Characterization of flexible ECoG electrode arrays for chronic recording in awake rats

被引:91
作者
Yeager, John D. [2 ]
Phillips, Derrick J. [1 ]
Rector, David M. [1 ]
Bahr, David F. [2 ]
机构
[1] Washington State Univ, Dept Vet Comparat Anat Physiol & Pharmacol, Pullman, WA 99164 USA
[2] Washington State Univ, Sch Mech & Mat Engn, Pullman, WA 99164 USA
关键词
surface-evoked potential; polyimide; Kapton (TM); biocompatibility; device durability;
D O I
10.1016/j.jneumeth.2008.06.024
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
We developed a 64-channel flexible polyimide ECoG electrode array and characterized its performance for long-term implantation, chronic cortical recording and high resolution mapping of surface-evoked potentials in awake rats. To achieve the longest possible recording periods, the flexibility of the electrode array, adhesion between the metals and carrier substrate, and biocompatibility were critical for maintaining the signal integrity. Experimental testing of thin film adhesion was applied to a gold-polyimide system in order to characterize relative interfacial fracture energies for several different adhesion layers, yielding an increase in overall device reliability. We tested several different adhesion techniques including the following: gold alone without an adhesion layer, titanium-tungsten, tantalum and chromium. We found titanium-tungsten to be a suitable adhesion layer considering the biocompatibility requirements as well as stability and delamination resistance. While chromium and tantalum produced stronger gold adhesion, concerns over biocompatibility of these materials require further testing. We implanted the polyimide ECoG electrode arrays through a slit made in the skull of rats and recorded cortical surface evoked responses. The arrays performed reliably over a period of at least 100 days and signals compared well with traditional screw electrodes, with better high frequency response characteristics. Since the ultimate goal of chronically implanted electrode arrays is for neural prosthetic devices that need to last many decades, other adhesion layers that would prove safe for implantation may be tested in the same way in order to improve the device reliability. (c) 2008 Elsevier B.V. All rights reserved.
引用
收藏
页码:279 / 285
页数:7
相关论文
共 33 条
[1]   Effects of insertion conditions on tissue strain and vascular damage during neuroprosthetic device insertion [J].
Bjornsson, C. S. ;
Oh, S. J. ;
Al-Kofahi, Y. A. ;
Lim, Y. J. ;
Smith, K. L. ;
Turner, J. N. ;
De, S. ;
Roysam, B. ;
Shain, W. ;
Kim, S. J. .
JOURNAL OF NEURAL ENGINEERING, 2006, 3 (03) :196-207
[2]   Flexible polyimide microelectrode array for in vivo recordings and current source density analysis [J].
Cheung, Karen C. ;
Renaud, Philippe ;
Tanila, Heikki ;
Djupsund, Kaj .
BIOSENSORS & BIOELECTRONICS, 2007, 22 (08) :1783-1790
[3]   Adhesion and debonding of multi-layer thin film structures [J].
Dauskardt, R ;
Lane, M ;
Ma, Q ;
Krishna, N .
ENGINEERING FRACTURE MECHANICS, 1998, 61 (01) :141-162
[4]   ADHESION MECHANISM OF GOLD-UNDERLAYER FILM COMBINATIONS TO OXIDE SUBSTRATES [J].
HAQ, KE ;
BEHRNDT, KH ;
KOBIN, I .
JOURNAL OF VACUUM SCIENCE & TECHNOLOGY, 1969, 6 (01) :148-&
[5]  
HETKE JF, 1990, SENSOR ACTUAT A-PHYS, V21, P999
[6]   A MEMS fabricated flexible electrode array for recording surface field potentials [J].
Hollenberg, Brian A. ;
Richards, Cecilia D. ;
Richards, Robert ;
Bahr, David F. ;
Rector, David M. .
JOURNAL OF NEUROSCIENCE METHODS, 2006, 153 (01) :147-153
[7]   Initiation and arrest of an interfacial crack in a four-point bend test [J].
Huang, ZY ;
Suo, Z ;
Xu, GH ;
He, J ;
Prévost, JH ;
Sukumar, N .
ENGINEERING FRACTURE MECHANICS, 2005, 72 (17) :2584-2601
[8]   Four-point bend adhesion measurements of copper and permalloy systems [J].
Hughey, MP ;
Morris, DJ ;
Cook, RF ;
Bozeman, SP ;
Kelly, BL ;
Chakravarty, SLN ;
Harkens, DP ;
Stearns, LC .
ENGINEERING FRACTURE MECHANICS, 2004, 71 (02) :245-261
[9]   The effect of sputtered interface metallic layers on reinforced core laminate making build-up structures [J].
Jalonen, P ;
Tuominen, A .
MICROELECTRONICS RELIABILITY, 2002, 42 (07) :1075-1079
[10]   Spatiotemporal organization of fast (>200 Hz) electrical oscillations in rat vibrissa/barrel cortex [J].
Jones, MS ;
Barth, DS .
JOURNAL OF NEUROPHYSIOLOGY, 1999, 82 (03) :1599-1609