Experimental evaluation of the holding power/stiffness of the self-tapping bone screws in normal and osteoporotic bone material

被引:57
作者
Battula, Suneel
Schoenfeld, Andrew
Vrabec, Gregory
Njus, Glen O.
机构
[1] Univ Akron, Dept Biomed Engn, Sydney Olson Res Ctr, Akron, OH 44325 USA
[2] Akron Gen Med Ctr, Dept Orthoped Surg, Akron, OH 44307 USA
关键词
biomechanical testing; screw pullout; holding power; osteoporosis; self-tapping screws; depth of insertion;
D O I
10.1016/j.clinbiomech.2005.12.020
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Objective. The goal of this study is to compare the holding power of the self-tapping bone screws in normal and osteoporotic bone materials. Background. Self-tapping screws are increasingly being used in orthopaedic surgery due to their advantages over the other bone screws. Methods. Screws were divided into five groups (six screws per group) based on the depth of insertion in the bone coupons that represented normal and osteoporotic bones. Screws were randomly inserted into the bone coupons with tips of the screws being -1 mm, 0 mm, I mm, 2 mm and 3 mm relative to the far cortex. Biomechanical testing was performed using an Instron 8511 in accordance with the American Society for Testing and Materials standards for bone screws. Two-factor analysis of variance (ANOVA) was used to determine if the holding power of the screws were different with respect to insertion depths and bone materials. Findings. The bone materials had a significant difference (P < 0.05) in the holding power and depths of insertion past the far cortex were significantly different from one another in holding power. The affect of the screw material on the holding power of the self-tapping screws in different bone materials was also examined. The performance of stainless steel screws was superior to that of titanium screws in the osteoporotic material. Interpretation. Based on the results it can be concluded that the depth of insertion of the tip of the screw for adequate fracture fixation in normal bone is 1 mm or more past the far cortex and in osteoporotic bone it is at least 2 mm past the far cortex. (c) 2006 Elsevier Ltd. All rights reserved.
引用
收藏
页码:533 / 537
页数:5
相关论文
共 9 条
[1]  
*ASTM, 2003, ANN BOOK ASTM STA 13
[2]  
BATTULA S, 2005, P 20 C INT SOC BIOM
[3]  
BATTULA S, 2004, P BMES ANN FALL M PH, P27
[4]   Pullout strength of self-tapping screws inserted to different depths [J].
Berkowitz, R ;
Njus, G ;
Vrabec, G .
JOURNAL OF ORTHOPAEDIC TRAUMA, 2005, 19 (07) :462-465
[5]  
NETO RB, 1999, REV HOSP CLIN SAO PA, V54, P6
[6]   THE WORLDWIDE PROBLEM OF OSTEOPOROSIS - INSIGHTS AFFORDED BY EPIDEMIOLOGY [J].
RIGGS, BL ;
MELTON, LJ .
BONE, 1995, 17 (05) :S505-S511
[7]   The degree of osteoporosis in patients with vertebral fracture and patients with hip fracture: Relationship to incidence of vertebral fracture [J].
Takahashi, M ;
Kushida, K ;
Naitou, K .
JOURNAL OF BONE AND MINERAL METABOLISM, 1999, 17 (03) :187-194
[8]  
Ugural A.C., 2003, ADV STRENGTH APPL EL
[9]   Effect of cutting flute design on cortical bone screw insertion torque and pullout strength [J].
Yerby, S ;
Scott, CC ;
Evans, NJ ;
Messing, KL ;
Carter, DR .
JOURNAL OF ORTHOPAEDIC TRAUMA, 2001, 15 (03) :216-221