Digital pulse-shape discrimination of fast neutrons and γ rays

被引:80
作者
Soderstrom, P. -A. [1 ]
Nyberg, J. [1 ]
Wolters, R. [1 ]
机构
[1] Uppsala Univ, Dept Phys & Astron, SE-75121 Uppsala, Sweden
基金
瑞典研究理事会;
关键词
digital pulse-shape discrimination; fast-neutron detection; liquid scintillator; BC-501; sampling ADC;
D O I
10.1016/j.nima.2008.06.004
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Discrimination of the detection of fast neutrons and 7 rays in a liquid scintillator detector has been investigated using digital pulse-processing techniques. An experimental setup with a Cf-252 source, a BC-501 liquid scintillator detector, and a BaF2 detector was used to collect waveforms with a 100 Ms/s, 14 bit sampling ADC. Three identical ADCs were combined to increase the sampling frequency to 300 Ms/s. Four different digital pulse-shape analysis algorithms were developed and compared to each other and to data obtained with an analogue neutron-gamma discrimination unit. Two of the digital algorithms were based on the charge comparison method, while the analogue unit and the other two digital algorithms were based on the zero-crossover method. Two different figure-of-merit parameters, which quantify the neutron-gamma discrimination properties, were evaluated for all four digital algorithms and for the analogue data set. All of the digital algorithms gave similar or better figure-of-merit values than what was obtained with the analogue setup. A detailed study of the discrimination properties as a function of sampling frequency and bit resolution of the ADC was performed. it was shown that a sampling ADC with a bit resolution of 12 bits and a sampling frequency of 100 Ms/s is adequate for achieving an optimal neutron-gamma discrimination for pulses having a dynamic range for deposited neutron energies of 0.3-12 MeV. An investigation of the influence of the sampling frequency on the time resolution was made. A FWHM of 1.7 ns was obtained at 100 Ms/s. (C) 2008 Elsevier B.V. All rights reserved.
引用
收藏
页码:79 / 89
页数:11
相关论文
共 29 条
  • [1] AN AMPLITUDE-INSENSITIVE SYSTEM THAT DISTINGUISHES PULSES OF DIFFEREENT SHAPES
    ALEXANDER, TK
    GOULDING, FS
    [J]. NUCLEAR INSTRUMENTS & METHODS, 1961, 13 (02): : 244 - 246
  • [2] A 2-PI NEUTRON AND GAMMA-RAY MULTIPLICITY FILTER FOR THE NORDBALL DETECTION SYSTEM
    ARNELL, SE
    ROTH, HA
    SKEPPSTEDT, O
    BIALKOWSKI, J
    MOSZYNSKI, M
    WOLSKI, D
    NYBERG, J
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1991, 300 (02) : 303 - 311
  • [3] Verification of the digital discrimination of neutrons and γ rays using pulse gradient analysis by digital measurement of time of flight
    Aspinall, M. D.
    D'Mellow, B.
    Mackin, R. O.
    Joyce, M. J.
    Hawkes, N. P.
    Thomas, D. J.
    Jarrah, Z.
    Peyton, A. J.
    Nolan, P. J.
    Boston, A. J.
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2007, 583 (2-3) : 432 - 438
  • [4] A SCINTILLATION COUNTER WITH NEUTRON AND GAMMA-RAY DISCRIMINATORS
    BROOKS, FD
    [J]. NUCLEAR INSTRUMENTS & METHODS, 1959, 4 (03): : 151 - 163
  • [5] IMPROVED PREDICTIONS OF NEUTRON DETECTION EFFICIENCY FOR HYDROCARBON SCINTILLATORS FROM 1 MEV TO ABOUT 300 MEV
    CECIL, RA
    ANDERSON, BD
    MADEY, R
    [J]. NUCLEAR INSTRUMENTS & METHODS, 1979, 161 (03): : 439 - 447
  • [6] Relative enhancement of weak two-neutron exit channels in heavy-ion induced fusion-evaporation reactions
    Cederkall, J
    Cederwall, B
    Johnson, A
    Palacz, M
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1997, 385 (01) : 166 - 170
  • [7] 1n and 2n transfer with the borromean nucleus 6He near the Coulomb barrier
    Chatterjee, A.
    Navin, A.
    Shrivastava, A.
    Bhattacharyya, S.
    Rejmund, M.
    Keeley, N.
    Nanal, V.
    Nyberg, J.
    Pillay, R. G.
    Ramachandran, K.
    Stefan, I.
    Bazin, D.
    Beaumel, D.
    Blumenfeld, Y.
    de France, G.
    Gupta, D.
    Labiche, M.
    Lemasson, A.
    Lemmon, R.
    Raabe, R.
    Scarpaci, J. A.
    Simenel, C.
    Timis, C.
    [J]. PHYSICAL REVIEW LETTERS, 2008, 101 (03)
  • [8] Digital discrimination of neutrons and γ-rays in liquid scintillators using pulse gradient analysis
    D'Mellow, B.
    Aspinall, M. D.
    Mackin, R. O.
    Joyce, M. J.
    Peyton, A. J.
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2007, 578 (01) : 191 - 197
  • [9] Friedman J. H., 1974, Proceedings of the 1974 CERN School of Computing, P271
  • [10] Gatti E., 1962, NUCL ELECT, V2, P265