Smart pumps and random safety audits in a Neonatal Intensive Care Unit: a new challenge for patient safety

被引:11
作者
Bergon-Sendin, Elena [1 ]
Perez-Grande, Carmen [1 ]
Lora-Pablos, David [1 ]
Teresa Moral-Pumarega, Maria [1 ]
Melgar-Bonis, Ana [1 ]
Pea-Peloche, Carmen [1 ]
Diezma-Rodino, Mercedes [1 ]
Garcia-San Jose, Lidia [1 ]
Cabanes-Alonso, Esther [1 ]
Rosa Pallas-Alonso, Carmen [1 ]
机构
[1] 12 Octubre Univ Hosp, Biomed Res Inst I 12, Dept Neonatol, Madrid 28041, Spain
来源
BMC PEDIATRICS | 2015年 / 15卷
关键词
Smart pumps; Random safety audits; Technology; Patient safety; Adverse events; Neonatal Intensive Care Unit; Newborn; ADMINISTRATION ERRORS; PEDIATRIC INPATIENTS; MEDICATION ERRORS; TECHNOLOGY; IMPACT;
D O I
10.1186/s12887-015-0521-6
中图分类号
R72 [儿科学];
学科分类号
100202 ;
摘要
Background: Random safety audits (RSA) are a safety tool enabling prevention of adverse events, but they have not been widely used in hospitals. The aim of this study was to use RSAs to assess and compare the frequency of appropriate use of infusion pump safety systems in a Neonatal Intensive Care Unit (NICU) before and after quality improvement interventions and to analyse the intravenous medication programming data. Methods: Prospective, observational study comparing the frequency of appropriate use of Alaris (R) CC smart pumps through RSAs over two periods, from 1 January to 31 December 2012 and from 1 November 2014 to 31 January 2015. Appropriate use was defined as all evaluated variables being correctly programmed into the same device. Between the two periods they were established interventions to improve the use of pumps. The information recorded at the pumps with the new security system, also extracted for one year. Results: Fifty-two measurements were collected during the first period and 160 measurements during the second period. The frequency of appropriate use was 73.13 % (117/160) in the second period versus 0 % (0/52) in the first period (p < 0.0001). Information was recorded on 44,924 infusions; in 46.03 % (20,680/44,924) of cases the drug name was recorded. In 2.5 % (532/20,680) of cases there was an attempt to exceed the absolute limit. Conclusions: Random Safety Audits were a very useful tool for detecting inappropriate use of pumps in the NICU. The improvement strategies were effective for improving appropriate use and programming of the intravenous medication infusion pumps in our NICU.
引用
收藏
页数:10
相关论文
共 23 条
[1]  
Antonucci Roberto, 2014, World J Clin Pediatr, V3, P37, DOI 10.5409/wjcp.v3.i3.37
[2]   Peripheral intravenous catheter infiltration: anesthesia providers do not adhere to their own ideas of best practice [J].
Ball, Ryan D. ;
Henao, John P. ;
Ibinson, James W. ;
Metro, David G. .
JOURNAL OF CLINICAL ANESTHESIA, 2013, 25 (02) :115-120
[3]   The effect of a multifaceted educational intervention on medication preparation and administration errors in neonatal intensive care [J].
Chedoe, Indra ;
Molendijk, Harry ;
Hospes, Wobbe ;
Van den Heuvel, Edwin R. ;
Taxis, Katja .
ARCHIVES OF DISEASE IN CHILDHOOD-FETAL AND NEONATAL EDITION, 2012, 97 (06) :F449-F455
[4]   Incidence of Nonelective Removal of Percutaneously Inserted Central Catheters According to Tip Position in Neonates [J].
Costa, Priscila ;
Bueno, Mariana ;
Aparecida Alves, Angelina Maria ;
Kimura, Amelia Fumiko .
JOGNN-JOURNAL OF OBSTETRIC GYNECOLOGIC AND NEONATAL NURSING, 2013, 42 (03) :348-356
[5]  
Douglas GC, 1986, PRESSURE TREND MONIT
[6]   Impact of smart infusion technology on administration of anticoagulants (unfractionated heparin, argatroban, lepirudin, and bivalirudin) [J].
Fanikos, John ;
Fiumara, Karen ;
Baroletti, Steve ;
Luppi, Carol ;
Saniuk, Catherine ;
Mehta, Amar ;
Silverman, Jon ;
Goldhaber, Samuel Z. .
AMERICAN JOURNAL OF CARDIOLOGY, 2007, 99 (07) :1002-1005
[7]   The incidence and nature of prescribing and medication administration errors in paediatric inpatients [J].
Ghaleb, Maisoon Abdullah ;
Barber, Nick ;
Franklin, Bryony Dean ;
Wong, Ian Chi Kei .
ARCHIVES OF DISEASE IN CHILDHOOD, 2010, 95 (02) :113-118
[8]  
Gomez-Baraza C, 2014, Farm Hosp, V38, P276, DOI 10.7399/fh.2014.38.4.1116
[9]   The Use of Smart Pumps for Preventing Medication Errors [J].
Hertzel, Camille ;
Sousa, Valmi D. .
JOURNAL OF INFUSION NURSING, 2009, 32 (05) :257-267
[10]   Medication errors and adverse drug events in pediatric inpatients [J].
Kaushal, R ;
Bates, DW ;
Landrigan, C ;
McKenna, DJ ;
Clapp, MD ;
Federico, F ;
Goldmann, DA .
JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 2001, 285 (16) :2114-2120