Improving spot-scanning proton therapy patient specific quality assurance with HPlusQA, a second-check dose calculation engine

被引:26
|
作者
Mackin, Dennis [1 ]
Li, Yupeng [1 ]
Taylor, Michael B. [1 ]
Kerr, Matthew [1 ]
Holmes, Charles [1 ]
Sahoo, Narayan [1 ]
Poenisch, Falk [1 ]
Li, Heng [1 ]
Lii, Jim [1 ]
Amos, Richard [1 ]
Wu, Richard [1 ]
Suzuki, Kazumichi [1 ]
Gillin, Michael T. [1 ]
Zhu, X. Ronald [1 ]
Zhang, Xiaodong [1 ]
机构
[1] Univ Texas MD Anderson Canc Ctr, Dept Radiat Phys, Houston, TX 77030 USA
关键词
spot scanning; proton therapy; IMPT; gamma index; patient quality assurance; TREATMENT PLANNING SYSTEM; ANDERSON CANCER CENTER; IMRT; VERIFICATION; BEAMS; HOUSTON; QA;
D O I
10.1118/1.4828775
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: The purpose of this study was to validate the use of HPlusQA, spot-scanning proton therapy (SSPT) dose calculation software developed at The University of Texas MD Anderson Cancer Center, as second-check dose calculation software for patient-specific quality assurance (PSQA). The authors also showed how HPlusQA can be used within the current PSQA framework. Methods: The authors compared the dose calculations of HPlusQA and the Eclipse treatment planning system with 106 planar dose measurements made as part of PSQA. To determine the relative performance and the degree of correlation between HPlusQA and Eclipse, the authors compared calculated with measured point doses. Then, to determine how well HPlusQA can predict when the comparisons between Eclipse calculations and the measured dose will exceed tolerance levels, the authors compared gamma index scores for HPlusQA versus Eclipse with those of measured doses versus Eclipse. The authors introduce the alpha beta gamma transformation as a way to more easily compare gamma scores. Results: The authors compared measured and calculated dose planes using the relative depth, z/R x 100%, where z is the depth of the measurement and R is the proton beam range. For relative depths than less than 80%, both Eclipse and HPlusQA calculations were within 2 cGy of dose measurements on average. When the relative depth was greater than 80%, the agreement between the calculations and measurements fell to 4 cGy. For relative depths less than 10%, the Eclipse and HPlusQA dose discrepancies showed a negative correlation, -0.21. Otherwise, the correlation between the dose discrepancies was positive and as large as 0.6. For the dose planes in this study, HPlusQA correctly predicted when Eclipse had and had not calculated the dose to within tolerance 92% and 79% of the time, respectively. In 4 of 106 cases, HPlusQA failed to predict when the comparison between measurement and Eclipse's calculation had exceeded the tolerance levels of 3% for dose and 3 mm for distance-to-agreement. Conclusions: The authors found HPlusQA to be reasonably effective (79% +/- 10%) in determining when the comparison between measured dose planes and the dose planes calculated by the Eclipse treatment planning system had exceeded the acceptable tolerance levels. When used as described in this study, HPlusQA can reduce the need for patient specific quality assurance measurements by 64%. The authors believe that the use of HPlusQA as a dose calculation second check can increase the efficiency and effectiveness of the QA process. (C) 2013 American Association of Physicists in Medicine.
引用
收藏
页数:10
相关论文
共 36 条
  • [21] Clinically Applicable Monte Carlo-based Biological Dose Optimization for the Treatment of Head and Neck Cancers With Spot-Scanning Proton Therapy
    Tseung, Hok Seum Wan Chan
    Ma, Jiasen
    Kreofsky, Cole R.
    Ma, Daniel J.
    Beltran, Chris
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2016, 95 (05): : 1535 - 1543
  • [22] Implication of spot position error on plan quality and patient safety in pencil-beam-scanning proton therapy
    Yu, Juan
    Beltran, Chris J.
    Herman, Michael G.
    MEDICAL PHYSICS, 2014, 41 (08) : 96 - 103
  • [23] Transitioning from measurement-based to combined patient-specific quality assurance for intensity-modulated proton therapy
    Chen, Mei
    Yepes, Pablo
    Hojo, Yoshifumi
    Poenisch, Falk
    Li, Yupeng
    Chen, Jiayi
    Xu, Cheng
    He, Xiaodong
    Gunn, G. Brandon
    Frank, Steven J.
    Sahoo, Narayan
    Li, Heng
    Zhu, Xiaorong Ronald
    Zhang, Xiaodong
    BRITISH JOURNAL OF RADIOLOGY, 2020, 93 (1107)
  • [24] Dosimetric validation of a GPU-based dose engine for a fast in silico patient-specific quality assurance program in light ion beam therapy
    Magro, Giuseppe
    Fassi, Martina
    Mirandola, Alfredo
    Rossi, Eleonora
    Molinelli, Silvia
    Russo, Stefania
    Bazani, Alessia
    Vai, Alessandro
    Ciocca, Mario
    Donetti, Marco
    Mairani, Andrea
    MEDICAL PHYSICS, 2022, 49 (12) : 7802 - 7814
  • [25] Feasibility study of utilizing XRV-124 scintillation detector for quality assurance of spot profile in pencil beam scanning proton therapy
    Rana, Suresh
    Samuel, E. James Jebaseelan
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2019, 66 : 15 - 20
  • [26] Patient-Specific Quality Assurance Using Monte Carlo Dose Calculation and Elekta Log Files for Prostate Volumetric-Modulated Arc Therapy
    Katsuta, Yoshiyuki
    Kadoya, Noriyuki
    Fujita, Yukio
    Shimizu, Eiji
    Matsunaga, Kenichi
    Sawada, Kinya
    Matsushita, Haruo
    Majima, Kazuhiro
    Jingu, Keiichi
    TECHNOLOGY IN CANCER RESEARCH & TREATMENT, 2017, 16 (06) : 1220 - 1225
  • [27] Patient-Specific Quality Assurance Protocol for Volumetric Modulated Arc Therapy using Dose Volume Histogram
    Low, Christopher
    Toye, Warren
    Phung, Peter
    Huston, Christopher
    JOURNAL OF MEDICAL PHYSICS, 2018, 43 (02) : 112 - 118
  • [28] Inverse Verification of the Dose Distribution for Intensity Modulated Radiation Therapy Patient-specific Quality Assurance Using Dynamic MLC Log Files
    Lee, Jeong-Woo
    Park, Jeong-Hoon
    Chung, Jin-Beom
    Park, Ji-Yeon
    Choe, Bo-Young
    Suh, Tae-Suk
    Lee, Doo-Hyun
    Hong, Semie
    Kang, Min-Young
    Choi, Kyoung-Sik
    JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2009, 55 (04) : 1649 - 1656
  • [29] Quality assurance-based optimization (QAO): Towards improving patient-specific quality assurance in volumetric modulated arc therapy plans using machine learning
    Wall, Phillip D. H.
    Fontenot, Jonas D.
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2021, 87 : 136 - 143
  • [30] Validity of one-time phantomless patient-specific quality assurance in proton therapy with regard to the reproducibility of beam delivery
    Wolter, Lukas Cornelius
    Hennings, Fabian
    Bokor, Jozef
    Richter, Christian
    Stuetzer, Kristin
    MEDICAL PHYSICS, 2025,