Improving spot-scanning proton therapy patient specific quality assurance with HPlusQA, a second-check dose calculation engine

被引:26
|
作者
Mackin, Dennis [1 ]
Li, Yupeng [1 ]
Taylor, Michael B. [1 ]
Kerr, Matthew [1 ]
Holmes, Charles [1 ]
Sahoo, Narayan [1 ]
Poenisch, Falk [1 ]
Li, Heng [1 ]
Lii, Jim [1 ]
Amos, Richard [1 ]
Wu, Richard [1 ]
Suzuki, Kazumichi [1 ]
Gillin, Michael T. [1 ]
Zhu, X. Ronald [1 ]
Zhang, Xiaodong [1 ]
机构
[1] Univ Texas MD Anderson Canc Ctr, Dept Radiat Phys, Houston, TX 77030 USA
关键词
spot scanning; proton therapy; IMPT; gamma index; patient quality assurance; TREATMENT PLANNING SYSTEM; ANDERSON CANCER CENTER; IMRT; VERIFICATION; BEAMS; HOUSTON; QA;
D O I
10.1118/1.4828775
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: The purpose of this study was to validate the use of HPlusQA, spot-scanning proton therapy (SSPT) dose calculation software developed at The University of Texas MD Anderson Cancer Center, as second-check dose calculation software for patient-specific quality assurance (PSQA). The authors also showed how HPlusQA can be used within the current PSQA framework. Methods: The authors compared the dose calculations of HPlusQA and the Eclipse treatment planning system with 106 planar dose measurements made as part of PSQA. To determine the relative performance and the degree of correlation between HPlusQA and Eclipse, the authors compared calculated with measured point doses. Then, to determine how well HPlusQA can predict when the comparisons between Eclipse calculations and the measured dose will exceed tolerance levels, the authors compared gamma index scores for HPlusQA versus Eclipse with those of measured doses versus Eclipse. The authors introduce the alpha beta gamma transformation as a way to more easily compare gamma scores. Results: The authors compared measured and calculated dose planes using the relative depth, z/R x 100%, where z is the depth of the measurement and R is the proton beam range. For relative depths than less than 80%, both Eclipse and HPlusQA calculations were within 2 cGy of dose measurements on average. When the relative depth was greater than 80%, the agreement between the calculations and measurements fell to 4 cGy. For relative depths less than 10%, the Eclipse and HPlusQA dose discrepancies showed a negative correlation, -0.21. Otherwise, the correlation between the dose discrepancies was positive and as large as 0.6. For the dose planes in this study, HPlusQA correctly predicted when Eclipse had and had not calculated the dose to within tolerance 92% and 79% of the time, respectively. In 4 of 106 cases, HPlusQA failed to predict when the comparison between measurement and Eclipse's calculation had exceeded the tolerance levels of 3% for dose and 3 mm for distance-to-agreement. Conclusions: The authors found HPlusQA to be reasonably effective (79% +/- 10%) in determining when the comparison between measured dose planes and the dose planes calculated by the Eclipse treatment planning system had exceeded the acceptable tolerance levels. When used as described in this study, HPlusQA can reduce the need for patient specific quality assurance measurements by 64%. The authors believe that the use of HPlusQA as a dose calculation second check can increase the efficiency and effectiveness of the QA process. (C) 2013 American Association of Physicists in Medicine.
引用
收藏
页数:10
相关论文
共 36 条
  • [1] Automation of routine elements for spot-scanning proton patient-specific quality assurance
    Morales, Danairis Hernandez
    Shan, Jie
    Liu, Wei
    Augustine, Kurt E.
    Bues, Martin
    Davis, Michael J.
    Fatyga, Mirek
    Johnson, Jedediah E.
    Mundy, Daniel W.
    Shen, Jiajian
    Younkin, James E.
    Stoker, Joshua B.
    MEDICAL PHYSICS, 2019, 46 (01) : 5 - 14
  • [2] Spot-Scanning Proton Therapy Patient-Specific Quality Assurance: Our Methodology and Results From 295 Treatment Plans
    Mackin, D.
    Zhang, X.
    Li, Y.
    Li, H.
    Wu, R.
    Poenisch, F.
    Suzuki, K.
    Kerr, M.
    Holmes, C.
    Zhu, X.
    Sahoo, N.
    Gillin, M.
    MEDICAL PHYSICS, 2014, 41 (06) : 429 - 429
  • [3] Towards Effective and Efficient Patient-Specific Quality Assurance for Spot Scanning Proton Therapy
    Zhu, X. Ronald.
    Li, Yupeng
    Mackin, Dennis
    Li, Heng
    Poenisch, Falk
    Lee, Andrew K.
    Mahajan, Anita
    Frank, Steven J.
    Gillin, Michael T.
    Sahoo, Narayan
    Zhang, Xiaodong
    CANCERS, 2015, 7 (02): : 631 - 647
  • [4] Use of treatment log files in spot scanning proton therapy as part of patient-specific quality assurance
    Li, Heng
    Sahoo, Narayan
    Poenisch, Falk
    Suzuki, Kazumichi
    Li, Yupeng
    Li, Xiaoqiang
    Zhang, Xiaodong
    Lee, Andrew K.
    Gillin, Michael T.
    Zhu, X. Ronald
    MEDICAL PHYSICS, 2013, 40 (02)
  • [5] PATIENT-SPECIFIC QUALITY ASSURANCE FOR PROSTATE CANCER PATIENTS RECEIVING SPOT SCANNING PROTON THERAPY USING SINGLE-FIELD UNIFORM DOSE
    Zhu, X. Ronald
    Poenisch, Falk
    Song, Xiaofei
    Johnson, Jennifer L.
    Ciangaru, George
    Taylor, M. Brad
    Lii, MingFwu
    Martin, Craig
    Arjomandy, Bijan
    Lee, Andrew K.
    Choi, Seungtaek
    Quynh Nhu Nguyen
    Gillin, Michael T.
    Sahoo, Narayan
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2011, 81 (02): : 552 - 559
  • [6] Feasibility of patient specific quality assurance for proton therapy based on independent dose calculation and predicted outcomes
    Meijers, Arturs
    Marmitt, Gabriel Guterres
    Siang, Kelvin Ng Wei
    van der Schaaf, Arjen
    Knopf, Antje C.
    Langendijk, Johannes A.
    Both, Stefan
    RADIOTHERAPY AND ONCOLOGY, 2020, 150 : 136 - 141
  • [7] Three-dimensional gamma criterion for patient-specific quality assurance of spot scanning proton beams
    Chang, Chang
    Poole, Kendra L.
    Teran, Anthony V.
    Luckman, Scott
    Mah, Dennis
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2015, 16 (05): : 381 - 388
  • [8] Analysis of measurement deviations for the patient-specific quality assurance using intensity-modulated spot-scanning particle beams
    Li, Yongqiang
    Hsi, Wen C.
    PHYSICS IN MEDICINE AND BIOLOGY, 2017, 62 (07) : 2675 - 2693
  • [9] Evaluation of a prototype array for daily quality assurance in spot scanning proton therapy
    Flatten, Veronika
    Devendranath, Henry-Aravinth
    Kroh, Janik
    Witt, Matthias
    Baumann, Kilian-Simon
    Gall, Kenneth
    Simon, Bill
    Wulff, Joerg
    Schoenfeld, Andreas A.
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2024, 25 (11):
  • [10] Highly efficient and sensitive patient-specific quality assurance for spot-scanned proton therapy
    Johnson, J. E.
    Beltran, C.
    Tseung, H. Wan Chan
    Mundy, D. W.
    Kruse, J. J.
    Whitaker, T. J.
    Herman, M. G.
    Furutani, K. M.
    PLOS ONE, 2019, 14 (02):