Quench protection design of a 9.4 T whole-body MRI superconducting magnet

被引:5
作者
Chen, Shunzhong [1 ]
Li, Yi [1 ]
Dai, Yinming [1 ]
Lei, Yuanzhong [1 ]
Yan, Luguang [1 ]
机构
[1] Chinese Acad Sci, Inst Elect Engn, Key Lab Appl Superconduct, Beijing 100190, Peoples R China
来源
PHYSICA C-SUPERCONDUCTIVITY AND ITS APPLICATIONS | 2014年 / 497卷
关键词
MRI; Quench; Protection; Heater;
D O I
10.1016/j.physc.2013.11.001
中图分类号
O59 [应用物理学];
学科分类号
摘要
A 9.4 T MRI superconducting magnet with a 800 mm clear warm bore in diameter is designed and fabricated for bioscience research. The superconducting magnet consisting of five coaxial solenoid coils is fabricated with NbTi Wire-in-Channel (WIC) conductor where the ratios of copper to non-copper are from 5 to 10. The four compensation solenoid coils are with rectangular NbTi/Cu strand wires. The magnet will be operated in a relative low nominal current of 224.5 A with a high level of stored energy, about 138 MJ. A protection method with the cold diodes and resistors in series across the subdivided sections and active trigger heater to accelerate quench is adopted to avoid the damage of the magnet. In the paper, the quench simulation results of currents, voltages and hot-spot temperatures based on the protection scheme are analyzed in details. (C) 2013 Elsevier B.V. All rights reserved.
引用
收藏
页码:49 / 53
页数:5
相关论文
共 50 条
  • [31] The Role of Whole-Body MRI in Pediatric Oncology
    Atkin, Karen L.
    Ditchfield, Michael R.
    JOURNAL OF PEDIATRIC HEMATOLOGY ONCOLOGY, 2014, 36 (05) : 342 - 352
  • [32] Quench Analysis of 5.8 T Conduction-Cooled Superconducting Magnet
    Zhang, Yongliang
    Xu, Aihua
    Dai, Chao
    Hao, Qiangwang
    Wu, Yu
    Ma, Yuanyuan
    Wu, Kaihong
    IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 2021, 31 (05)
  • [33] Whole-Body Functional MRI and PET/MRI in Multiple Myeloma
    Mule, Sebastien
    Reizine, Edouard
    Blanc-Durand, Paul
    Baranes, Laurence
    Zerbib, Pierre
    Burns, Robert
    Nouri, Refaat
    Itti, Emmanuel
    Luciani, Alain
    CANCERS, 2020, 12 (11) : 1 - 11
  • [34] Diagnostic accuracy of whole-body PET/MRI and whole-body PET/CT for TNM staging in oncology
    Heusch, Philipp
    Nensa, Felix
    Schaarschmidt, Benedikt
    Sivanesapillai, Rupika
    Beiderwellen, Karsten
    Gomez, Benedikt
    Koehler, Jens
    Reis, Henning
    Ruhlmann, Verena
    Buchbender, Christian
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2015, 42 (01) : 42 - 48
  • [35] Commissioning of the Iseult CEA 11.7 T whole-body MRI: current status, gradient–magnet interaction tests and first imaging experience
    Nicolas Boulant
    Lionel Quettier
    Magnetic Resonance Materials in Physics, Biology and Medicine, 2023, 36 : 175 - 189
  • [36] Three-Line Microstrip Array for Whole-Body MRI System at 7 T
    Hernandez, Daniel
    APPLIED SCIENCES-BASEL, 2021, 11 (01): : 1 - 11
  • [37] Analysis of Powering and Quench Protection of the SIGRUM Superconducting Combined-Function Dipole Magnet
    Ferrentino, Vittorio
    Arpaia, Pasquale
    Gilardi, Antonio
    Karppinen, Mikko
    Kokkinos, Charilaos
    Ravaioli, Emmanuele
    IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 2023, 33 (07)
  • [38] Whole-body MRI for Detecting Bone Marrow Metastases
    Kwee, Thomas C.
    Takahara, Taro
    Katahira, Kazuhiro
    Nakanishi, Katsuyuki
    PET CLINICS, 2010, 5 (03) : 297 - 309
  • [39] Postmortem Whole-Body MRI in Traumatic Causes of Death
    Ross, Steffen
    Ebner, Lukas
    Flach, Patricia
    Brodhage, Rolf
    Bolliger, Stephan A.
    Christe, Andreas
    Thali, Michael J.
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2012, 199 (06) : 1186 - 1192
  • [40] Commissioning of the Iseult CEA 11.7 T whole-body MRI: current status, gradient-magnet interaction tests and first imaging experience
    Boulant, Nicolas
    Quettier, Lionel
    MAGNETIC RESONANCE MATERIALS IN PHYSICS BIOLOGY AND MEDICINE, 2023, 36 (02) : 175 - 189