Improving spot-scanning proton therapy patient specific quality assurance with HPlusQA, a second-check dose calculation engine

被引:26
|
作者
Mackin, Dennis [1 ]
Li, Yupeng [1 ]
Taylor, Michael B. [1 ]
Kerr, Matthew [1 ]
Holmes, Charles [1 ]
Sahoo, Narayan [1 ]
Poenisch, Falk [1 ]
Li, Heng [1 ]
Lii, Jim [1 ]
Amos, Richard [1 ]
Wu, Richard [1 ]
Suzuki, Kazumichi [1 ]
Gillin, Michael T. [1 ]
Zhu, X. Ronald [1 ]
Zhang, Xiaodong [1 ]
机构
[1] Univ Texas MD Anderson Canc Ctr, Dept Radiat Phys, Houston, TX 77030 USA
关键词
spot scanning; proton therapy; IMPT; gamma index; patient quality assurance; TREATMENT PLANNING SYSTEM; ANDERSON CANCER CENTER; IMRT; VERIFICATION; BEAMS; HOUSTON; QA;
D O I
10.1118/1.4828775
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: The purpose of this study was to validate the use of HPlusQA, spot-scanning proton therapy (SSPT) dose calculation software developed at The University of Texas MD Anderson Cancer Center, as second-check dose calculation software for patient-specific quality assurance (PSQA). The authors also showed how HPlusQA can be used within the current PSQA framework. Methods: The authors compared the dose calculations of HPlusQA and the Eclipse treatment planning system with 106 planar dose measurements made as part of PSQA. To determine the relative performance and the degree of correlation between HPlusQA and Eclipse, the authors compared calculated with measured point doses. Then, to determine how well HPlusQA can predict when the comparisons between Eclipse calculations and the measured dose will exceed tolerance levels, the authors compared gamma index scores for HPlusQA versus Eclipse with those of measured doses versus Eclipse. The authors introduce the alpha beta gamma transformation as a way to more easily compare gamma scores. Results: The authors compared measured and calculated dose planes using the relative depth, z/R x 100%, where z is the depth of the measurement and R is the proton beam range. For relative depths than less than 80%, both Eclipse and HPlusQA calculations were within 2 cGy of dose measurements on average. When the relative depth was greater than 80%, the agreement between the calculations and measurements fell to 4 cGy. For relative depths less than 10%, the Eclipse and HPlusQA dose discrepancies showed a negative correlation, -0.21. Otherwise, the correlation between the dose discrepancies was positive and as large as 0.6. For the dose planes in this study, HPlusQA correctly predicted when Eclipse had and had not calculated the dose to within tolerance 92% and 79% of the time, respectively. In 4 of 106 cases, HPlusQA failed to predict when the comparison between measurement and Eclipse's calculation had exceeded the tolerance levels of 3% for dose and 3 mm for distance-to-agreement. Conclusions: The authors found HPlusQA to be reasonably effective (79% +/- 10%) in determining when the comparison between measured dose planes and the dose planes calculated by the Eclipse treatment planning system had exceeded the acceptable tolerance levels. When used as described in this study, HPlusQA can reduce the need for patient specific quality assurance measurements by 64%. The authors believe that the use of HPlusQA as a dose calculation second check can increase the efficiency and effectiveness of the QA process. (C) 2013 American Association of Physicists in Medicine.
引用
收藏
页数:10
相关论文
共 36 条
  • [31] Assessing the radiation-induced second cancer risk in proton therapy for pediatric brain tumors: the impact of employing a patient-specific aperture in pencil beam scanning
    Geng, Changran
    Moteabbed, Maryam
    Xie, Yunhe
    Schuemann, Jan
    Yock, Torunn
    Paganetti, Harald
    PHYSICS IN MEDICINE AND BIOLOGY, 2016, 61 (01) : 12 - 22
  • [32] Effect of dose grid resolution on the results of patient-specific quality assurance for intensity-modulated radiation therapy and volumetric modulated arc therapy
    Chun, M.
    Kim, J., I
    Oh, G. H.
    Wu, H. G.
    Park, J. M.
    INTERNATIONAL JOURNAL OF RADIATION RESEARCH, 2020, 18 (03): : 521 - 530
  • [33] Correlation between patient-specific quality assurance in volumetric modulated arc therapy and 2D dose image features
    Guo, Yixiao
    Hu, Jinyan
    Li, Yang
    Ran, Juntao
    Cai, Hongyi
    SCIENTIFIC REPORTS, 2023, 13 (01)
  • [34] Enhancing Patient-specific Quality Assurance in Carbon-ion Radiation Therapy: Recalculating Delivered Dose Distribution Using Log Data
    Yun, Yongdo
    Han, Min cheol
    Kim, Changhwan
    Hong, Chae-Seon
    Kim, Dong wook
    Lee, Ho
    Kim, Hojin
    Han, Soorim
    Souda, Hikaru
    Lee, Sung hyun
    Miyasaka, Yuya
    Iwai, Takeo
    Kim, Jin sung
    IN VIVO, 2025, 39 (02): : 1086 - 1093
  • [35] Patient-specific quality assurance for proton depth dose distribution using a multi-layer ionization chamber in a single-ring wobbling method
    Kato, Takahiro
    Arai, Kazuhiro
    Sagara, Tatsuhiko
    Kato, Ryohei
    Yamazaki, Yuhei
    Oyama, Sho
    RADIOLOGICAL PHYSICS AND TECHNOLOGY, 2019, 12 (03) : 305 - 311
  • [36] Influence of beam efficiency through the patient-specific collimator on secondary neutron dose equivalent in double scattering and uniform scanning modes of proton therapy
    Hecksel, D.
    Anferov, V.
    Fitzek, M.
    Shahnazi, K.
    MEDICAL PHYSICS, 2010, 37 (06) : 2910 - 2917