摘要
An analysis is carried out to investigate the accuracy of kinetic parameters obtained using surface plasmon resonance methodology with a BIAcore instrument. The Cramer Rao lower bound for the least possible variance of an estimator of the kinetic parameters is determined. Using simulations it is shown that the standard least-squares estimation technique provides estimates that achieve this bound. The theoretical and simulation results are compared with experimental data obtained from an analysis of the interaction of the myc peptide with the anti-myc antibody, 9E10. This investigation indicates that the accuracy of the results depends on the signal level which has particular relevance to the design of experiments with low signal levels. It is shown how the accuracy of the estimates of the kinetic constants depends on the kinetic constants themselves and how the accuracy of the association constants depends on the concentration of the analyte that is used in the experiment. In addition, the effects of increasing the number of data points in the analysis of dissociation data on the accuracy of the estimates are quantitated. It is also demonstrated that signal averaging of data derived from repeat sensorgrams can result in a significant decrease in the standard deviation of the estimates. (C) 1999 Academic Press.
引用
收藏
页码:49 / 59
页数:11
相关论文
共 50 条
  • [31] Influence of black phosphorous on performance of surface plasmon resonance biosensor
    Pal, Sarika
    Verma, Alka
    Prajapati, Y. K.
    Saini, J. P.
    OPTICAL AND QUANTUM ELECTRONICS, 2017, 49 (12)
  • [32] Influence of black phosphorous on performance of surface plasmon resonance biosensor
    Sarika Pal
    Alka Verma
    Y. K. Prajapati
    J. P. Saini
    Optical and Quantum Electronics, 2017, 49
  • [33] Plasmon resonance measurements in pool boiling, condensation and evaporation experiments on a nanostructured surface
    Lefevre, F.
    Al Masri, M.
    Ibrahim, J.
    Veillas, C.
    Verrier, I
    Celle, F.
    Parriaux, O.
    Jourlin, Y.
    Cioulachtjian, S.
    INTERNATIONAL JOURNAL OF HEAT AND MASS TRANSFER, 2022, 184
  • [34] Enzyme kinetic assays with surface plasmon resonance (BIAcore) based on competition between enzyme and creatinine antibody
    Stöcklein, WFM
    Behrsing, O
    Scharte, G
    Micheel, B
    Benkert, A
    Schössler, W
    Warsinke, A
    Scheller, FW
    BIOSENSORS & BIOELECTRONICS, 2000, 15 (7-8) : 377 - 382
  • [35] Kinetic analysis of human topoisomerase IIα and β DNA binding by surface plasmon resonance
    Leontiou, C
    Lightowlers, R
    Lakey, JH
    Austin, CA
    FEBS LETTERS, 2003, 554 (1-2) : 206 - 210
  • [36] Theoretical investigation of detection accuracy of surface plasmon resonance sensor with dielectric layer
    Lin, Chengyou
    Chen, Shujing
    JOURNAL OF NANOPHOTONICS, 2017, 11 (04)
  • [37] Binding of polylysine to protein kinase CK2, measured by surface plasmon resonance
    Benitez, MJ
    Mier, G
    Brione, F
    Moreno, FJ
    Jiménez, JS
    MOLECULAR AND CELLULAR BIOCHEMISTRY, 1999, 191 (1-2) : 29 - 33
  • [38] Gold Nanoparticles Used as Protein Scavengers Enhance Surface Plasmon Resonance Signal
    de Macedo, Erenildo Ferreira
    Ducatti Formaggio, Daniela Maria
    Santos, Nivia Salles
    Tada, Dayane Batista
    SENSORS, 2017, 17 (12)
  • [39] Interaction between hammerhead ribozyme and RNA substrates measured by a surface plasmon resonance biosensor
    Nyholm, T
    Andäng, M
    Bandholtz, Å
    Maijgren, C
    Persson, B
    Hotchkiss, G
    Fehniger, TE
    Larsson, S
    Åhrlund-Richter, L
    JOURNAL OF BIOCHEMICAL AND BIOPHYSICAL METHODS, 2000, 44 (1-2): : 41 - 57
  • [40] Kinetics of dissolution of Concanavalin A dextran sols in response to glucose measured by surface plasmon resonance
    Ballerstadt, R
    Schultz, JS
    SENSORS AND ACTUATORS B-CHEMICAL, 1998, 46 (01) : 50 - 55