Monte Carlo Design Study of a Gamma Detector System to Locate Distal Dose Falloff in Proton Therapy

被引:24
作者
Kang, Byung-Hwi [1 ]
Kim, Jong-Won [2 ]
机构
[1] Seoul Natl Univ, Dept Phys, Seoul 151742, South Korea
[2] Natl Canc Ctr, Goyang 410769, Kyonggi, South Korea
关键词
Distal dose falloff; gaseous detector; Monte Carlo; prompt gamma; proton therapy; time projection chamber; RAY IMAGING DETECTOR; RADIATION;
D O I
10.1109/TNS.2008.2005189
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Monte Carlo simulations have been performed to design a prompt gamma detector system for use as a quality assurance device in proton therapy. The system measures prompt gammas emitted in the direction normal to the incident therapeutic protons to correlate with the location of the distal dose falloff. The system consists of a drift chamber to measure the energies and directions of recoil electrons and a scintillator array to measure the positions and energies of scattered gammas in coincidence with the electrons. To ensure the use of the system in a therapeutic environment, the detection rate of Compton-scattered prompt gammas was simulated. The simulation predicts that the reconstruction uncertainty is around 1 cm for a practical configuration of the detector system. The resolution of the reconstruction was evaluated for varying dimensions and energy resolutions assumed for the detector. An actual system is,being constructed based on this work.
引用
收藏
页码:46 / 50
页数:5
相关论文
共 14 条
[1]  
[Anonymous], 2003, 39 MEET PART
[2]   Radiation in the treatment of cancer [J].
Boyer, AL ;
Goitein, M ;
Lomax, AJ ;
Pedroni, ES .
PHYSICS TODAY, 2002, 55 (09) :34-36
[3]  
BRUN R, 1994, GEANT3 DETECTOR DESC
[4]  
FASSO A, 2005, INFCTC0511
[5]   The GEANT low energy Compton scattering (GLECS) package for use in simulating advanced Compton telescopes [J].
Kippen, RM .
NEW ASTRONOMY REVIEWS, 2004, 48 (1-4) :221-225
[6]   Ultralow threshold on-chip microcavity nanocrystal quantum dot lasers [J].
Min, Bumki ;
Kim, Sungjee ;
Okamoto, Koichi ;
Yang, Lan ;
Scherer, Axel ;
Atwater, Harry ;
Vahala, Kerry .
APPLIED PHYSICS LETTERS, 2006, 89 (19)
[7]   Determination of distal dose edge location by measuring right-angled prompt-gamma rays from a 38 MeV proton beam [J].
Min, C. H. ;
Kim, J. W. ;
Youn, M. Y. ;
Kim, C. H. .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2007, 580 (01) :562-565
[8]   Dose-volume delivery guided proton therapy using beam on-line PET system [J].
Nishio, Teiji ;
Ogino, Takashi ;
Nomura, Kazuhiro ;
Uchida, Hiroshi .
MEDICAL PHYSICS, 2006, 33 (11) :4190-4197
[9]   A novel design of the MeV gamma-ray imaging detector with micro-TPC [J].
Orito, R ;
Kubo, H ;
Miuchi, K ;
Nagayoshi, T ;
Takada, A ;
Tanimori, T ;
Ueno, M .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2003, 513 (1-2) :408-412
[10]   Ion beam therapy in 2004 [J].
Sisterson, J .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 2005, 241 (1-4) :713-716