Experimental limitations in impedance spectroscopy .5. Apparatus contributions and corrections

被引:52
作者
Edwards, DD
Hwang, JH
Ford, SJ
Mason, TO
机构
[1] NORTHWESTERN UNIV,DEPT MAT SCI & ENGN,EVANSTON,IL 60208
[2] NORTHWESTERN UNIV,MAT RES CTR,EVANSTON,IL 60208
[3] NORTHWESTERN UNIV,CTR ADV CEMENT BASED MAT,EVANSTON,IL 60208
关键词
impedance spectroscopy; apparatus; simulation;
D O I
10.1016/S0167-2738(97)00206-3
中图分类号
O64 [物理化学(理论化学)、化学物理学];
学科分类号
070304 ; 081704 ;
摘要
Overlooked apparatus (cabling, leads, sample holder) contributions to experimental impedance spectra can be significant, especially at high frequencies, and can obscure the true sample response. Instrumental Limitations are discussed and high-frequency artifacts arising from apparatus contributions are investigated as they pertain to accurate impedance measurements of materials systems. Remediation strategies are presented, including geometrical adjustments, lead shielding, and null-correction procedures.
引用
收藏
页码:85 / 93
页数:9
相关论文
共 16 条
[1]   STUDY OF SOLID ELECTROLYTE POLARIZATION BY A COMPLEX ADMITTANCE METHOD [J].
BAUERLE, JE .
JOURNAL OF PHYSICS AND CHEMISTRY OF SOLIDS, 1969, 30 (12) :2657-&
[2]  
BOUKAMP BA, 1990, EQUIVCRT DEP CHEM TE
[3]   IMPEDANCE SPECTROSCOPY OF HYDRATING CEMENT-EASED MATERIALS - MEASUREMENT, INTERPRETATION, AND APPLICATION [J].
CHRISTENSEN, BJ ;
COVERDALE, RT ;
OLSON, RA ;
FORD, SJ ;
GARBOCZI, EJ ;
JENNINGS, HM ;
MASON, TO .
JOURNAL OF THE AMERICAN CERAMIC SOCIETY, 1994, 77 (11) :2789-2804
[4]  
COOPER EA, 1992, THESIS NW U EVANSTON
[5]   Experimental limitations in impedance spectroscopy .3. Effect of reference electrode geometry/position [J].
Hsieh, G ;
Mason, TO ;
Garboczi, EJ ;
Pederson, LR .
SOLID STATE IONICS, 1997, 96 (3-4) :153-172
[6]   Experimental limitations in impedance spectroscopy .2. Electrode artifacts in three-point measurements on Pt/YSZ [J].
Hsieh, G ;
Mason, TO ;
Pederson, LR .
SOLID STATE IONICS, 1996, 91 (3-4) :203-212
[7]   Experimental limitations in impedance spectroscopy .1. Simulation of reference electrode artifacts in three-point measurements [J].
Hsieh, G ;
Ford, SJ ;
Mason, TO ;
Pederson, LR .
SOLID STATE IONICS, 1996, 91 (3-4) :191-201
[8]   Experimental limitations in impedance spectroscopy .4. Electrode contact effects [J].
Hwang, JH ;
Kirkpatrick, KS ;
Mason, TO ;
Garboczi, EJ .
SOLID STATE IONICS, 1997, 98 (1-2) :93-104
[9]   Characterization of humidity-sensing NiO-Ni(OH)(2) nanocomposites by impedance spectroscopy [J].
Hwang, JH ;
Mason, TO ;
Buehler, MF ;
Darab, JG ;
Matson, DW ;
Linehan, JC .
MATERIALS SCIENCE AND ENGINEERING A-STRUCTURAL MATERIALS PROPERTIES MICROSTRUCTURE AND PROCESSING, 1995, 204 (1-2) :252-257
[10]  
Irvine J. T. S., 1990, Advanced Materials, V2, P132, DOI 10.1002/adma.19900020304