Effect of Computed Tomography Scanning Parameters on Gold Nanoparticle and Iodine Contrast

被引:86
作者
Galper, Merav Weill [1 ]
Saung, May Tun [1 ]
Fuster, Valentin [2 ,3 ]
Roessl, Ewald [4 ]
Thran, Axel [4 ]
Proksa, Roland [4 ]
Fayad, Zahi Adel [1 ]
Cormode, David Peter [1 ]
机构
[1] Mt Sinai Sch Med, Translat & Mol Imaging Inst, New York, NY 10029 USA
[2] Mt Sinai Sch Med, Zena & Michael A Wiener Cardiovasc Inst, New York, NY 10029 USA
[3] CNIC, Madrid, Spain
[4] Philips Res Europe, Hamburg, Germany
基金
美国国家卫生研究院;
关键词
computed tomography; gold nanoparticles; contrast agents; iodine; BLOOD-POOL; AGENT; CT; MICE; DELIVERY;
D O I
10.1097/RLI.0b013e3182562ab9
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: Gold nanoparticles (gold-NPs) have lately been proposed as alternative contrast agents to iodine-based contrast agents (iodine-CA) for computed tomography (CT) angiography. The aims of this study were to confirm an appropriate environment in which to evaluate such novel contrast agents, to investigate the comparative contrast of iodine-CA versus gold-NP, and to determine optimal scanning parameters for gold-NP. Materials and Methods: Three different clinical scanners were used to acquire CT images. A range of concentrations (10 mM to 1.5 M) of gold-NP and iodine-CA were scanned with varying x-ray tube voltages and currents, reconstruction kernels, protocols, and scanner models. The different environments investigated were air, water, and water with a bone simulant (Ca-3(PO4)(2)). Regression coefficients were derived from the attenuation values plotted against concentration and compared for statistical significance using t values. Results: As expected, contrast was linearly related to concentrations up to 500 to 1000 mM, depending on the conditions used, whereupon a plateau of 3000 Hounsfield units was reached. Attenuation was significantly different depending on the environment used (air, water, or water and bone simulant). Contrast is dependent on the x-ray tube voltage used, with the contrast produced from iodine-CA sharply declining with increasing voltage, whereas the contrast of gold-NP varied less with tube voltage but was maximal at 120 kV in water with bone simulant. Current, reconstruction kernels, protocols, and scanner model had less effect on contrast. Conclusion: Water with a bone simulant is a preferable environment for evaluating novel cardiac CT contrast agents. Relative iodine-CA versus gold-NP contrast is dependent on the scanning conditions used. Optimal scanning conditions for gold-NP will likely use an x-ray tube voltage of 120 kV.
引用
收藏
页码:475 / 481
页数:7
相关论文
共 50 条
  • [41] Computed tomography numbers obtained for varying iodine contrast concentrations by different-generation dual-energy computed tomography scanners
    Tsukamoto, Tomokatsu
    Masuda, Takanori
    Takahata, Takashi
    Kawamoto, Yoshinori
    Uenaka, Osamu
    Mori, Hiroki
    RADIATION PROTECTION DOSIMETRY, 2024, 200 (14) : 1358 - 1364
  • [42] Gold Nanoparticle Quantitation by Whole Cell Tomography
    Sanders, Aric W.
    Jeerage, Kavita M.
    Schwartz, Cindi L.
    Curtin, Alexandra E.
    Chiaramonti, Ann N.
    ACS NANO, 2015, 9 (12) : 11792 - 11799
  • [43] Poly(acrylic acid) Bridged Gadolinium Metal-Organic Framework-Gold Nanoparticle Composites as Contrast Agents for Computed Tomography and Magnetic Resonance Bimodal Imaging
    Tian, Chixia
    Zhu, Liping
    Lin, Feng
    Boyes, Stephen G.
    ACS APPLIED MATERIALS & INTERFACES, 2015, 7 (32) : 17765 - 17775
  • [44] Nanoparticles as computed tomography contrast agents: current status and future perspectives
    Shilo, Malka
    Reuveni, Tobi
    Motiei, Menachem
    Popovtzer, Rachela
    NANOMEDICINE, 2012, 7 (02) : 257 - 269
  • [45] Nanoparticle contrast-enhanced computed tomography and magnetic resonance imaging of vascularization of a subcutaneous niche for islet transplantation
    Capuani, Simone
    Campa-Carranza, Jocelyn Nikita
    Hernandez, Nathanael
    Menon, Renuka T. R.
    Bhavane, Rohan
    Rome, Gabrielle E.
    Devkota, Laxman
    Ghaghada, Ketan B.
    Annapragada, Ananth V.
    Chua, Corrine Ying Xuan
    Badachhape, Andrew A.
    Grattoni, Alessandro
    BIOENGINEERING & TRANSLATIONAL MEDICINE, 2024,
  • [46] Volumetric computed tomography: Principles, parameters
    Jankowski, A.
    Ferretti, G.
    REVUE DES MALADIES RESPIRATOIRES, 2010, 27 (08) : 964 - 969
  • [47] The Comparison of Computed Tomography Perfusion, Contrast-Enhanced Computed Tomography and Positron-Emission Tomography/Computed Tomography for the Detection of Primary Esophageal Carcinoma
    Genc, Berhan
    Kantarci, Mecit
    Sade, Recep
    Orsal, Ebru
    Ogul, Hayri
    Okur, Aylin
    Aydin, Yener
    Karaca, Leyla
    Eroglu, Atilla
    MEDICAL PRINCIPLES AND PRACTICE, 2016, 25 (03) : 254 - 259
  • [48] Folic acid-cysteamine modified gold nanoparticle as a nanoprobe for targeted computed tomography imaging of cancer cells
    Khademi, Sara
    Sarkar, Saeed
    Shakeri-Zadeh, Ali
    Attaran, Neda
    Kharrazi, Sharmin
    Ay, Mohammad Reza
    Ghadiri, Hossein
    MATERIALS SCIENCE & ENGINEERING C-MATERIALS FOR BIOLOGICAL APPLICATIONS, 2018, 89 : 182 - 193
  • [49] Effect of contrast phase on quantitative analysis of skeletal muscle and adipose tissue by computed tomography
    Vedder, Issi R.
    Levolger, Stef
    Dierckx, Rudi A. J. O.
    Viddeleer, Alain R.
    Bokkers, Reinoud P. H.
    NUTRITION, 2024, 126
  • [50] Dual energy computed tomography analysis in cancer patients: What factors affect iodine concentration in contrast enhanced studies?
    Corrias, Giuseppe
    Sawan, Peter
    Mahmood, Usman
    Zheng, Junting
    Capanu, Marinela
    Salvatore, Marco
    Spinato, Giacomo
    Saba, Luca
    Mannelli, Lorenzo
    EUROPEAN JOURNAL OF RADIOLOGY, 2019, 120