    van Erven, Lieselot
    Bax, Jeroen J.
    Schalij, Martin J.
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2011, 57 (05) : 556 - 562
  • [5] Inappropriate shocks in the subcutaneous ICD: Incidence, predictors and management
    Nordkamp, Louise R. A. Olde
    Brouwer, Tom F.
    Barr, Craig
    Theuns, Dominic A. M. J.
    Boersma, Lucas V. A.
    Johansen, Jens B.
    Neuzil, Petr
    Wilde, Arthur A. M.
    Carter, Nathan
    Husby, Michael
    Lambiase, Pier D.
    Knops, Reinoud E.
    INTERNATIONAL JOURNAL OF CARDIOLOGY, 2015, 195 : 126 - 133
  • [6] Prognostic impact of inappropriate defibrillator shocks in a population cohort
    Deyell, Marc William
    Qi, Anzhen
    Chakrabarti, Santabhanu
    Yeung-Lai-Wah, John Ah-Fan
    Tung, Stanley
    Khoo, Clarence
    Bennett, Mathew T.
    Qian, Hong
    Kerr, Charles R.
    HEART, 2013, 99 (17) : 1250 - 1255
  • [7] Inappropriate defibrillator shocks and mortality
    Liew, Reginald
    HEART, 2013, 99 (17) : 1223 - 1224
  • [8] Inappropriate Implantable Defibrillator Shocks
    Raitt, Merritt H.
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2013, 62 (15) : 1351 - 1352
  • [9] Revision of subcutaneous implantable cardioverter defibrillator to intermuscular pocket to prevent inappropriate shocks
    Madjarov, Jeko M.
    Madzharov, Svetozar
    Abedinov, Filip
    Chappuis, Stefanie
    Fedor, John M.
    CLINICAL CASE REPORTS, 2020, 8 (06): : 1081 - 1085
  • [10] Incidence of Appropriate and Inappropriate Shocks in Implantable Cardioverter-Defibrillator Patients With Obstructive Sleep Apnea
    Naksuk, Niyada
    Selcuk, Adabag
    JOURNAL OF CARDIAC FAILURE, 2012, 18 (08) : S54 - S54