NaI detector nonlinearity for PGNAA applications

被引:16
作者
Gardner, RP [1 ]
Mayo, CW [1 ]
机构
[1] N Carolina State Univ, Ctr Engn Applicat Radioisotopes, Raleigh, NC 27695 USA
关键词
prompt gamma; NAI nonlinearity; PGNAA;
D O I
10.1016/S0969-8043(98)00183-3
中图分类号
O61 [无机化学];
学科分类号
070301 ; 081704 ;
摘要
The nonlinearity of a 6 " x 6 " NaI detector used for PGNAA has been measured. The prompt gamma-rays from pure element samples mixed with graphite were used for this determination in the range from 1.712 to 10.829 MeV and gamma-rays from radioisotope sources were used in the range from 0.356 to 1.333 MeV. A surprising result is that the pulse height per unit absorbed energy is slightly higher from single escape as opposed to full energy peaks. One explanation for this is that the light production efficiency for positrons may be different than that for electrons. By interpolation a value for the coincidence spectral full energy peak exhibited by NaI detectors that are neutron activated is obtained. The value reported is 6.885 MeV which is slightly higher than the cross section weighted value of 6.834 MeV. (The reported S-N values for I and Na are 6.826 and 6.959 MeV, respectively.) (C) 1999 Elsevier Science Ltd. All rights reserved.
引用
收藏
页码:189 / 195
页数:7
相关论文
共 7 条
[1]  
Bevington R., 1969, DATA REDUCTION ERROR
[2]  
GARDNER RP, 1998, ANAL DESIGN RADIATIO
[3]  
Heath R.L., 1964, IDO168801
[4]  
MARQUARDT DW, 1963, J SOC IND APPL MATH, V11, P2
[5]  
MARSHALL JH, 1989, NUCL GEOPHYS, V3, P445
[6]  
SAMON L, 1961, NUCL INSTRUM METHODS, V14, P193
[7]  
SHYU CM, 1993, NUCL GEOPHYS, V7, P241