Hippocampal-sparing whole-brain radiotherapy using the Elekta equipment

被引:26
作者
Nevelsky, Alexander [1 ]
Ieumwananonthachai, Nantakan [2 ]
Kaidar-Person, Orit [1 ]
Bar-Deroma, Raquel [1 ]
Nasrallah, Haitam [1 ]
Ben-Yosef, Rahamim [1 ]
Kuten, Abraham [1 ]
机构
[1] Rambam Hlth Care Campus, Div Oncol, IL-31096 Haifa, Israel
[2] Chulabhorn Canc Ctr, Bangkok, Thailand
关键词
brain metastases; conformal avoidance; hippocampus; linear accelerator; radiotherapy; whole brain; DOSE CALCULATION; RADIATION-THERAPY; METASTASES; IMRT; DISTRIBUTIONS; IRRADIATION; ALGORITHMS; QUALITY;
D O I
10.1120/jacmp.v14i3.4205
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The purpose of this study was to evaluate the feasibility of hippocampal-sparing whole-brain radiotherapy (HS WBRT) using the Elekta Infinity linear accelerator and Monaco treatment planning system (TPS). Ten treatment plans were created for HS-WBRT to a dose of 30 Gy (10 fractions). RTOG 0933 recommendations were applied for treatment planning. Intensity-modulated radiotherapy (IMRT) plans for the Elekta Infinity linear accelerator were created using Monaco 3.1 TPS-based on a nine-field arrangement and step-and-shoot delivery method. Plan evaluation was performed using D2% and D98% for the whole-brain PTV (defined as whole brain excluding hippocampus avoidance region), D100% and maximum dose to the hippocampus, and maximum dose to optic nerves and chiasm. Homogeneity index (HI) defined as (D2%-D98%)/D-median was used to quantify dose homogeneity in the PTV. The whole-brain PTV D2% mean value was 37.28 Gy (range 36.95-37.49 Gy), and D98% mean value was 25.37 Gy (range 25.40-25.89 Gy). The hippocampus D100% mean value was 8.37 Gy (range 7.48-8.97 Gy) and the hippocampus maximum dose mean value was 14.35 Gy (range 13.48-15.40 Gy). The maximum dose to optic nerves and optic chiasm for all patients did not exceed 37.50 Gy. HI mean value was 0.36 (range 0.34-0.37). Mean number of segments was 105 (range 88-122) and mean number of monitor units was 1724 (range 1622-1914). Gamma evaluation showed that all plans passed 3%, 3 mm criteria with more than 99% of the measured points. These results indicate that Elekta equipment (Elekta Infinity linac and Monaco TPS) can be used for HS WBRT planning according to compliance criteria defined by the RTOG 0933 protocol.
引用
收藏
页码:113 / 120
页数:8
相关论文
共 26 条
[1]  
[Anonymous], 2010, J ICRU, DOI [10.1093/jicru_ndq002, DOI 10.1093/JICRU_NDQ002]
[2]   Evaluation of the Delta4 phantom for IMRT and VMAT verification [J].
Bedford, James L. ;
Lee, Young K. ;
Wai, Philip ;
South, Christopher P. ;
Warrington, Alan P. .
PHYSICS IN MEDICINE AND BIOLOGY, 2009, 54 (09) :N167-N176
[3]   RADIATION-INDUCED DEMENTIA IN PATIENTS CURED OF BRAIN METASTASES [J].
DEANGELIS, LM ;
DELATTRE, JY ;
POSNER, JB .
NEUROLOGY, 1989, 39 (06) :789-796
[4]   Evaluation of a biplanar diode array dosimeter for quality assurance of step-and-shoot IMRT [J].
Feygelman, Vladimir ;
Forster, Kenneth ;
Opp, Daniel ;
Nilsson, Gorgen .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2009, 10 (04) :64-78
[5]   On the dosimetric behaviour of photon dose calculation algorithms in the presence of simple geometric heterogeneities:: comparison with Monte Carlo calculations [J].
Fogliata, Antonella ;
Vanetti, Eugenio ;
Albers, Dirk ;
Brink, Carsten ;
Clivio, Alessandro ;
Knoos, Tommy ;
Nicolini, Giorgia ;
Cozzi, Luca .
PHYSICS IN MEDICINE AND BIOLOGY, 2007, 52 (05) :1363-1385
[6]  
Fotina I, 2011, STRAHLENTHER ONKOL, V187, P433, DOI 10.1007/s00066-011-2215-9
[7]   Advanced kernel methods vs. Monte Carlo-based dose calculation for high energy photon beams [J].
Fotina, Irina ;
Winkler, Peter ;
Kuenzler, Thomas ;
Reiterer, Jochen ;
Simmat, Isabell ;
Georg, Dietmar .
RADIOTHERAPY AND ONCOLOGY, 2009, 93 (03) :645-653
[8]   Distribution of brain metastases in relation to the hippocampus:: Implications for neurocognitive functional preservation [J].
Ghia, Amol ;
Tome, Wolfgang A. ;
Thomas, Sayana ;
Cannon, George ;
Khuntia, Deepak ;
Kuo, John S. ;
Mehta, Mineshi P. .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2007, 68 (04) :971-977
[9]   HIPPOCAMPAL-SPARING WHOLE-BRAIN RADIOTHERAPY: A "HOW-TO" TECHNIQUE USING HELICAL TOMOTHERAPY AND LINEAR ACCELERATOR-BASED INTENSITY-MODULATED RADIOTHERAPY [J].
Gondi, Vinai ;
Tolakanahalli, Ranjini ;
Mehta, Minesh P. ;
Tewatia, Dinesh ;
Rowley, Howard ;
Kuo, John S. ;
Khuntia, Deepak ;
Tome, Wolfgang A. .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2010, 78 (04) :1244-1252
[10]   Estimated risk of perihippocampal disease progression after hippocampal avoidance during whole-brain radiotherapy: Safety profile for RTOG 0933 [J].
Gondi, Vinai ;
Tome, Wolfgang A. ;
Marsh, James ;
Struck, Aaron ;
Ghia, Amol ;
Turian, Julius V. ;
Bentzen, Soren M. ;
Kuo, John S. ;
Khuntia, Deepak ;
Mehta, Minesh P. .
RADIOTHERAPY AND ONCOLOGY, 2010, 95 (03) :327-331