Transcatheter closure of atrial septal defects with the CardioSEAL-device

被引:1
作者
Kaulitz, R [1 ]
Bertram, H [1 ]
Peuster, M [1 ]
Zutz, M [1 ]
Fink, C [1 ]
Hausdorf, G [1 ]
机构
[1] Hannover Med Sch, Abt Padiatr Kardiol & Padiat Intens Med, D-30625 Hannover, Germany
来源
ZEITSCHRIFT FUR KARDIOLOGIE | 2002年 / 91卷 / 02期
关键词
congenital heart defects; atrial septal defect; interventional cardiology; transcatheter defect closure;
D O I
10.1007/s003920200007
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Data on mid- and long-term follow-up for the recent devices for closure of secundum-type atrial septal defects are limited. The purpose of our retrospective study was to report the effectiveness of transcatheter closure in patients with various morphological types of atrial septal defect other than centrally located defects within the oval fossa using the CardioSEAL and CardioSEAL-Starflex occluder. A total of 91 patients (age 1.5-71 years, median 6 years) underwent transcatheter closure. On the transesophageal echocardiogram, defect size varied from 6 to 18 mm with an estimated stretched diameter of 11 to 24 mm, median 15 mm; the ratio of the stretched diameter to septal length ranged from 0.28 to 0.68. Mean follow-up was 28.7 +/- 11,9 months (range 3-46 months). Isolated secundum-type defects were present in 59 patients (65%), multiple septal defects including patients with perforated atrial septal aneurysms and defects with deficient atrial rim in 32 patients (35%). Occlusion rate using device diameters from 23 to 40 mm increased from 66% (60/91 patients) immediately after implantation to 86% (48/56 patients) 24 months after implantation. Patients with isolated secundum-type defects presented with a significantly higher primary closure rate (45/59 patients, 76%) compared to patients with various defect morphology. Closure rate did not depend on the type of implanted device modification. No thrombus formation, sustained atrial arrhythmia or infective endocarditis occurred. Serial transthoracic echocardiographic findings revealed protrusion of one left-sided arm onto the right atrial aspect in 5 patients; malposition of one right-sided superior arm of the device was observed in 7 patients. Fluoroscopy showed single fatigue fracture in 7 patients (7.7%) within the first 6 months after implantation. These results demonstrate that transcatheter closure with the double umbrella device was effective and safe on medium-term follow-up and could be extended to atrial septal defects of various morphology.
引用
收藏
页码:169 / 177
页数:9
相关论文
共 28 条
[1]   Influence of atrial septal defect anatomy in patient selection and assessment of closure with the Cardioseal device - A three-dimensional transoesophageal echocardiographic reconstruction [J].
Acar, P ;
Saliba, Z ;
Bonhoeffer, P ;
Aggoun, Y ;
Bonnet, D ;
Sidi, D ;
Kachaner, J .
EUROPEAN HEART JOURNAL, 2000, 21 (07) :573-581
[2]   ECHOCARDIOGRAPHIC FOLLOW-UP OF ATRIAL SEPTAL-DEFECT AFTER CATHETER CLOSURE BY DOUBLE-UMBRELLA DEVICE [J].
BOUTIN, C ;
MUSEWE, NN ;
SMALLHORN, JF ;
DYCK, JD ;
KOBAYASHI, T ;
BENSON, LN .
CIRCULATION, 1993, 88 (02) :621-627
[3]   Transcatheter closure of multiple atrial septal defects - Initial results and value of two- and three-dimensional transoesophageal echocardiography [J].
Cao, QL ;
Radtke, W ;
Berger, F ;
Zhu, W ;
Hijazi, ZM .
EUROPEAN HEART JOURNAL, 2000, 21 (11) :941-947
[4]  
CHAN KC, 1993, BRIT HEART J, V69, P52
[5]   Morphology of perforated atrial septal aneurysm suitable for closure by transcatheter device placement [J].
Ewert, P ;
Berger, F ;
Vogel, M ;
Dähnert, I ;
Alexi-Meshkishvili, V ;
Lange, PE .
HEART, 2000, 84 (03) :327-331
[6]  
FERREIRA SMAG, 1992, BRIT HEART J, V67, P316
[7]   Transcatheter closure of secundum atrial septal defects with the new self-centering Amplatzer Septal Occluder [J].
Fischer, G ;
Kramer, HH ;
Stieh, J ;
Harding, P ;
Jung, O .
EUROPEAN HEART JOURNAL, 1999, 20 (07) :541-549
[8]  
Franke A, 1997, CIRCULATION, V96, P323
[9]   2-DIMENSIONAL ECHOCARDIOGRAPHIC AND COLOR DOPPLER MEASUREMENTS OF ATRIAL SEPTAL-DEFECT, AND COMPARISON WITH THE BALLOON-STRETCHED DIAMETER [J].
GODART, F ;
REY, C ;
FRANCART, C ;
JARRAR, M ;
VAKSMANN, G .
AMERICAN JOURNAL OF CARDIOLOGY, 1993, 72 (14) :1095-1097
[10]   Transcatheter closure of atrial septal defect with a new flexible, self-centering device - (The STARFlex occluder) [J].
Hausdorf, G ;
Kaulitz, R ;
Paul, T ;
Carminati, M ;
Lock, J .
AMERICAN JOURNAL OF CARDIOLOGY, 1999, 84 (09) :1113-+