High Resolution Ultrasonic Attenuation Measurement in Pulse-echo Setup

被引:0
|
作者
Dubey, P. K. [1 ]
Rajagopalan, S. [2 ]
Vyaghra, V. R. [2 ]
Pendsey, V. M. [3 ]
Sharma, S. J. [4 ]
机构
[1] Natl Phys Lab, Radio & Atmospher Sci Div, New Delhi 110012, India
[2] RSTM Nagpur Univ Campus, Dept Elect, Nagpur 440033, Maharashtra, India
[3] Sindhu Mahavidyalaya, Dept Elect, Nagpur 440017, Maharashtra, India
[4] SK Porwal Coll, Dept Elect, Nagpur 441002, Maharashtra, India
来源
关键词
D O I
暂无
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
In the present work, a new technique for ultrasonic attenuation measurement in pulse echo setup is designed using off the shelf components. The technique is based on the comparison of echo amplitudes with a programmable reference voltage generated by the personal computer. The designed circuit can measure the echo amplitudes up to seven echoes with a resolution of 4mV, which is sufficient for estimation of ultrasonic attenuation. It does not require the fast analog to digital converters. A program has been developed in "C" to record echo amplitudes and display ultrasonic attenuation on screen. The instrument developed is a stand alone system and does not require any other instrument except PC. The developed technique has been tested for its functionality on some standard liquids.
引用
收藏
页码:245 / 252
页数:8
相关论文
共 50 条