Improving spot-scanning proton therapy patient specific quality assurance with HPlusQA, a second-check dose calculation engine

被引:26
|
作者
Mackin, Dennis [1 ]
Li, Yupeng [1 ]
Taylor, Michael B. [1 ]
Kerr, Matthew [1 ]
Holmes, Charles [1 ]
Sahoo, Narayan [1 ]
Poenisch, Falk [1 ]
Li, Heng [1 ]
Lii, Jim [1 ]
Amos, Richard [1 ]
Wu, Richard [1 ]
Suzuki, Kazumichi [1 ]
Gillin, Michael T. [1 ]
Zhu, X. Ronald [1 ]
Zhang, Xiaodong [1 ]
机构
[1] Univ Texas MD Anderson Canc Ctr, Dept Radiat Phys, Houston, TX 77030 USA
关键词
spot scanning; proton therapy; IMPT; gamma index; patient quality assurance; TREATMENT PLANNING SYSTEM; ANDERSON CANCER CENTER; IMRT; VERIFICATION; BEAMS; HOUSTON; QA;
D O I
10.1118/1.4828775
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: The purpose of this study was to validate the use of HPlusQA, spot-scanning proton therapy (SSPT) dose calculation software developed at The University of Texas MD Anderson Cancer Center, as second-check dose calculation software for patient-specific quality assurance (PSQA). The authors also showed how HPlusQA can be used within the current PSQA framework. Methods: The authors compared the dose calculations of HPlusQA and the Eclipse treatment planning system with 106 planar dose measurements made as part of PSQA. To determine the relative performance and the degree of correlation between HPlusQA and Eclipse, the authors compared calculated with measured point doses. Then, to determine how well HPlusQA can predict when the comparisons between Eclipse calculations and the measured dose will exceed tolerance levels, the authors compared gamma index scores for HPlusQA versus Eclipse with those of measured doses versus Eclipse. The authors introduce the alpha beta gamma transformation as a way to more easily compare gamma scores. Results: The authors compared measured and calculated dose planes using the relative depth, z/R x 100%, where z is the depth of the measurement and R is the proton beam range. For relative depths than less than 80%, both Eclipse and HPlusQA calculations were within 2 cGy of dose measurements on average. When the relative depth was greater than 80%, the agreement between the calculations and measurements fell to 4 cGy. For relative depths less than 10%, the Eclipse and HPlusQA dose discrepancies showed a negative correlation, -0.21. Otherwise, the correlation between the dose discrepancies was positive and as large as 0.6. For the dose planes in this study, HPlusQA correctly predicted when Eclipse had and had not calculated the dose to within tolerance 92% and 79% of the time, respectively. In 4 of 106 cases, HPlusQA failed to predict when the comparison between measurement and Eclipse's calculation had exceeded the tolerance levels of 3% for dose and 3 mm for distance-to-agreement. Conclusions: The authors found HPlusQA to be reasonably effective (79% +/- 10%) in determining when the comparison between measured dose planes and the dose planes calculated by the Eclipse treatment planning system had exceeded the acceptable tolerance levels. When used as described in this study, HPlusQA can reduce the need for patient specific quality assurance measurements by 64%. The authors believe that the use of HPlusQA as a dose calculation second check can increase the efficiency and effectiveness of the QA process. (C) 2013 American Association of Physicists in Medicine.
引用
收藏
页数:10
相关论文
共 36 条
  • [11] An analytical model for the upper bound estimation of respiratory motion-induced dose uncertainty in spot-scanning proton beam therapy
    Li, Heng
    Zhang, Xiaodong
    Li, Yupeng
    Zhu, Ronald X.
    MEDICAL PHYSICS, 2019, 46 (11) : 5249 - 5261
  • [12] An analytical dose-averaged LET calculation algorithm considering the off-axis LET enhancement by secondary protons for spot-scanning proton therapy
    Hirayama, Shusuke
    Matsuura, Taeko
    Ueda, Hideaki
    Fujii, Yusuke
    Fujii, Takaaki
    Takao, Seishin
    Miyamoto, Naoki
    Shimizu, Shinichi
    Fujimoto, Rintaro
    Umegaki, Kikuo
    Shirato, Hiroki
    MEDICAL PHYSICS, 2018, 45 (07) : 3404 - 3416
  • [13] Fast spot-scanning proton dose calculation method with uncertainty quantification using a three-dimensional convolutional neural network
    Nomura, Yusuke
    Wang, Jeff
    Shirato, Hiroki
    Shimizu, Shinichi
    Xing, Lei
    PHYSICS IN MEDICINE AND BIOLOGY, 2020, 65 (21)
  • [14] Evaluation and second check of a commercial Monte Carlo dose engine for small-field apertures in pencil beam scanning proton therapy
    Holmes, Jason
    Shen, Jiajian
    Shan, Jie
    Patrick, Christopher L.
    Wong, William W.
    Foote, Robert L.
    Patel, Samir H.
    Bues, Martin
    Liu, Wei
    MEDICAL PHYSICS, 2022, 49 (05) : 3497 - 3506
  • [15] Analysis of dose comparison techniques for patient-specific quality assurance in radiation therapy
    Yu, Liting
    Tang, Timothy L. S.
    Cassim, Naasiha
    Livingstone, Alexander
    Cassidy, Darren
    Kairn, Tanya
    Crowe, Scott B.
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2019, 20 (11): : 189 - 198
  • [16] Validation of dose distribution for liver tumors treated with real-time-image gated spot-scanning proton therapy by log data based dose reconstruction
    Yamada, Takahiro
    Takao, Seishin
    Koyano, Hidenori
    Nihongi, Hideaki
    Fujii, Yusuke
    Hirayama, Shusuke
    Miyamoto, Naoki
    Matsuura, Taeko
    Umegaki, Kikuo
    Katoh, Norio
    Yokota, Isao
    Shirato, Hiroki
    Shimizu, Shinichi
    JOURNAL OF RADIATION RESEARCH, 2021, 62 (04) : 626 - 633
  • [17] Patient-specific quality assurance and plan dose errors on breast intensity-modulated proton therapy
    Liu, Chunbo
    Zheng, Dandan
    Bradley, Julie A.
    Vega, Raymond B. Mailhot
    Li, Zuofeng
    Mendenhall, Nancy P.
    Liang, Xiaoying
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2020, 77 : 84 - 91
  • [18] Patient-specific quality assurance of dynamically-collimated proton therapy treatment plans
    Bennett, Laura C.
    Hyer, Daniel E.
    Vu, Justin
    Patwardhan, Kaustubh
    Erhart, Kevin
    Gutierrez, Alonso N.
    Pons, Eduardo
    Jensen, Eric
    Ubau, Manual
    Zapata, Julio
    Wroe, Andrew
    Wake, Karsten
    Nelson, Nicholas P.
    Culberson, Wesley S.
    Smith, Blake R.
    Hill, Patrick M.
    Flynn, Ryan T.
    MEDICAL PHYSICS, 2024, : 5901 - 5910
  • [19] A fast and reliable method for daily quality assurance in spot scanning proton therapy with a compact and inexpensive phantom
    Bizzocchi, Nicola
    Fracchiolla, Francesco
    Schwarz, Marco
    Algranati, Carlo
    MEDICAL DOSIMETRY, 2017, 42 (03) : 238 - 246
  • [20] Dose distribution effects of spot-scanning proton beam therapy equipped with a multi-leaf collimator for pediatric brain tumors
    Fukumitsu, Nobuyoshi
    Yamashita, Tomohiro
    Mima, Masayuki
    Demizu, Yusuke
    Suzuki, Takeshi
    Soejima, Toshinori
    ONCOLOGY LETTERS, 2021, 22 (02)