Comparison of dinamap PRO-100 and mercury sphygmomanometer blood pressure measurements in a population-based study

被引:36
作者
Ni, HY
Wu, C
Prineas, R
Shea, S
Liu, K
Kronmal, R
Bild, D
机构
[1] NHLBI, Epidemiol Biometry Program, Div Epidemiol & Clin Applicat, NIH, Bethesda, MD 20892 USA
[2] Wake Forest Univ, Sch Med, Winston Salem, NC 27109 USA
[3] Columbia Univ, Dept Med & Epidemiol, New York, NY 10027 USA
[4] Northwestern Univ, Sch Publ Hlth, Chicago, IL 60208 USA
[5] Univ Washington, Sch Publ Hlth, Seattle, WA 98195 USA
基金
美国国家卫生研究院;
关键词
blood pressure; sphygmomanometer; Dinamap monitor;
D O I
10.1016/j.amjhyper.2005.10.020
中图分类号
R6 [外科学];
学科分类号
1002 ; 100210 ;
摘要
Background: The accuracy of automated oscillometric devices has been questioned. In addition the acceptability of these devices for research under the field conditions is unclear. Methods: We compared blood pressure (BP) readings obtained using the Dinamap PRO-100 with readings obtained using a standard mercury sphygmomanometer in 305 participants aged 48 to 86 years who were enrolled in the ongoing Multi-Ethnic Study of Atherosclerosis. The BP was measured three times by each device in random order in each participant. Results: Approximately one half of the participants were male and 46.6% had hypertension. The Dinamap and mercury measurements were well correlated (r = 0.89 for systolic BP and r = 0.81 for diastolic BP). Overall the Dinamap underestimated BP: the mean difference (Dinamap - mercury sphygmomanometer) was -0.5 mm Hg (P = .36, SD = 9.8 mm Hg) for systolic BP and -2.9 mm Hg (P < .001, SD = 6.6 mm Hg) for diastolic BP. However, the Dinamap device tended to overestimate systolic BP in participants who were 75 to 86 years of age, who had a pulse pressure >= 60 mm Hg, or who had stages I to III hypertension. On the other hand, the Dinamap underestimated diastolic BP among these same subgroups but with a smaller underestimate than for the rest of the study sample. Conclusions: Although the BP measurements obtained by the Dinamap PRO-100 tend on average to be slightly lower than those obtained by the standard mercury sphygmomanometer in middle-aged and older persons, the discrepancies may vary with age, pulse pressure, and BP. Health care providers and researchers should know of this variation and should interpret Dinamap-measured BP with caution. (c) 2006 American Journal of Hypertension, Ltd.
引用
收藏
页码:353 / 360
页数:8
相关论文
共 31 条
[1]  
[Anonymous], 1993, ANSI AAMI SP
[2]   Accuracy of the Dinamap 1846 XT automated blood pressure monitor [J].
Beaubien, ER ;
Card, CM ;
Card, SE ;
Biem, HJ ;
Wilson, TW .
JOURNAL OF HUMAN HYPERTENSION, 2002, 16 (09) :647-652
[3]   Multi-ethnic study of atherosclerosis: Objectives and design [J].
Bild, DE ;
Bluemke, DA ;
Burke, GL ;
Detrano, R ;
Roux, AVD ;
Folsom, AR ;
Greenland, P ;
Jacobs, DR ;
Kronmal, R ;
Liu, K ;
Nelson, JC ;
O'Leary, D ;
Saad, MF ;
Shea, S ;
Szklo, M ;
Tracy, RP .
AMERICAN JOURNAL OF EPIDEMIOLOGY, 2002, 156 (09) :871-881
[4]   Measuring agreement in method comparison studies [J].
Bland, JM ;
Altman, DG .
STATISTICAL METHODS IN MEDICAL RESEARCH, 1999, 8 (02) :135-160
[5]   The aging of elastic and muscular arteries - A comparison of diabetic and nondiabetic subjects [J].
Cameron, JD ;
Bulpitt, CJ ;
Pinto, ES ;
Rajkumar, C .
DIABETES CARE, 2003, 26 (07) :2133-2138
[6]   Seventh Report of the Joint National Committee on Prevention, Detection, Evaluation, and Treatment of High Blood Pressure [J].
Chobanian, AV ;
Bakris, GL ;
Black, HR ;
Cushman, WC ;
Green, LA ;
Izzo, JL ;
Jones, DW ;
Materson, BJ ;
Oparil, S ;
Wright, JT ;
Roccella, EJ .
HYPERTENSION, 2003, 42 (06) :1206-1252
[7]  
Gardner AW, 1998, ANGIOLOGY, V49, P723
[8]  
GOONASEKERA CDA, 1995, J HUM HYPERTENS, V9, P885
[9]   COMPARISON OF AUTOMATIC OSCILLOMETRIC ARTERIAL-PRESSURE MEASUREMENT WITH CONVENTIONAL AUSCULTATORY MEASUREMENT IN THE LABOR WARD [J].
HASAN, MA ;
THOMAS, TA ;
PRYSROBERTS, C .
BRITISH JOURNAL OF ANAESTHESIA, 1993, 70 (02) :141-144
[10]   Coronary artery calcification and aortic pulse wave velocity in chronic kidney disease patients [J].
Haydar, AA ;
Covic, A ;
Colhoun, H ;
Rubens, M ;
Goldsmith, DJA .
KIDNEY INTERNATIONAL, 2004, 65 (05) :1790-1794