Diagnostic accuracy of a portable recording device (MESAM IV) in suspected obstructive sleep apnoea

被引:79
作者
Esnaola, S
Duran, J
InfanteRivard, C
Rubio, R
Fernandez, A
机构
[1] HOSP TXAGORRITXU, SLEEP UNIT, VITORIA, SPAIN
[2] MCGILL UNIV, FAC MED, DEPT OCCUPAT HLTH, MONTREAL, PQ, CANADA
关键词
heart rate; monitoring (physiologic); obstructive sleep apnoea syndrome; oxygen; snoring;
D O I
10.1183/09031936.96.09122597
中图分类号
R56 [呼吸系及胸部疾病];
学科分类号
摘要
This study evaluated the accuracy of a portable recording device (MESAM IV) in identifying obstructive sleep apnoea (OSA). The MESAM IV system measures arterial oxygen saturation (Sa,O-2), heart rate, snoring sounds and body position, and allows both automatic and manual scoring of the recordings, Nocturnal polysomnography and MESAM IV recordings were performed simultaneously in 150 patients with suspected OSA, and were analysed blindly by a different observer. Patients with an apnoea-hypopnoea index (AHI) greater than or equal to 10 were diagnosed as having OSA, In the evaluation of the discriminatory ability of MESAM IV scores, the cut-off point was set to minimize first the exclusion of truly diseased patients (i.e. false-negative interpretations), and then the confirmation of nondiseased subjects (i.e. false-positives). When used as an exclusion test, the portable device reached a sensitivity of 0.98 and a specificity of 0.78; as a confirmation test, these values were 0.69 and 0.97, respectively, These results were achieved with manual scoring, which was superior to automatic scoring, Manual scoring was also better than automatic scoring when OSA was defined according to other threshold values (greater than or equal to, 15 and 20) for the AHI. The combination of MESAM IV manual scores could reduce the need for diagnostic polysomnography in three quarters of the patients clinically suspected of having obstructive sleep apnoea, substantially reducing costs associated with diagnostic procedures.
引用
收藏
页码:2597 / 2605
页数:9
相关论文
共 27 条
  • [1] [Anonymous], 1989, AM REV RESPIR DIS, V139, P559
  • [2] LUNG-FUNCTION TESTING - SELECTION OF REFERENCE VALUES AND INTERPRETATIVE STRATEGIES
    不详
    [J]. AMERICAN REVIEW OF RESPIRATORY DISEASE, 1991, 144 (05): : 1202 - 1218
  • [3] ARMSTRONG BK, 1994, PRINCIPLES EXPOSURE, P104
  • [4] STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
    BLAND, JM
    ALTMAN, DG
    [J]. LANCET, 1986, 1 (8476) : 307 - 310
  • [5] CANTOLLA JD, 1994, ARCH BRONCONEUMOL, V30, P331
  • [6] A LEISURELY LOOK AT THE BOOTSTRAP, THE JACKKNIFE, AND CROSS-VALIDATION
    EFRON, B
    GONG, G
    [J]. AMERICAN STATISTICIAN, 1983, 37 (01) : 36 - 48
  • [7] EFFECT OF CONTINUOUS POSITIVE AIRWAY PRESSURE TREATMENT ON DAYTIME FUNCTION IN SLEEP APNOEA/HYPOPNOEA SYNDROME
    ENGLEMAN, HM
    MARTIN, SE
    DEARY, IJ
    DOUGLAS, NJ
    [J]. LANCET, 1994, 343 (8897) : 572 - 575
  • [8] Esnaola S., 1995, European Respiratory Journal, V8, p436S
  • [9] PORTABLE RECORDING IN THE ASSESSMENT OF OBSTRUCTIVE SLEEP-APNEA
    FERBER, R
    MILLMAN, R
    COPPOLA, M
    FLEETHAM, J
    MURRAY, CF
    IBER, C
    MCCALL, V
    NINOMURCIA, G
    PRESSMAN, M
    SANDERS, M
    STROHL, K
    VOTTERI, B
    WILLIAMS, A
    [J]. SLEEP, 1994, 17 (04) : 378 - 392
  • [10] FINDLEY LJ, 1992, CLIN CHEST MED, V13, P427